{"title":"使用聚乙烯和聚甲基丙烯酸甲酯慢化剂的被动中子光谱仪的概念设计","authors":"Hector Rene Vega-Carrillo ,&nbsp;Ahmed M. El-Khayatt ,&nbsp;Antonio Baltazar-Raigosa ,&nbsp;Victor Martin Hernandez-Davila ,&nbsp;Ayed M. Binzowaimil ,&nbsp;Miguel Angel Salas-Luevano","doi":"10.1016/j.radphyschem.2025.113257","DOIUrl":null,"url":null,"abstract":"<div><div>Two passive neutron spectrometers were designed using Monte Carlo methods. Each spectrometer utilizes six 20.5 cm-diameter cylindrical moderators of varying thicknesses (0.5, 4.5, 8.5, 12.5, 16.5, and 20.5 cm). One spectrometer employs polyethylene moderators, while the other uses polymethylmethacrylate. Thermal neutrons are measured using pairs of thermoluminescent chips with different <sup>6</sup>Li content (TLD600 and TLD700). Responses to γ-rays of these TLDs and are approximately the same because have the same elemental composition, the same effective atomic number and similar densities. For both spectrometers, fluence response functions and absorbed doses from neutrons and secondary γ-rays were estimated for 45 monoenergetic neutron energies ranging from 10<sup>−9</sup> to 20 MeV. As monoenergetic neutrons are moderated to thermal energies, they reach the TLDs with an energy distribution, inducing <sup>6</sup>Li (n, α) reactions that define the fluence response function. Energy deposition from these neutrons and γ-rays within the TLDs allows for absorbed dose calculation. The fluence response matrices are similar for both moderator materials; however, neutron absorbed dose is consistently higher in TLD600 than in TLD700, independent of the moderator. Conversely, γ-ray absorbed dose is similar across TLD types and moderators. These spectrometers offer a smaller size and lower weight compared to a passive Bonner sphere spectrometer, while also being easier to construct and more portable.</div></div>","PeriodicalId":20861,"journal":{"name":"Radiation Physics and Chemistry","volume":"239 ","pages":"Article 113257"},"PeriodicalIF":2.8000,"publicationDate":"2025-08-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Conceptual design of Passive Neutron Spectrometers using polyethylene and polymethylmethacrylate moderators\",\"authors\":\"Hector Rene Vega-Carrillo ,&nbsp;Ahmed M. El-Khayatt ,&nbsp;Antonio Baltazar-Raigosa ,&nbsp;Victor Martin Hernandez-Davila ,&nbsp;Ayed M. Binzowaimil ,&nbsp;Miguel Angel Salas-Luevano\",\"doi\":\"10.1016/j.radphyschem.2025.113257\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><div>Two passive neutron spectrometers were designed using Monte Carlo methods. Each spectrometer utilizes six 20.5 cm-diameter cylindrical moderators of varying thicknesses (0.5, 4.5, 8.5, 12.5, 16.5, and 20.5 cm). One spectrometer employs polyethylene moderators, while the other uses polymethylmethacrylate. Thermal neutrons are measured using pairs of thermoluminescent chips with different <sup>6</sup>Li content (TLD600 and TLD700). Responses to γ-rays of these TLDs and are approximately the same because have the same elemental composition, the same effective atomic number and similar densities. For both spectrometers, fluence response functions and absorbed doses from neutrons and secondary γ-rays were estimated for 45 monoenergetic neutron energies ranging from 10<sup>−9</sup> to 20 MeV. As monoenergetic neutrons are moderated to thermal energies, they reach the TLDs with an energy distribution, inducing <sup>6</sup>Li (n, α) reactions that define the fluence response function. Energy deposition from these neutrons and γ-rays within the TLDs allows for absorbed dose calculation. The fluence response matrices are similar for both moderator materials; however, neutron absorbed dose is consistently higher in TLD600 than in TLD700, independent of the moderator. Conversely, γ-ray absorbed dose is similar across TLD types and moderators. These spectrometers offer a smaller size and lower weight compared to a passive Bonner sphere spectrometer, while also being easier to construct and more portable.</div></div>\",\"PeriodicalId\":20861,\"journal\":{\"name\":\"Radiation Physics and Chemistry\",\"volume\":\"239 \",\"pages\":\"Article 113257\"},\"PeriodicalIF\":2.8000,\"publicationDate\":\"2025-08-22\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Radiation Physics and Chemistry\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0969806X25007492\",\"RegionNum\":3,\"RegionCategory\":\"物理与天体物理\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CHEMISTRY, PHYSICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation Physics and Chemistry","FirstCategoryId":"92","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0969806X25007492","RegionNum":3,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, PHYSICAL","Score":null,"Total":0}
引用次数: 0

摘要

采用蒙特卡罗方法设计了两台无源中子能谱仪。每个光谱仪使用6个直径20.5厘米的圆柱形减速器,这些减速器的厚度不同(0.5、4.5、8.5、12.5、16.5和20.5厘米)。一台光谱仪使用聚乙烯缓和剂,而另一台使用聚甲基丙烯酸甲酯。采用不同6Li含量(TLD600和TLD700)的热释光芯片对热中子进行测量。由于具有相同的元素组成、相同的有效原子序数和相似的密度,这些tld和tld对γ射线的响应大致相同。对于这两种光谱仪,估计了45个单能中子的通量响应函数和中子和次级γ射线的吸收剂量,范围从10-9到20 MeV。当单能中子被缓和为热能时,它们以一定的能量分布到达tld,诱导6Li (n, α)反应,该反应定义了通量响应函数。这些中子和γ射线在tld内的能量沉积使吸收剂量计算成为可能。两种慢化剂材料的影响度响应矩阵相似;然而,与慢化剂无关,TLD600的中子吸收剂量始终高于TLD700。相反,γ射线吸收剂量在TLD类型和调节剂之间相似。与被动邦纳球光谱仪相比,这些光谱仪具有更小的尺寸和更轻的重量,同时也更容易构造和更便携。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Conceptual design of Passive Neutron Spectrometers using polyethylene and polymethylmethacrylate moderators
Two passive neutron spectrometers were designed using Monte Carlo methods. Each spectrometer utilizes six 20.5 cm-diameter cylindrical moderators of varying thicknesses (0.5, 4.5, 8.5, 12.5, 16.5, and 20.5 cm). One spectrometer employs polyethylene moderators, while the other uses polymethylmethacrylate. Thermal neutrons are measured using pairs of thermoluminescent chips with different 6Li content (TLD600 and TLD700). Responses to γ-rays of these TLDs and are approximately the same because have the same elemental composition, the same effective atomic number and similar densities. For both spectrometers, fluence response functions and absorbed doses from neutrons and secondary γ-rays were estimated for 45 monoenergetic neutron energies ranging from 10−9 to 20 MeV. As monoenergetic neutrons are moderated to thermal energies, they reach the TLDs with an energy distribution, inducing 6Li (n, α) reactions that define the fluence response function. Energy deposition from these neutrons and γ-rays within the TLDs allows for absorbed dose calculation. The fluence response matrices are similar for both moderator materials; however, neutron absorbed dose is consistently higher in TLD600 than in TLD700, independent of the moderator. Conversely, γ-ray absorbed dose is similar across TLD types and moderators. These spectrometers offer a smaller size and lower weight compared to a passive Bonner sphere spectrometer, while also being easier to construct and more portable.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Radiation Physics and Chemistry
Radiation Physics and Chemistry 化学-核科学技术
CiteScore
5.60
自引率
17.20%
发文量
574
审稿时长
12 weeks
期刊介绍: Radiation Physics and Chemistry is a multidisciplinary journal that provides a medium for publication of substantial and original papers, reviews, and short communications which focus on research and developments involving ionizing radiation in radiation physics, radiation chemistry and radiation processing. The journal aims to publish papers with significance to an international audience, containing substantial novelty and scientific impact. The Editors reserve the rights to reject, with or without external review, papers that do not meet these criteria. This could include papers that are very similar to previous publications, only with changed target substrates, employed materials, analyzed sites and experimental methods, report results without presenting new insights and/or hypothesis testing, or do not focus on the radiation effects.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信