{"title":"新生儿黄疸经皮胆红素测量技术及其与血清胆红素水平一致性的比较分析:一项前瞻性研究","authors":"Merve Kucukoglu Keser, Evrim Alyamac Dizdar, Esra Beser, Esin Okman, Fatma Nur Sari","doi":"10.1155/ijcp/1688130","DOIUrl":null,"url":null,"abstract":"<div>\n <p><b>Aim:</b> This prospective study aimed to assess the accuracy and agreement of two transcutaneous bilirubin (TcB) devices—BiliCare and MBJ20—compared with serum total bilirubin (STB) in neonates with jaundice.</p>\n <p><b>Methods:</b> Eighty-eight infants born at ≥ 34 weeks of gestation were included. TcB was measured from the auricle using BiliCare and from the sternum using MBJ20, concurrently with STB testing.</p>\n <p><b>Results:</b> The median gestational age was 38 weeks and mean birth weight was 3108 g; 39% were delivered by cesarean section and 49% were male. Mean bilirubin levels were 14.6 mg/dL (MBJ20), 12.9 mg/dL (BiliCare), and 14.4 mg/dL (STB). Both devices showed strong correlation with STB (MBJ20: <i>r</i> = 0.756, BiliCare: <i>r</i> = 0.772; <i>p</i> < 0.001). Mean differences from STB were −0.12 mg/dL for MBJ20 and 1.54 mg/dL for BiliCare. BiliCare showed higher predictive accuracy (AUC = 0.814) than MBJ20 (AUC = 0.751). Optimal TcB cutoff values were 14.2 mg/dL (BiliCare) and 15.6 mg/dL (MBJ20). Regression analysis showed no significant bias for either device. Based on STB measurements, 15% of the infants required phototherapy (PT).</p>\n <p><b>Conclusion:</b> Both devices provide reliable noninvasive screening though BiliCare has a better accuracy in predicting the need for PT.</p>\n </div>","PeriodicalId":13782,"journal":{"name":"International Journal of Clinical Practice","volume":"2025 1","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2025-07-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1155/ijcp/1688130","citationCount":"0","resultStr":"{\"title\":\"Comparative Analysis of Transcutaneous Bilirubin Measurement Techniques and Their Agreement With Serum Bilirubin Levels in Neonatal Jaundice: A Prospective Study\",\"authors\":\"Merve Kucukoglu Keser, Evrim Alyamac Dizdar, Esra Beser, Esin Okman, Fatma Nur Sari\",\"doi\":\"10.1155/ijcp/1688130\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n <p><b>Aim:</b> This prospective study aimed to assess the accuracy and agreement of two transcutaneous bilirubin (TcB) devices—BiliCare and MBJ20—compared with serum total bilirubin (STB) in neonates with jaundice.</p>\\n <p><b>Methods:</b> Eighty-eight infants born at ≥ 34 weeks of gestation were included. TcB was measured from the auricle using BiliCare and from the sternum using MBJ20, concurrently with STB testing.</p>\\n <p><b>Results:</b> The median gestational age was 38 weeks and mean birth weight was 3108 g; 39% were delivered by cesarean section and 49% were male. Mean bilirubin levels were 14.6 mg/dL (MBJ20), 12.9 mg/dL (BiliCare), and 14.4 mg/dL (STB). Both devices showed strong correlation with STB (MBJ20: <i>r</i> = 0.756, BiliCare: <i>r</i> = 0.772; <i>p</i> < 0.001). Mean differences from STB were −0.12 mg/dL for MBJ20 and 1.54 mg/dL for BiliCare. BiliCare showed higher predictive accuracy (AUC = 0.814) than MBJ20 (AUC = 0.751). Optimal TcB cutoff values were 14.2 mg/dL (BiliCare) and 15.6 mg/dL (MBJ20). Regression analysis showed no significant bias for either device. Based on STB measurements, 15% of the infants required phototherapy (PT).</p>\\n <p><b>Conclusion:</b> Both devices provide reliable noninvasive screening though BiliCare has a better accuracy in predicting the need for PT.</p>\\n </div>\",\"PeriodicalId\":13782,\"journal\":{\"name\":\"International Journal of Clinical Practice\",\"volume\":\"2025 1\",\"pages\":\"\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2025-07-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1155/ijcp/1688130\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Clinical Practice\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1155/ijcp/1688130\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICINE, GENERAL & INTERNAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Clinical Practice","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1155/ijcp/1688130","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
Comparative Analysis of Transcutaneous Bilirubin Measurement Techniques and Their Agreement With Serum Bilirubin Levels in Neonatal Jaundice: A Prospective Study
Aim: This prospective study aimed to assess the accuracy and agreement of two transcutaneous bilirubin (TcB) devices—BiliCare and MBJ20—compared with serum total bilirubin (STB) in neonates with jaundice.
Methods: Eighty-eight infants born at ≥ 34 weeks of gestation were included. TcB was measured from the auricle using BiliCare and from the sternum using MBJ20, concurrently with STB testing.
Results: The median gestational age was 38 weeks and mean birth weight was 3108 g; 39% were delivered by cesarean section and 49% were male. Mean bilirubin levels were 14.6 mg/dL (MBJ20), 12.9 mg/dL (BiliCare), and 14.4 mg/dL (STB). Both devices showed strong correlation with STB (MBJ20: r = 0.756, BiliCare: r = 0.772; p < 0.001). Mean differences from STB were −0.12 mg/dL for MBJ20 and 1.54 mg/dL for BiliCare. BiliCare showed higher predictive accuracy (AUC = 0.814) than MBJ20 (AUC = 0.751). Optimal TcB cutoff values were 14.2 mg/dL (BiliCare) and 15.6 mg/dL (MBJ20). Regression analysis showed no significant bias for either device. Based on STB measurements, 15% of the infants required phototherapy (PT).
Conclusion: Both devices provide reliable noninvasive screening though BiliCare has a better accuracy in predicting the need for PT.
期刊介绍:
IJCP is a general medical journal. IJCP gives special priority to work that has international appeal.
IJCP publishes:
Editorials. IJCP Editorials are commissioned. [Peer reviewed at the editor''s discretion]
Perspectives. Most IJCP Perspectives are commissioned. Example. [Peer reviewed at the editor''s discretion]
Study design and interpretation. Example. [Always peer reviewed]
Original data from clinical investigations. In particular: Primary research papers from RCTs, observational studies, epidemiological studies; pre-specified sub-analyses; pooled analyses. [Always peer reviewed]
Meta-analyses. [Always peer reviewed]
Systematic reviews. From October 2009, special priority will be given to systematic reviews. [Always peer reviewed]
Non-systematic/narrative reviews. From October 2009, reviews that are not systematic will be considered only if they include a discrete Methods section that must explicitly describe the authors'' approach. Special priority will, however, be given to systematic reviews. [Always peer reviewed]
''How to…'' papers. Example. [Always peer reviewed]
Consensus statements. [Always peer reviewed] Short reports. [Always peer reviewed]
Letters. [Peer reviewed at the editor''s discretion]
International scope
IJCP publishes work from investigators globally. Around 30% of IJCP articles list an author from the UK. Around 30% of IJCP articles list an author from the USA or Canada. Around 45% of IJCP articles list an author from a European country that is not the UK. Around 15% of articles published in IJCP list an author from a country in the Asia-Pacific region.