雅培Alinity i化学发光分析仪用于五项甲状腺功能检测的性能验证

IF 1.3 Q3 MEDICAL LABORATORY TECHNOLOGY
Hongyu Zhang , Baixiu Wu , Liuhua Ke , Zheng Peng
{"title":"雅培Alinity i化学发光分析仪用于五项甲状腺功能检测的性能验证","authors":"Hongyu Zhang ,&nbsp;Baixiu Wu ,&nbsp;Liuhua Ke ,&nbsp;Zheng Peng","doi":"10.1016/j.plabm.2025.e00487","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>To verify and evaluate the performance of the Abbott Alinity i chemiluminescence analyzer for five thyroid function tests.</div></div><div><h3>Method</h3><div>Referring to the relevant documents of the Clinical and Laboratory Standards Institute (CLSI) and related literature, the precision, accuracy, linear range, reference interval, and sample carryover effect of Abbott Alinity i immunoassay system for measuring FT3, FT4, T3, T4, and TSH were verified and analyzed.</div></div><div><h3>Results</h3><div>Precision: Repeatability ranged from 1.23 % to 6.11 %, and intermediate precision ranged from 1.84 % to 7.33 %, both meeting the quality targets (≤6.25 % and ≤8.33 %, respectively). Accuracy: The deviation between the mean value and the target value was less than 12.5 %, indicating good agreement. Linearity: For T3, T4, and TSH, the 95 % confidence intervals of the deviation from linearity (difference between measured and expected value) were entirely within the allowable deviation limits (ADL). For FT3 and FT4, manufacturer guidelines preclude dilution; thus, linearity verification was omitted. Reference interval: All test values of 20 healthy individuals were within the reference intervals provided by the manufacturer. Sample carryover effect: The sample carryover effect ranged from −0.4 % to 0.17 %, which met the requirement of less than 1 %.</div></div><div><h3>Conclusion</h3><div>The Abbott Alinity i analyzer demonstrated acceptable performance in precision, accuracy, linearity (for T3, T4, and TSH), reference interval, and carry-over effect for the five thyroid function tests, meeting manufacturer specifications.</div></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"46 ","pages":"Article e00487"},"PeriodicalIF":1.3000,"publicationDate":"2025-06-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance validation of Abbott Alinity i chemiluminescence analyzer for five thyroid function tests\",\"authors\":\"Hongyu Zhang ,&nbsp;Baixiu Wu ,&nbsp;Liuhua Ke ,&nbsp;Zheng Peng\",\"doi\":\"10.1016/j.plabm.2025.e00487\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objective</h3><div>To verify and evaluate the performance of the Abbott Alinity i chemiluminescence analyzer for five thyroid function tests.</div></div><div><h3>Method</h3><div>Referring to the relevant documents of the Clinical and Laboratory Standards Institute (CLSI) and related literature, the precision, accuracy, linear range, reference interval, and sample carryover effect of Abbott Alinity i immunoassay system for measuring FT3, FT4, T3, T4, and TSH were verified and analyzed.</div></div><div><h3>Results</h3><div>Precision: Repeatability ranged from 1.23 % to 6.11 %, and intermediate precision ranged from 1.84 % to 7.33 %, both meeting the quality targets (≤6.25 % and ≤8.33 %, respectively). Accuracy: The deviation between the mean value and the target value was less than 12.5 %, indicating good agreement. Linearity: For T3, T4, and TSH, the 95 % confidence intervals of the deviation from linearity (difference between measured and expected value) were entirely within the allowable deviation limits (ADL). For FT3 and FT4, manufacturer guidelines preclude dilution; thus, linearity verification was omitted. Reference interval: All test values of 20 healthy individuals were within the reference intervals provided by the manufacturer. Sample carryover effect: The sample carryover effect ranged from −0.4 % to 0.17 %, which met the requirement of less than 1 %.</div></div><div><h3>Conclusion</h3><div>The Abbott Alinity i analyzer demonstrated acceptable performance in precision, accuracy, linearity (for T3, T4, and TSH), reference interval, and carry-over effect for the five thyroid function tests, meeting manufacturer specifications.</div></div>\",\"PeriodicalId\":20421,\"journal\":{\"name\":\"Practical Laboratory Medicine\",\"volume\":\"46 \",\"pages\":\"Article e00487\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2025-06-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Practical Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S235255172500040X\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S235255172500040X","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的验证和评价雅培Alinity i化学发光分析仪在甲状腺功能五项检测中的性能。方法参考美国临床与实验室标准协会(CLSI)的相关文件及相关文献,对雅培Alinity i免疫测定系统测量FT3、FT4、T3、T4、TSH的精密度、准确度、线性范围、参考区间及样本携带效应进行验证和分析。结果精密度:重复性为1.23% ~ 6.11%,中间精密度为1.84% ~ 7.33%,均满足质量指标(分别≤6.25%和≤8.33%)。准确度:平均值与目标值偏差小于12.5%,吻合良好。线性:对于T3、T4和TSH,偏离线性(实测值与期望值之差)的95%置信区间完全在允许偏差限(ADL)内。对于FT3和FT4,制造商指南禁止稀释;因此,线性验证被省略。参考区间:20名健康个体的所有测试值均在制造商提供的参考区间内。样品残留效应:样品残留效应范围为- 0.4% ~ 0.17%,满足小于1%的要求。结论雅培Alinity i分析仪在精密度、准确度、线性度(T3、T4和TSH)、参考区间和五项甲状腺功能检测的结转效应方面表现良好,符合生产厂家的要求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Performance validation of Abbott Alinity i chemiluminescence analyzer for five thyroid function tests

Objective

To verify and evaluate the performance of the Abbott Alinity i chemiluminescence analyzer for five thyroid function tests.

Method

Referring to the relevant documents of the Clinical and Laboratory Standards Institute (CLSI) and related literature, the precision, accuracy, linear range, reference interval, and sample carryover effect of Abbott Alinity i immunoassay system for measuring FT3, FT4, T3, T4, and TSH were verified and analyzed.

Results

Precision: Repeatability ranged from 1.23 % to 6.11 %, and intermediate precision ranged from 1.84 % to 7.33 %, both meeting the quality targets (≤6.25 % and ≤8.33 %, respectively). Accuracy: The deviation between the mean value and the target value was less than 12.5 %, indicating good agreement. Linearity: For T3, T4, and TSH, the 95 % confidence intervals of the deviation from linearity (difference between measured and expected value) were entirely within the allowable deviation limits (ADL). For FT3 and FT4, manufacturer guidelines preclude dilution; thus, linearity verification was omitted. Reference interval: All test values of 20 healthy individuals were within the reference intervals provided by the manufacturer. Sample carryover effect: The sample carryover effect ranged from −0.4 % to 0.17 %, which met the requirement of less than 1 %.

Conclusion

The Abbott Alinity i analyzer demonstrated acceptable performance in precision, accuracy, linearity (for T3, T4, and TSH), reference interval, and carry-over effect for the five thyroid function tests, meeting manufacturer specifications.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Practical Laboratory Medicine
Practical Laboratory Medicine Health Professions-Radiological and Ultrasound Technology
CiteScore
3.50
自引率
0.00%
发文量
40
审稿时长
7 weeks
期刊介绍: Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信