{"title":"高能x射线辐射光致发光玻璃剂量计剂量学特性的实验研究。","authors":"Hao Qi, Xiang Du, Sheng-Ri Li, Chunyong Yang, Jin Wang","doi":"10.1093/rpd/ncaf079","DOIUrl":null,"url":null,"abstract":"<p><p>This study evaluates the dosimetric performance of GD-302M radiophotoluminescent glass dosemeters (RPLGDs) under high-energy X-rays (6-15 MV) as a potential alternative to thermoluminescent dosemeters for radiotherapy audits in China. Key properties-dose linearity, uniformity, reproducibility, energy response, fading, build-up effects, and signal depletion-were systematically assessed using a medical linear accelerator under reference conditions. The results demonstrated excellent dose linearity (R2 = 0.9987, 1-4 Gy), uniformity (coefficient of variation = 0.9%), and reproducibility (standard deviation = 0.42%). Energy-dependent variations remained within 1.3%, while fading effects showed a cumulative signal loss of 3% over 110 d. Unpreheated RPLGDs retained 98% of preheated signals after 30 d, with minimal signal depletion per readout (0.016%). A water-to-solid phantom conversion coefficient of 0.998 was established. The total experimental uncertainty was 2.4%. These findings validate RPLGDs as a stable, reproducible, and cost-effective tool for enhancing radiotherapy quality assurance in resource-constrained settings.</p>","PeriodicalId":20795,"journal":{"name":"Radiation protection dosimetry","volume":" ","pages":"841-851"},"PeriodicalIF":0.7000,"publicationDate":"2025-08-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Experimental study of dosimetric properties of radiophotoluminescent glass dosemeters using high-energy X-rays.\",\"authors\":\"Hao Qi, Xiang Du, Sheng-Ri Li, Chunyong Yang, Jin Wang\",\"doi\":\"10.1093/rpd/ncaf079\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>This study evaluates the dosimetric performance of GD-302M radiophotoluminescent glass dosemeters (RPLGDs) under high-energy X-rays (6-15 MV) as a potential alternative to thermoluminescent dosemeters for radiotherapy audits in China. Key properties-dose linearity, uniformity, reproducibility, energy response, fading, build-up effects, and signal depletion-were systematically assessed using a medical linear accelerator under reference conditions. The results demonstrated excellent dose linearity (R2 = 0.9987, 1-4 Gy), uniformity (coefficient of variation = 0.9%), and reproducibility (standard deviation = 0.42%). Energy-dependent variations remained within 1.3%, while fading effects showed a cumulative signal loss of 3% over 110 d. Unpreheated RPLGDs retained 98% of preheated signals after 30 d, with minimal signal depletion per readout (0.016%). A water-to-solid phantom conversion coefficient of 0.998 was established. The total experimental uncertainty was 2.4%. These findings validate RPLGDs as a stable, reproducible, and cost-effective tool for enhancing radiotherapy quality assurance in resource-constrained settings.</p>\",\"PeriodicalId\":20795,\"journal\":{\"name\":\"Radiation protection dosimetry\",\"volume\":\" \",\"pages\":\"841-851\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2025-08-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Radiation protection dosimetry\",\"FirstCategoryId\":\"93\",\"ListUrlMain\":\"https://doi.org/10.1093/rpd/ncaf079\",\"RegionNum\":4,\"RegionCategory\":\"环境科学与生态学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENVIRONMENTAL SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation protection dosimetry","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1093/rpd/ncaf079","RegionNum":4,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENVIRONMENTAL SCIENCES","Score":null,"Total":0}
引用次数: 0

摘要

本研究评估了GD-302M放射光致发光玻璃剂量计(RPLGDs)在高能x射线(6-15 MV)下的剂量学性能,作为中国放射审计中热致发光剂量计的潜在替代品。在参考条件下,使用医用直线加速器系统地评估了关键特性-剂量线性、均匀性、再现性、能量响应、衰落、累积效应和信号耗尽。结果具有良好的剂量线性(R2 = 0.9987, 1 ~ 4 Gy)、均匀性(变异系数= 0.9%)和重复性(标准差= 0.42%)。能量依赖的变化保持在1.3%以内,而衰减效应显示110 d内的累计信号损失为3%。未预热的rplgd在30 d后保留了98%的预热信号,每次读出的信号损耗最小(0.016%)。建立了水固幻相转换系数为0.998。实验总不确定度为2.4%。这些发现证实了RPLGDs是一种稳定、可重复、具有成本效益的工具,可在资源有限的情况下提高放疗质量保证。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Experimental study of dosimetric properties of radiophotoluminescent glass dosemeters using high-energy X-rays.

This study evaluates the dosimetric performance of GD-302M radiophotoluminescent glass dosemeters (RPLGDs) under high-energy X-rays (6-15 MV) as a potential alternative to thermoluminescent dosemeters for radiotherapy audits in China. Key properties-dose linearity, uniformity, reproducibility, energy response, fading, build-up effects, and signal depletion-were systematically assessed using a medical linear accelerator under reference conditions. The results demonstrated excellent dose linearity (R2 = 0.9987, 1-4 Gy), uniformity (coefficient of variation = 0.9%), and reproducibility (standard deviation = 0.42%). Energy-dependent variations remained within 1.3%, while fading effects showed a cumulative signal loss of 3% over 110 d. Unpreheated RPLGDs retained 98% of preheated signals after 30 d, with minimal signal depletion per readout (0.016%). A water-to-solid phantom conversion coefficient of 0.998 was established. The total experimental uncertainty was 2.4%. These findings validate RPLGDs as a stable, reproducible, and cost-effective tool for enhancing radiotherapy quality assurance in resource-constrained settings.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Radiation protection dosimetry
Radiation protection dosimetry 环境科学-公共卫生、环境卫生与职业卫生
CiteScore
1.40
自引率
10.00%
发文量
223
审稿时长
6-12 weeks
期刊介绍: Radiation Protection Dosimetry covers all aspects of personal and environmental dosimetry and monitoring, for both ionising and non-ionising radiations. This includes biological aspects, physical concepts, biophysical dosimetry, external and internal personal dosimetry and monitoring, environmental and workplace monitoring, accident dosimetry, and dosimetry related to the protection of patients. Particular emphasis is placed on papers covering the fundamentals of dosimetry; units, radiation quantities and conversion factors. Papers covering archaeological dating are included only if the fundamental measurement method or technique, such as thermoluminescence, has direct application to personal dosimetry measurements. Papers covering the dosimetric aspects of radon or other naturally occurring radioactive materials and low level radiation are included. Animal experiments and ecological sample measurements are not included unless there is a significant relevant content reason.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信