{"title":"雅培Alinity i化学发光分析仪用于五项甲状腺功能检测的性能验证","authors":"Hongyu Zhang , Baixiu Wu , Liuhua Ke , Zheng Peng","doi":"10.1016/j.plabm.2025.e00487","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>To verify and evaluate the performance of the Abbott Alinity i chemiluminescence analyzer for five thyroid function tests.</div></div><div><h3>Method</h3><div>Referring to the relevant documents of the Clinical and Laboratory Standards Institute (CLSI) and related literature, the precision, accuracy, linear range, reference interval, and sample carryover effect of Abbott Alinity i immunoassay system for measuring FT3, FT4, T3, T4, and TSH were verified and analyzed.</div></div><div><h3>Results</h3><div>Precision: Repeatability ranged from 1.23 % to 6.11 %, and intermediate precision ranged from 1.84 % to 7.33 %, both meeting the quality targets (≤6.25 % and ≤8.33 %, respectively). Accuracy: The deviation between the mean value and the target value was less than 12.5 %, indicating good agreement. Linearity: For T3, T4, and TSH, the 95 % confidence intervals of the deviation from linearity (difference between measured and expected value) were entirely within the allowable deviation limits (ADL). For FT3 and FT4, manufacturer guidelines preclude dilution; thus, linearity verification was omitted. Reference interval: All test values of 20 healthy individuals were within the reference intervals provided by the manufacturer. Sample carryover effect: The sample carryover effect ranged from −0.4 % to 0.17 %, which met the requirement of less than 1 %.</div></div><div><h3>Conclusion</h3><div>The Abbott Alinity i analyzer demonstrated acceptable performance in precision, accuracy, linearity (for T3, T4, and TSH), reference interval, and carry-over effect for the five thyroid function tests, meeting manufacturer specifications.</div></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"46 ","pages":"Article e00487"},"PeriodicalIF":1.3000,"publicationDate":"2025-06-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance validation of Abbott Alinity i chemiluminescence analyzer for five thyroid function tests\",\"authors\":\"Hongyu Zhang , Baixiu Wu , Liuhua Ke , Zheng Peng\",\"doi\":\"10.1016/j.plabm.2025.e00487\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objective</h3><div>To verify and evaluate the performance of the Abbott Alinity i chemiluminescence analyzer for five thyroid function tests.</div></div><div><h3>Method</h3><div>Referring to the relevant documents of the Clinical and Laboratory Standards Institute (CLSI) and related literature, the precision, accuracy, linear range, reference interval, and sample carryover effect of Abbott Alinity i immunoassay system for measuring FT3, FT4, T3, T4, and TSH were verified and analyzed.</div></div><div><h3>Results</h3><div>Precision: Repeatability ranged from 1.23 % to 6.11 %, and intermediate precision ranged from 1.84 % to 7.33 %, both meeting the quality targets (≤6.25 % and ≤8.33 %, respectively). Accuracy: The deviation between the mean value and the target value was less than 12.5 %, indicating good agreement. Linearity: For T3, T4, and TSH, the 95 % confidence intervals of the deviation from linearity (difference between measured and expected value) were entirely within the allowable deviation limits (ADL). For FT3 and FT4, manufacturer guidelines preclude dilution; thus, linearity verification was omitted. Reference interval: All test values of 20 healthy individuals were within the reference intervals provided by the manufacturer. Sample carryover effect: The sample carryover effect ranged from −0.4 % to 0.17 %, which met the requirement of less than 1 %.</div></div><div><h3>Conclusion</h3><div>The Abbott Alinity i analyzer demonstrated acceptable performance in precision, accuracy, linearity (for T3, T4, and TSH), reference interval, and carry-over effect for the five thyroid function tests, meeting manufacturer specifications.</div></div>\",\"PeriodicalId\":20421,\"journal\":{\"name\":\"Practical Laboratory Medicine\",\"volume\":\"46 \",\"pages\":\"Article e00487\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2025-06-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Practical Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S235255172500040X\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S235255172500040X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
Performance validation of Abbott Alinity i chemiluminescence analyzer for five thyroid function tests
Objective
To verify and evaluate the performance of the Abbott Alinity i chemiluminescence analyzer for five thyroid function tests.
Method
Referring to the relevant documents of the Clinical and Laboratory Standards Institute (CLSI) and related literature, the precision, accuracy, linear range, reference interval, and sample carryover effect of Abbott Alinity i immunoassay system for measuring FT3, FT4, T3, T4, and TSH were verified and analyzed.
Results
Precision: Repeatability ranged from 1.23 % to 6.11 %, and intermediate precision ranged from 1.84 % to 7.33 %, both meeting the quality targets (≤6.25 % and ≤8.33 %, respectively). Accuracy: The deviation between the mean value and the target value was less than 12.5 %, indicating good agreement. Linearity: For T3, T4, and TSH, the 95 % confidence intervals of the deviation from linearity (difference between measured and expected value) were entirely within the allowable deviation limits (ADL). For FT3 and FT4, manufacturer guidelines preclude dilution; thus, linearity verification was omitted. Reference interval: All test values of 20 healthy individuals were within the reference intervals provided by the manufacturer. Sample carryover effect: The sample carryover effect ranged from −0.4 % to 0.17 %, which met the requirement of less than 1 %.
Conclusion
The Abbott Alinity i analyzer demonstrated acceptable performance in precision, accuracy, linearity (for T3, T4, and TSH), reference interval, and carry-over effect for the five thyroid function tests, meeting manufacturer specifications.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.