Zixin Deng , Yi Li , Zhizhi Xiang , Yi Liu , Jingbo Tang , Song Zhang , Guangchao Zang , Yingying Gao , Lei Ma
{"title":"一种新型全自动红细胞渗透脆性分析系统的临床应用研究","authors":"Zixin Deng , Yi Li , Zhizhi Xiang , Yi Liu , Jingbo Tang , Song Zhang , Guangchao Zang , Yingying Gao , Lei Ma","doi":"10.1016/j.bbrep.2025.102224","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Existing erythrocyte osmotic fragility test (EFT) methods are constrained by subjectivity, poor reproducibility, and lack of standardization, limiting their clinical utility in thalassemia screening. This study evaluates the RA-800 Plus, a novel fully automated EFT analyzer, as a scalable tool for thalassemia screening and aims to establish reference intervals for healthy individuals. By addressing the key limitations of manual EFTs, this work seeks to promote methodological innovation and enable standardized, high-throughput screening in diverse clinical settings.</div></div><div><h3>Methods</h3><div>EFTs were performed on 273 healthy adults via an RA-800 Plus analyzer. The light scattering turbidity method was employed to establish a reference interval within the normal range. The neonatal samples were tested in the same way to determine the newborn-specific range. Moreover, 97 samples underwent dual tests of RA-800 Plus analysis and the gold standard (genetic testing) to evaluate the diagnostic performance. A total of 103 pairs of samples were detected and compared via an automated system and the traditional manual direct colorimetric method, further verifying the consistency of the two methods. The effects of anticoagulant type, sample storage, and rewarming were also investigated.</div></div><div><h3>Results</h3><div>RA-800 Plus demonstrated high reliability, with minimal influence from anticoagulants and sample handling conditions. The analyzer showed an 84 % detection rate for β-thalassemia, indicating superior effectiveness for β-thalassemia screening. The test results were consistent with the manual method (Kappa ≥0.6). ROC curve analysis confirmed the suitability of both methods for thalassemia screening, with AUCs of 0.91 for β-thalassemia and 0.72 for α-thalassemia.</div></div><div><h3>Conclusions</h3><div>The RA-800 Plus offers a fully automated, reliable, and clinically viable alternative for thalassemia screening, particularly for β-thalassemia. Its stability, accuracy, and ease of use make it a valuable tool for improving thalassemia diagnostics.</div></div>","PeriodicalId":8771,"journal":{"name":"Biochemistry and Biophysics Reports","volume":"44 ","pages":"Article 102224"},"PeriodicalIF":2.2000,"publicationDate":"2025-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Clinical application study of a novel fully automatic erythrocyte osmotic fragility analysis system\",\"authors\":\"Zixin Deng , Yi Li , Zhizhi Xiang , Yi Liu , Jingbo Tang , Song Zhang , Guangchao Zang , Yingying Gao , Lei Ma\",\"doi\":\"10.1016/j.bbrep.2025.102224\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><div>Existing erythrocyte osmotic fragility test (EFT) methods are constrained by subjectivity, poor reproducibility, and lack of standardization, limiting their clinical utility in thalassemia screening. This study evaluates the RA-800 Plus, a novel fully automated EFT analyzer, as a scalable tool for thalassemia screening and aims to establish reference intervals for healthy individuals. By addressing the key limitations of manual EFTs, this work seeks to promote methodological innovation and enable standardized, high-throughput screening in diverse clinical settings.</div></div><div><h3>Methods</h3><div>EFTs were performed on 273 healthy adults via an RA-800 Plus analyzer. The light scattering turbidity method was employed to establish a reference interval within the normal range. The neonatal samples were tested in the same way to determine the newborn-specific range. Moreover, 97 samples underwent dual tests of RA-800 Plus analysis and the gold standard (genetic testing) to evaluate the diagnostic performance. A total of 103 pairs of samples were detected and compared via an automated system and the traditional manual direct colorimetric method, further verifying the consistency of the two methods. The effects of anticoagulant type, sample storage, and rewarming were also investigated.</div></div><div><h3>Results</h3><div>RA-800 Plus demonstrated high reliability, with minimal influence from anticoagulants and sample handling conditions. The analyzer showed an 84 % detection rate for β-thalassemia, indicating superior effectiveness for β-thalassemia screening. The test results were consistent with the manual method (Kappa ≥0.6). ROC curve analysis confirmed the suitability of both methods for thalassemia screening, with AUCs of 0.91 for β-thalassemia and 0.72 for α-thalassemia.</div></div><div><h3>Conclusions</h3><div>The RA-800 Plus offers a fully automated, reliable, and clinically viable alternative for thalassemia screening, particularly for β-thalassemia. Its stability, accuracy, and ease of use make it a valuable tool for improving thalassemia diagnostics.</div></div>\",\"PeriodicalId\":8771,\"journal\":{\"name\":\"Biochemistry and Biophysics Reports\",\"volume\":\"44 \",\"pages\":\"Article 102224\"},\"PeriodicalIF\":2.2000,\"publicationDate\":\"2025-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biochemistry and Biophysics Reports\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2405580825003115\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"BIOCHEMISTRY & MOLECULAR BIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biochemistry and Biophysics Reports","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405580825003115","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"BIOCHEMISTRY & MOLECULAR BIOLOGY","Score":null,"Total":0}
Clinical application study of a novel fully automatic erythrocyte osmotic fragility analysis system
Background
Existing erythrocyte osmotic fragility test (EFT) methods are constrained by subjectivity, poor reproducibility, and lack of standardization, limiting their clinical utility in thalassemia screening. This study evaluates the RA-800 Plus, a novel fully automated EFT analyzer, as a scalable tool for thalassemia screening and aims to establish reference intervals for healthy individuals. By addressing the key limitations of manual EFTs, this work seeks to promote methodological innovation and enable standardized, high-throughput screening in diverse clinical settings.
Methods
EFTs were performed on 273 healthy adults via an RA-800 Plus analyzer. The light scattering turbidity method was employed to establish a reference interval within the normal range. The neonatal samples were tested in the same way to determine the newborn-specific range. Moreover, 97 samples underwent dual tests of RA-800 Plus analysis and the gold standard (genetic testing) to evaluate the diagnostic performance. A total of 103 pairs of samples were detected and compared via an automated system and the traditional manual direct colorimetric method, further verifying the consistency of the two methods. The effects of anticoagulant type, sample storage, and rewarming were also investigated.
Results
RA-800 Plus demonstrated high reliability, with minimal influence from anticoagulants and sample handling conditions. The analyzer showed an 84 % detection rate for β-thalassemia, indicating superior effectiveness for β-thalassemia screening. The test results were consistent with the manual method (Kappa ≥0.6). ROC curve analysis confirmed the suitability of both methods for thalassemia screening, with AUCs of 0.91 for β-thalassemia and 0.72 for α-thalassemia.
Conclusions
The RA-800 Plus offers a fully automated, reliable, and clinically viable alternative for thalassemia screening, particularly for β-thalassemia. Its stability, accuracy, and ease of use make it a valuable tool for improving thalassemia diagnostics.
期刊介绍:
Open access, online only, peer-reviewed international journal in the Life Sciences, established in 2014 Biochemistry and Biophysics Reports (BB Reports) publishes original research in all aspects of Biochemistry, Biophysics and related areas like Molecular and Cell Biology. BB Reports welcomes solid though more preliminary, descriptive and small scale results if they have the potential to stimulate and/or contribute to future research, leading to new insights or hypothesis. Primary criteria for acceptance is that the work is original, scientifically and technically sound and provides valuable knowledge to life sciences research. We strongly believe all results deserve to be published and documented for the advancement of science. BB Reports specifically appreciates receiving reports on: Negative results, Replication studies, Reanalysis of previous datasets.