Hao Tang, Yixin Zhou, Baojun Mai, Binjie Zhu, Ping Chen, Yujia Fu, Guangzhi Wang
{"title":"基于惯性测量单元的智能试验系统在全髋关节置换术中的部件方位测量。","authors":"Hao Tang, Yixin Zhou, Baojun Mai, Binjie Zhu, Ping Chen, Yujia Fu, Guangzhi Wang","doi":"10.1186/s42836-025-00312-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Intraoperative measurement of component orientation represents a basis for optimizing outcomes after total hip arthroplasty (THA). Although the use of computer navigation systems in THA has improved the accuracy of component positioning, they have not gained widespread popularity due to their complexity, time demands, and time-consuming protocols.</p><p><strong>Methods: </strong>We developed an Inertial Measurement Unit-based Hip Smart Trial system (IMUHST) to assist with intra-operative monitoring of hip posture. An in vitro validation experiment was conducted using a sawbones with a three-dimensional (3D) measurement model as the reference standard.</p><p><strong>Results: </strong>The absolute mean error, Bland-Altman analysis, and Intra-class Correlation Coefficient revealed that the accuracy and precision of this system meet the threshold for clinical application.</p><p><strong>Conclusions: </strong>In conclusion, this in vitro validation demonstrates that the IMUHST system provides accurate component orientation measurements while eliminating the cost and complexity of optical navigation, offering a practical solution for widespread adoption. Video Abstract.</p>","PeriodicalId":52831,"journal":{"name":"Arthroplasty","volume":"7 1","pages":"36"},"PeriodicalIF":2.3000,"publicationDate":"2025-06-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12188666/pdf/","citationCount":"0","resultStr":"{\"title\":\"Component orientation measurements in total hip arthroplasty using an inertial measurement unit-based smart trial system.\",\"authors\":\"Hao Tang, Yixin Zhou, Baojun Mai, Binjie Zhu, Ping Chen, Yujia Fu, Guangzhi Wang\",\"doi\":\"10.1186/s42836-025-00312-3\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Intraoperative measurement of component orientation represents a basis for optimizing outcomes after total hip arthroplasty (THA). Although the use of computer navigation systems in THA has improved the accuracy of component positioning, they have not gained widespread popularity due to their complexity, time demands, and time-consuming protocols.</p><p><strong>Methods: </strong>We developed an Inertial Measurement Unit-based Hip Smart Trial system (IMUHST) to assist with intra-operative monitoring of hip posture. An in vitro validation experiment was conducted using a sawbones with a three-dimensional (3D) measurement model as the reference standard.</p><p><strong>Results: </strong>The absolute mean error, Bland-Altman analysis, and Intra-class Correlation Coefficient revealed that the accuracy and precision of this system meet the threshold for clinical application.</p><p><strong>Conclusions: </strong>In conclusion, this in vitro validation demonstrates that the IMUHST system provides accurate component orientation measurements while eliminating the cost and complexity of optical navigation, offering a practical solution for widespread adoption. Video Abstract.</p>\",\"PeriodicalId\":52831,\"journal\":{\"name\":\"Arthroplasty\",\"volume\":\"7 1\",\"pages\":\"36\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2025-06-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12188666/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Arthroplasty\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1186/s42836-025-00312-3\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ORTHOPEDICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arthroplasty","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s42836-025-00312-3","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
Component orientation measurements in total hip arthroplasty using an inertial measurement unit-based smart trial system.
Background: Intraoperative measurement of component orientation represents a basis for optimizing outcomes after total hip arthroplasty (THA). Although the use of computer navigation systems in THA has improved the accuracy of component positioning, they have not gained widespread popularity due to their complexity, time demands, and time-consuming protocols.
Methods: We developed an Inertial Measurement Unit-based Hip Smart Trial system (IMUHST) to assist with intra-operative monitoring of hip posture. An in vitro validation experiment was conducted using a sawbones with a three-dimensional (3D) measurement model as the reference standard.
Results: The absolute mean error, Bland-Altman analysis, and Intra-class Correlation Coefficient revealed that the accuracy and precision of this system meet the threshold for clinical application.
Conclusions: In conclusion, this in vitro validation demonstrates that the IMUHST system provides accurate component orientation measurements while eliminating the cost and complexity of optical navigation, offering a practical solution for widespread adoption. Video Abstract.