siemens immulite®2000 tsi法测定促甲状腺激素受体抗体的分析和临床特点评价。

IF 1.5 4区 医学 Q4 BIOCHEMISTRY & MOLECULAR BIOLOGY
Neda Milinković, Marija Sarić, Ana Ružanović, Miloš Žarković, Jasmina Ćirić, Iva Perović Blagojević, Ana-Marija Mastilović, Svetlana Ignjatović, Biljana Nedeljković Beleslin
{"title":"siemens immulite®2000 tsi法测定促甲状腺激素受体抗体的分析和临床特点评价。","authors":"Neda Milinković, Marija Sarić, Ana Ružanović, Miloš Žarković, Jasmina Ćirić, Iva Perović Blagojević, Ana-Marija Mastilović, Svetlana Ignjatović, Biljana Nedeljković Beleslin","doi":"10.5937/jomb0-55873","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Despite commercially improved, standardised routine methods used in medical laboratories, precision laboratory medicine lacks harmonisation of results to make the laboratory result useful for its intended purpose. Furthermore, to obtain reliable laboratory results and precise diagnoses, it is important and recommended that each laboratory confirms the analytical and clinical characteristics of the method used. This study aimed to evaluate the analytical and clinical performance of the IMMULITE®2000 TSI bridge immunoassay to determine autoreactive thyroid stimulating hormone receptor antibodies (SH-R-Ab).</p><p><strong>Methods: </strong>A total of 86 patients with clinically present Graves' orbitopathy and 23 healthy volunteers as a control group were included in the study. The total TSH-R-Ab concentration was determined using an ECLIA (Elecsys Anti-TSHR Immunoassay Roche Diagnostics, GmbH, Mannheim, Germany) on the Cobas e411 analyser (Roche, Diagnostics, GmbH). The TSH-R-Ab concentration was measured using a CLIA method (IMMULITE TSI 2000, Siemens Healthcare Diagnostics, UK). The inaccuracy of the method was investigated using two levels of commercial control samples (low and high analyte concentration).</p><p><strong>Results: </strong>The results obtained meet the general minimum requirements for the analytical performance of laboratory methods (CV<5%). The overall laboratory inaccuracy was acceptable according to FDA guidelines (CV<20%). The results showed a statistically significant correlation between the analysed methods (r=0.9041, p < 0.0001) but with a relative bias of 24.5%. The best ratio of sensitivity and specificity determined by the ROC analysis (93.3% and 100%, respectively) was obtained for a cut-off value of 0.1215 IU/L, which is significantly lower compared to the cut-off value specified by the manufacturer (0.55 IU/L).</p><p><strong>Conclusions: </strong>The IMMULITE 2000 TSI bridge immunoassay for TSH-R-Ab quantification confirmed adequate precision, which is essential for routine use. However, further studies are required to evaluate its analytical specificity.</p>","PeriodicalId":16175,"journal":{"name":"Journal of Medical Biochemistry","volume":"44 3","pages":"438-446"},"PeriodicalIF":1.5000,"publicationDate":"2025-06-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12357635/pdf/","citationCount":"0","resultStr":"{\"title\":\"Evaluation of the analytical and clinical characteristics of the siemens immulite®2000 tsi method for determining thyrotropin receptor antibodies.\",\"authors\":\"Neda Milinković, Marija Sarić, Ana Ružanović, Miloš Žarković, Jasmina Ćirić, Iva Perović Blagojević, Ana-Marija Mastilović, Svetlana Ignjatović, Biljana Nedeljković Beleslin\",\"doi\":\"10.5937/jomb0-55873\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Despite commercially improved, standardised routine methods used in medical laboratories, precision laboratory medicine lacks harmonisation of results to make the laboratory result useful for its intended purpose. Furthermore, to obtain reliable laboratory results and precise diagnoses, it is important and recommended that each laboratory confirms the analytical and clinical characteristics of the method used. This study aimed to evaluate the analytical and clinical performance of the IMMULITE®2000 TSI bridge immunoassay to determine autoreactive thyroid stimulating hormone receptor antibodies (SH-R-Ab).</p><p><strong>Methods: </strong>A total of 86 patients with clinically present Graves' orbitopathy and 23 healthy volunteers as a control group were included in the study. The total TSH-R-Ab concentration was determined using an ECLIA (Elecsys Anti-TSHR Immunoassay Roche Diagnostics, GmbH, Mannheim, Germany) on the Cobas e411 analyser (Roche, Diagnostics, GmbH). The TSH-R-Ab concentration was measured using a CLIA method (IMMULITE TSI 2000, Siemens Healthcare Diagnostics, UK). The inaccuracy of the method was investigated using two levels of commercial control samples (low and high analyte concentration).</p><p><strong>Results: </strong>The results obtained meet the general minimum requirements for the analytical performance of laboratory methods (CV<5%). The overall laboratory inaccuracy was acceptable according to FDA guidelines (CV<20%). The results showed a statistically significant correlation between the analysed methods (r=0.9041, p < 0.0001) but with a relative bias of 24.5%. The best ratio of sensitivity and specificity determined by the ROC analysis (93.3% and 100%, respectively) was obtained for a cut-off value of 0.1215 IU/L, which is significantly lower compared to the cut-off value specified by the manufacturer (0.55 IU/L).</p><p><strong>Conclusions: </strong>The IMMULITE 2000 TSI bridge immunoassay for TSH-R-Ab quantification confirmed adequate precision, which is essential for routine use. However, further studies are required to evaluate its analytical specificity.</p>\",\"PeriodicalId\":16175,\"journal\":{\"name\":\"Journal of Medical Biochemistry\",\"volume\":\"44 3\",\"pages\":\"438-446\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2025-06-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12357635/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Biochemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.5937/jomb0-55873\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"BIOCHEMISTRY & MOLECULAR BIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Biochemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5937/jomb0-55873","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOCHEMISTRY & MOLECULAR BIOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景:尽管商业上改进了医学实验室使用的标准化常规方法,但精密检验医学缺乏结果的协调,无法使实验室结果对其预期目的有用。此外,为了获得可靠的实验室结果和精确的诊断,重要的是建议每个实验室确认所使用方法的分析和临床特征。本研究旨在评估IMMULITE®2000 TSI桥式免疫测定法测定自身反应性促甲状腺激素受体抗体(SH-R-Ab)的分析和临床性能。方法:选取临床表现为Graves眼病的患者86例,健康志愿者23例作为对照组。使用ECLIA (Elecsys Anti-TSHR Immunoassay, Roche Diagnostics, GmbH, Mannheim, Germany)在Cobas e411分析仪(Roche, Diagnostics, GmbH)上测定总TSH-R-Ab浓度。使用CLIA方法测量TSH-R-Ab浓度(IMMULITE TSI 2000, Siemens Healthcare Diagnostics,英国)。使用两种水平的商业对照样品(低和高分析物浓度)研究了该方法的不准确性。结果:所得结果满足实验室方法分析性能的一般最低要求(cvv)。结论:IMMULITE 2000 TSI桥式免疫分析法用于TSH-R-Ab定量具有足够的精密度,可作为常规方法使用。然而,需要进一步的研究来评估其分析特异性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of the analytical and clinical characteristics of the siemens immulite®2000 tsi method for determining thyrotropin receptor antibodies.

Background: Despite commercially improved, standardised routine methods used in medical laboratories, precision laboratory medicine lacks harmonisation of results to make the laboratory result useful for its intended purpose. Furthermore, to obtain reliable laboratory results and precise diagnoses, it is important and recommended that each laboratory confirms the analytical and clinical characteristics of the method used. This study aimed to evaluate the analytical and clinical performance of the IMMULITE®2000 TSI bridge immunoassay to determine autoreactive thyroid stimulating hormone receptor antibodies (SH-R-Ab).

Methods: A total of 86 patients with clinically present Graves' orbitopathy and 23 healthy volunteers as a control group were included in the study. The total TSH-R-Ab concentration was determined using an ECLIA (Elecsys Anti-TSHR Immunoassay Roche Diagnostics, GmbH, Mannheim, Germany) on the Cobas e411 analyser (Roche, Diagnostics, GmbH). The TSH-R-Ab concentration was measured using a CLIA method (IMMULITE TSI 2000, Siemens Healthcare Diagnostics, UK). The inaccuracy of the method was investigated using two levels of commercial control samples (low and high analyte concentration).

Results: The results obtained meet the general minimum requirements for the analytical performance of laboratory methods (CV<5%). The overall laboratory inaccuracy was acceptable according to FDA guidelines (CV<20%). The results showed a statistically significant correlation between the analysed methods (r=0.9041, p < 0.0001) but with a relative bias of 24.5%. The best ratio of sensitivity and specificity determined by the ROC analysis (93.3% and 100%, respectively) was obtained for a cut-off value of 0.1215 IU/L, which is significantly lower compared to the cut-off value specified by the manufacturer (0.55 IU/L).

Conclusions: The IMMULITE 2000 TSI bridge immunoassay for TSH-R-Ab quantification confirmed adequate precision, which is essential for routine use. However, further studies are required to evaluate its analytical specificity.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Biochemistry
Journal of Medical Biochemistry BIOCHEMISTRY & MOLECULAR BIOLOGY-
CiteScore
3.00
自引率
12.00%
发文量
60
审稿时长
>12 weeks
期刊介绍: The JOURNAL OF MEDICAL BIOCHEMISTRY (J MED BIOCHEM) is the official journal of the Society of Medical Biochemists of Serbia with international peer-review. Papers are independently reviewed by at least two reviewers selected by the Editors as Blind Peer Reviews. The Journal of Medical Biochemistry is published quarterly. The Journal publishes original scientific and specialized articles on all aspects of clinical and medical biochemistry, molecular medicine, clinical hematology and coagulation, clinical immunology and autoimmunity, clinical microbiology, virology, clinical genomics and molecular biology, genetic epidemiology, drug measurement, evaluation of diagnostic markers, new reagents and laboratory equipment, reference materials and methods, reference values, laboratory organization, automation, quality control, clinical metrology, all related scientific disciplines where chemistry, biochemistry, molecular biology and immunochemistry deal with the study of normal and pathologic processes in human beings.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信