{"title":"一种新型全自动红细胞渗透脆性分析系统的临床应用研究","authors":"Zixin Deng ,&nbsp;Yi Li ,&nbsp;Zhizhi Xiang ,&nbsp;Yi Liu ,&nbsp;Jingbo Tang ,&nbsp;Song Zhang ,&nbsp;Guangchao Zang ,&nbsp;Yingying Gao ,&nbsp;Lei Ma","doi":"10.1016/j.bbrep.2025.102224","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Existing erythrocyte osmotic fragility test (EFT) methods are constrained by subjectivity, poor reproducibility, and lack of standardization, limiting their clinical utility in thalassemia screening. This study evaluates the RA-800 Plus, a novel fully automated EFT analyzer, as a scalable tool for thalassemia screening and aims to establish reference intervals for healthy individuals. By addressing the key limitations of manual EFTs, this work seeks to promote methodological innovation and enable standardized, high-throughput screening in diverse clinical settings.</div></div><div><h3>Methods</h3><div>EFTs were performed on 273 healthy adults via an RA-800 Plus analyzer. The light scattering turbidity method was employed to establish a reference interval within the normal range. The neonatal samples were tested in the same way to determine the newborn-specific range. Moreover, 97 samples underwent dual tests of RA-800 Plus analysis and the gold standard (genetic testing) to evaluate the diagnostic performance. A total of 103 pairs of samples were detected and compared via an automated system and the traditional manual direct colorimetric method, further verifying the consistency of the two methods. The effects of anticoagulant type, sample storage, and rewarming were also investigated.</div></div><div><h3>Results</h3><div>RA-800 Plus demonstrated high reliability, with minimal influence from anticoagulants and sample handling conditions. The analyzer showed an 84 % detection rate for β-thalassemia, indicating superior effectiveness for β-thalassemia screening. The test results were consistent with the manual method (Kappa ≥0.6). ROC curve analysis confirmed the suitability of both methods for thalassemia screening, with AUCs of 0.91 for β-thalassemia and 0.72 for α-thalassemia.</div></div><div><h3>Conclusions</h3><div>The RA-800 Plus offers a fully automated, reliable, and clinically viable alternative for thalassemia screening, particularly for β-thalassemia. Its stability, accuracy, and ease of use make it a valuable tool for improving thalassemia diagnostics.</div></div>","PeriodicalId":8771,"journal":{"name":"Biochemistry and Biophysics Reports","volume":"44 ","pages":"Article 102224"},"PeriodicalIF":2.2000,"publicationDate":"2025-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Clinical application study of a novel fully automatic erythrocyte osmotic fragility analysis system\",\"authors\":\"Zixin Deng ,&nbsp;Yi Li ,&nbsp;Zhizhi Xiang ,&nbsp;Yi Liu ,&nbsp;Jingbo Tang ,&nbsp;Song Zhang ,&nbsp;Guangchao Zang ,&nbsp;Yingying Gao ,&nbsp;Lei Ma\",\"doi\":\"10.1016/j.bbrep.2025.102224\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><div>Existing erythrocyte osmotic fragility test (EFT) methods are constrained by subjectivity, poor reproducibility, and lack of standardization, limiting their clinical utility in thalassemia screening. This study evaluates the RA-800 Plus, a novel fully automated EFT analyzer, as a scalable tool for thalassemia screening and aims to establish reference intervals for healthy individuals. By addressing the key limitations of manual EFTs, this work seeks to promote methodological innovation and enable standardized, high-throughput screening in diverse clinical settings.</div></div><div><h3>Methods</h3><div>EFTs were performed on 273 healthy adults via an RA-800 Plus analyzer. The light scattering turbidity method was employed to establish a reference interval within the normal range. The neonatal samples were tested in the same way to determine the newborn-specific range. Moreover, 97 samples underwent dual tests of RA-800 Plus analysis and the gold standard (genetic testing) to evaluate the diagnostic performance. A total of 103 pairs of samples were detected and compared via an automated system and the traditional manual direct colorimetric method, further verifying the consistency of the two methods. The effects of anticoagulant type, sample storage, and rewarming were also investigated.</div></div><div><h3>Results</h3><div>RA-800 Plus demonstrated high reliability, with minimal influence from anticoagulants and sample handling conditions. The analyzer showed an 84 % detection rate for β-thalassemia, indicating superior effectiveness for β-thalassemia screening. The test results were consistent with the manual method (Kappa ≥0.6). ROC curve analysis confirmed the suitability of both methods for thalassemia screening, with AUCs of 0.91 for β-thalassemia and 0.72 for α-thalassemia.</div></div><div><h3>Conclusions</h3><div>The RA-800 Plus offers a fully automated, reliable, and clinically viable alternative for thalassemia screening, particularly for β-thalassemia. Its stability, accuracy, and ease of use make it a valuable tool for improving thalassemia diagnostics.</div></div>\",\"PeriodicalId\":8771,\"journal\":{\"name\":\"Biochemistry and Biophysics Reports\",\"volume\":\"44 \",\"pages\":\"Article 102224\"},\"PeriodicalIF\":2.2000,\"publicationDate\":\"2025-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biochemistry and Biophysics Reports\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2405580825003115\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"BIOCHEMISTRY & MOLECULAR BIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biochemistry and Biophysics Reports","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405580825003115","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"BIOCHEMISTRY & MOLECULAR BIOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

现有的红细胞渗透脆性试验(EFT)方法存在主观性、可重复性差、缺乏标准化等问题,限制了其在地中海贫血筛查中的临床应用。本研究评估了RA-800 Plus,一种新型全自动EFT分析仪,作为地中海贫血筛查的可扩展工具,旨在为健康个体建立参考区间。通过解决手工eft的主要局限性,这项工作旨在促进方法创新,并在不同的临床环境中实现标准化、高通量筛查。方法采用RA-800 Plus分析仪对273名健康成人进行sefts检测。采用光散射浊度法在正常范围内建立参考区间。新生儿样本以同样的方式进行测试,以确定新生儿特异性范围。此外,97个样本接受了RA-800 Plus分析和金标准(基因测试)的双重测试,以评估诊断性能。通过自动化系统和传统的手工直接比色法对103对样品进行检测和比较,进一步验证了两种方法的一致性。研究了抗凝剂类型、样品保存和复温对抗凝效果的影响。结果ra -800 Plus具有较高的可靠性,受抗凝剂和样品处理条件的影响最小。该分析仪对β-地中海贫血的检出率为84%,显示了β-地中海贫血筛查的优越效果。试验结果与手工方法一致(Kappa≥0.6)。ROC曲线分析证实了两种方法筛查地中海贫血的适用性,β-地中海贫血的auc为0.91,α-地中海贫血的auc为0.72。RA-800 Plus提供了一种全自动、可靠且临床可行的地中海贫血筛查方法,特别是β-地中海贫血筛查。其稳定性、准确性和易用性使其成为改进地中海贫血诊断的宝贵工具。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Clinical application study of a novel fully automatic erythrocyte osmotic fragility analysis system

Background

Existing erythrocyte osmotic fragility test (EFT) methods are constrained by subjectivity, poor reproducibility, and lack of standardization, limiting their clinical utility in thalassemia screening. This study evaluates the RA-800 Plus, a novel fully automated EFT analyzer, as a scalable tool for thalassemia screening and aims to establish reference intervals for healthy individuals. By addressing the key limitations of manual EFTs, this work seeks to promote methodological innovation and enable standardized, high-throughput screening in diverse clinical settings.

Methods

EFTs were performed on 273 healthy adults via an RA-800 Plus analyzer. The light scattering turbidity method was employed to establish a reference interval within the normal range. The neonatal samples were tested in the same way to determine the newborn-specific range. Moreover, 97 samples underwent dual tests of RA-800 Plus analysis and the gold standard (genetic testing) to evaluate the diagnostic performance. A total of 103 pairs of samples were detected and compared via an automated system and the traditional manual direct colorimetric method, further verifying the consistency of the two methods. The effects of anticoagulant type, sample storage, and rewarming were also investigated.

Results

RA-800 Plus demonstrated high reliability, with minimal influence from anticoagulants and sample handling conditions. The analyzer showed an 84 % detection rate for β-thalassemia, indicating superior effectiveness for β-thalassemia screening. The test results were consistent with the manual method (Kappa ≥0.6). ROC curve analysis confirmed the suitability of both methods for thalassemia screening, with AUCs of 0.91 for β-thalassemia and 0.72 for α-thalassemia.

Conclusions

The RA-800 Plus offers a fully automated, reliable, and clinically viable alternative for thalassemia screening, particularly for β-thalassemia. Its stability, accuracy, and ease of use make it a valuable tool for improving thalassemia diagnostics.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Biochemistry and Biophysics Reports
Biochemistry and Biophysics Reports Biochemistry, Genetics and Molecular Biology-Biophysics
CiteScore
4.60
自引率
0.00%
发文量
191
审稿时长
59 days
期刊介绍: Open access, online only, peer-reviewed international journal in the Life Sciences, established in 2014 Biochemistry and Biophysics Reports (BB Reports) publishes original research in all aspects of Biochemistry, Biophysics and related areas like Molecular and Cell Biology. BB Reports welcomes solid though more preliminary, descriptive and small scale results if they have the potential to stimulate and/or contribute to future research, leading to new insights or hypothesis. Primary criteria for acceptance is that the work is original, scientifically and technically sound and provides valuable knowledge to life sciences research. We strongly believe all results deserve to be published and documented for the advancement of science. BB Reports specifically appreciates receiving reports on: Negative results, Replication studies, Reanalysis of previous datasets.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信