{"title":"RS-1光谱域光学相干层析成像与OA-2000生物计测量轴向长度的比较。","authors":"Hitomi Fujihara, Shinji Kakihara, Takao Hirano, Toshinori Murata","doi":"10.18240/ijo.2025.06.15","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>To determine the utility of the RS-1 spectral domain optical coherence tomography (SD-OCT), which incorporates an \"OCT Analysis Correction Parameter\" for approximating axial length, in comparison with measurements obtained from the OA-2000.</p><p><strong>Methods: </strong>Twenty-five right eyes of healthy individuals were included. Two horizontal line scans were conducted using the RS-1, and OCT Analysis Correction Parameters were recorded. Axial length was measured twice per eye using the OA-2000. Correlation between devices was assessed using Pearson correlation coefficient, coefficient of repeatability (CR%), and Bland-Altman analysis.</p><p><strong>Results: </strong>High correlation was found between axial length measurements from RS-1 and OA-2000 (<i>r</i>=0.986, <i>P</i><0.0001). The coefficient of repeatability was 1.56% for RS-1 and 0.115% for OA-2000. Mean axial length was 25.12±1.38 mm for RS-1 and 24.90±1.54 mm for OA-2000, with RS-1 showing a statistically significant larger value (paired <i>t</i>-test, <i>P</i>=0.0009).</p><p><strong>Conclusion: </strong>The axial length measurements of RS-1 demonstrate a strong positive correlation with those from the OA-2000. This indicates that it could potentially be used to forecast scan ranges and choose suitable databases according to axial length categories.</p>","PeriodicalId":14312,"journal":{"name":"International journal of ophthalmology","volume":"18 6","pages":"1085-1089"},"PeriodicalIF":1.9000,"publicationDate":"2025-06-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12120460/pdf/","citationCount":"0","resultStr":"{\"title\":\"Comparison of axial length measurements with RS-1 spectral domain optical coherence tomography and OA-2000 biometer.\",\"authors\":\"Hitomi Fujihara, Shinji Kakihara, Takao Hirano, Toshinori Murata\",\"doi\":\"10.18240/ijo.2025.06.15\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Aim: </strong>To determine the utility of the RS-1 spectral domain optical coherence tomography (SD-OCT), which incorporates an \\\"OCT Analysis Correction Parameter\\\" for approximating axial length, in comparison with measurements obtained from the OA-2000.</p><p><strong>Methods: </strong>Twenty-five right eyes of healthy individuals were included. Two horizontal line scans were conducted using the RS-1, and OCT Analysis Correction Parameters were recorded. Axial length was measured twice per eye using the OA-2000. Correlation between devices was assessed using Pearson correlation coefficient, coefficient of repeatability (CR%), and Bland-Altman analysis.</p><p><strong>Results: </strong>High correlation was found between axial length measurements from RS-1 and OA-2000 (<i>r</i>=0.986, <i>P</i><0.0001). The coefficient of repeatability was 1.56% for RS-1 and 0.115% for OA-2000. Mean axial length was 25.12±1.38 mm for RS-1 and 24.90±1.54 mm for OA-2000, with RS-1 showing a statistically significant larger value (paired <i>t</i>-test, <i>P</i>=0.0009).</p><p><strong>Conclusion: </strong>The axial length measurements of RS-1 demonstrate a strong positive correlation with those from the OA-2000. This indicates that it could potentially be used to forecast scan ranges and choose suitable databases according to axial length categories.</p>\",\"PeriodicalId\":14312,\"journal\":{\"name\":\"International journal of ophthalmology\",\"volume\":\"18 6\",\"pages\":\"1085-1089\"},\"PeriodicalIF\":1.9000,\"publicationDate\":\"2025-06-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12120460/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International journal of ophthalmology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.18240/ijo.2025.06.15\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q2\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.18240/ijo.2025.06.15","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
Comparison of axial length measurements with RS-1 spectral domain optical coherence tomography and OA-2000 biometer.
Aim: To determine the utility of the RS-1 spectral domain optical coherence tomography (SD-OCT), which incorporates an "OCT Analysis Correction Parameter" for approximating axial length, in comparison with measurements obtained from the OA-2000.
Methods: Twenty-five right eyes of healthy individuals were included. Two horizontal line scans were conducted using the RS-1, and OCT Analysis Correction Parameters were recorded. Axial length was measured twice per eye using the OA-2000. Correlation between devices was assessed using Pearson correlation coefficient, coefficient of repeatability (CR%), and Bland-Altman analysis.
Results: High correlation was found between axial length measurements from RS-1 and OA-2000 (r=0.986, P<0.0001). The coefficient of repeatability was 1.56% for RS-1 and 0.115% for OA-2000. Mean axial length was 25.12±1.38 mm for RS-1 and 24.90±1.54 mm for OA-2000, with RS-1 showing a statistically significant larger value (paired t-test, P=0.0009).
Conclusion: The axial length measurements of RS-1 demonstrate a strong positive correlation with those from the OA-2000. This indicates that it could potentially be used to forecast scan ranges and choose suitable databases according to axial length categories.
期刊介绍:
· International Journal of Ophthalmology-IJO (English edition) is a global ophthalmological scientific publication
and a peer-reviewed open access periodical (ISSN 2222-3959 print, ISSN 2227-4898 online).
This journal is sponsored by Chinese Medical Association Xi’an Branch and obtains guidance and support from
WHO and ICO (International Council of Ophthalmology). It has been indexed in SCIE, PubMed,
PubMed-Central, Chemical Abstracts, Scopus, EMBASE , and DOAJ. IJO JCR IF in 2017 is 1.166.
IJO was established in 2008, with editorial office in Xi’an, China. It is a monthly publication. General Scientific
Advisors include Prof. Hugh Taylor (President of ICO); Prof.Bruce Spivey (Immediate Past President of ICO);
Prof.Mark Tso (Ex-Vice President of ICO) and Prof.Daiming Fan (Academician and Vice President,
Chinese Academy of Engineering.
International Scientific Advisors include Prof. Serge Resnikoff (WHO Senior Speciatist for Prevention of
blindness), Prof. Chi-Chao Chan (National Eye Institute, USA) and Prof. Richard L Abbott (Ex-President of
AAO/PAAO) et al.
Honorary Editors-in-Chief: Prof. Li-Xin Xie(Academician of Chinese Academy of
Engineering/Honorary President of Chinese Ophthalmological Society); Prof. Dennis Lam (President of APAO) and
Prof. Xiao-Xin Li (Ex-President of Chinese Ophthalmological Society).
Chief Editor: Prof. Xiu-Wen Hu (President of IJO Press).
Editors-in-Chief: Prof. Yan-Nian Hui (Ex-Director, Eye Institute of Chinese PLA) and
Prof. George Chiou (Founding chief editor of Journal of Ocular Pharmacology & Therapeutics).
Associate Editors-in-Chief include:
Prof. Ning-Li Wang (President Elect of APAO);
Prof. Ke Yao (President of Chinese Ophthalmological Society) ;
Prof.William Smiddy (Bascom Palmer Eye instituteUSA) ;
Prof.Joel Schuman (President of Association of University Professors of Ophthalmology,USA);
Prof.Yizhi Liu (Vice President of Chinese Ophtlalmology Society);
Prof.Yu-Sheng Wang (Director of Eye Institute of Chinese PLA);
Prof.Ling-Yun Cheng (Director of Ocular Pharmacology, Shiley Eye Center, USA).
IJO accepts contributions in English from all over the world. It includes mainly original articles and review articles,
both basic and clinical papers.
Instruction is Welcome Contribution is Welcome Citation is Welcome
Cooperation organization
International Council of Ophthalmology(ICO), PubMed, PMC, American Academy of Ophthalmology, Asia-Pacific, Thomson Reuters, The Charlesworth Group, Crossref,Scopus,Publons, DOAJ etc.