Brian King, Jake Dove, Scott J McGonigle, Warwick A Ames, Zachary A Vesoulis
{"title":"确定两种脉搏血氧仪系统在模拟新生儿危急情况下的性能差异。","authors":"Brian King, Jake Dove, Scott J McGonigle, Warwick A Ames, Zachary A Vesoulis","doi":"10.1038/s41372-025-02364-4","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Pulse oximetry is used to guide critical clinical decisions in neonatology. We used a vital signs simulator to compare performance of two pulse oximetry systems in conditions not tested in standardized clinical verification studies.</p><p><strong>Study design: </strong>We devised a set of simulated tissue translucency, perfusion, peripheral oxygen saturation (SpO<sub>2</sub>)<sub>,</sub> and heart rate (HR) parameters to mimic challenging real-world neonatal data and applied them to two marketed pulse oximetry systems (Nellcor™ and Masimo®). At each combination of input parameters, we used the response from both systems to assess SpO<sub>2</sub> error.</p><p><strong>Results: </strong>The mean SpO<sub>2</sub> error for Nellcor™ was below 1.1% across all parameters explored, while Masimo® showed significantly higher (p < 0.005) error at lower translucencies.</p><p><strong>Conclusion: </strong>Significant performance differences can be observed when comparing pulse oximeters at low translucency and perfusion conditions. Patient simulators cannot replace clinical testing but provide a safe and cost-effective method for additional performance profiling.</p>","PeriodicalId":16690,"journal":{"name":"Journal of Perinatology","volume":" ","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2025-07-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Identifying performance differences between two pulse oximetry systems in simulated critical neonatal conditions.\",\"authors\":\"Brian King, Jake Dove, Scott J McGonigle, Warwick A Ames, Zachary A Vesoulis\",\"doi\":\"10.1038/s41372-025-02364-4\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>Pulse oximetry is used to guide critical clinical decisions in neonatology. We used a vital signs simulator to compare performance of two pulse oximetry systems in conditions not tested in standardized clinical verification studies.</p><p><strong>Study design: </strong>We devised a set of simulated tissue translucency, perfusion, peripheral oxygen saturation (SpO<sub>2</sub>)<sub>,</sub> and heart rate (HR) parameters to mimic challenging real-world neonatal data and applied them to two marketed pulse oximetry systems (Nellcor™ and Masimo®). At each combination of input parameters, we used the response from both systems to assess SpO<sub>2</sub> error.</p><p><strong>Results: </strong>The mean SpO<sub>2</sub> error for Nellcor™ was below 1.1% across all parameters explored, while Masimo® showed significantly higher (p < 0.005) error at lower translucencies.</p><p><strong>Conclusion: </strong>Significant performance differences can be observed when comparing pulse oximeters at low translucency and perfusion conditions. Patient simulators cannot replace clinical testing but provide a safe and cost-effective method for additional performance profiling.</p>\",\"PeriodicalId\":16690,\"journal\":{\"name\":\"Journal of Perinatology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2025-07-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Perinatology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1038/s41372-025-02364-4\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"OBSTETRICS & GYNECOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Perinatology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1038/s41372-025-02364-4","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OBSTETRICS & GYNECOLOGY","Score":null,"Total":0}
Identifying performance differences between two pulse oximetry systems in simulated critical neonatal conditions.
Objective: Pulse oximetry is used to guide critical clinical decisions in neonatology. We used a vital signs simulator to compare performance of two pulse oximetry systems in conditions not tested in standardized clinical verification studies.
Study design: We devised a set of simulated tissue translucency, perfusion, peripheral oxygen saturation (SpO2), and heart rate (HR) parameters to mimic challenging real-world neonatal data and applied them to two marketed pulse oximetry systems (Nellcor™ and Masimo®). At each combination of input parameters, we used the response from both systems to assess SpO2 error.
Results: The mean SpO2 error for Nellcor™ was below 1.1% across all parameters explored, while Masimo® showed significantly higher (p < 0.005) error at lower translucencies.
Conclusion: Significant performance differences can be observed when comparing pulse oximeters at low translucency and perfusion conditions. Patient simulators cannot replace clinical testing but provide a safe and cost-effective method for additional performance profiling.
期刊介绍:
The Journal of Perinatology provides members of the perinatal/neonatal healthcare team with original information pertinent to improving maternal/fetal and neonatal care. We publish peer-reviewed clinical research articles, state-of-the art reviews, comments, quality improvement reports, and letters to the editor. Articles published in the Journal of Perinatology embrace the full scope of the specialty, including clinical, professional, political, administrative and educational aspects. The Journal also explores legal and ethical issues, neonatal technology and product development.
The Journal’s audience includes all those that participate in perinatal/neonatal care, including, but not limited to neonatologists, perinatologists, perinatal epidemiologists, pediatricians and pediatric subspecialists, surgeons, neonatal and perinatal nurses, respiratory therapists, pharmacists, social workers, dieticians, speech and hearing experts, other allied health professionals, as well as subspecialists who participate in patient care including radiologists, laboratory medicine and pathologists.