新型阻尼容量分析装置在临床实践中评估种植体稳定性的比较分析。

IF 3.2 4区 医学 Q2 DENTISTRY, ORAL SURGERY & MEDICINE
Su-Bin Hong, Bo-Ah Lee, Woo-Ri Lee, Jung-Seok Lee, Young-Taek Kim
{"title":"新型阻尼容量分析装置在临床实践中评估种植体稳定性的比较分析。","authors":"Su-Bin Hong, Bo-Ah Lee, Woo-Ri Lee, Jung-Seok Lee, Young-Taek Kim","doi":"10.5051/jpis.2404680234","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Implant stability testing is crucial for verifying osseointegration before prosthetic loading. Several methods have been developed to assess osseointegration. Among these, damping capacity analysis (DCA) devices offer a user-friendly and non-invasive approach. In this study, we evaluated the accuracy and reliability of newly available DCA devices in assessing dental implant stability.</p><p><strong>Methods: </strong>This study included 58 implants from 37 patients over a 1-month period. Three measurements per implant were obtained with healing abutments in place, following the manufacturers' guidelines, using the DCA-P (Periotest M, Medizintechnik Gulden), DCA-A (Anycheck, Neobiotech), and DCA-T (The Trust, Dentium) devices. Factors such as healing abutment height, time since placement, bone grafting, fixture diameter, and fixture length were evaluated. Accuracy was assessed using DCA-P as the reference, and reproducibility was statistically analyzed using 3 measurements per implant. Statistical analyses were performed with SPSS Statistics 23.0 (IBM Corp.).</p><p><strong>Results: </strong>All implants that passed the stability tests using DCA devices were restored with definitive prostheses and showed no signs of early failure. The DCA-A value demonstrated a very strong correlation with the DCA-P value (DCAV-P), whereas the DCA-T value exhibited only a moderate correlation with DCAV-P. DCA-P also showed the highest reliability, followed by DCA-A and then DCA-T. The reliability of DCA-A and DCA-T was not significantly affected by any of the assessed factors; in contrast, DCA-P's reliability was significantly influenced by arch location and specific quadrant position.</p><p><strong>Conclusions: </strong>With DCA-P as the reference, DCA-A demonstrated superior accuracy compared to DCA-T. Although DCA-P exhibited the highest reliability, its performance was significantly affected by the positional factors of the target implant.</p>","PeriodicalId":48795,"journal":{"name":"Journal of Periodontal and Implant Science","volume":" ","pages":""},"PeriodicalIF":3.2000,"publicationDate":"2025-04-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparative analysis of novel damping capacity analysis devices for assessing implant stability in clinical practice.\",\"authors\":\"Su-Bin Hong, Bo-Ah Lee, Woo-Ri Lee, Jung-Seok Lee, Young-Taek Kim\",\"doi\":\"10.5051/jpis.2404680234\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Implant stability testing is crucial for verifying osseointegration before prosthetic loading. Several methods have been developed to assess osseointegration. Among these, damping capacity analysis (DCA) devices offer a user-friendly and non-invasive approach. In this study, we evaluated the accuracy and reliability of newly available DCA devices in assessing dental implant stability.</p><p><strong>Methods: </strong>This study included 58 implants from 37 patients over a 1-month period. Three measurements per implant were obtained with healing abutments in place, following the manufacturers' guidelines, using the DCA-P (Periotest M, Medizintechnik Gulden), DCA-A (Anycheck, Neobiotech), and DCA-T (The Trust, Dentium) devices. Factors such as healing abutment height, time since placement, bone grafting, fixture diameter, and fixture length were evaluated. Accuracy was assessed using DCA-P as the reference, and reproducibility was statistically analyzed using 3 measurements per implant. Statistical analyses were performed with SPSS Statistics 23.0 (IBM Corp.).</p><p><strong>Results: </strong>All implants that passed the stability tests using DCA devices were restored with definitive prostheses and showed no signs of early failure. The DCA-A value demonstrated a very strong correlation with the DCA-P value (DCAV-P), whereas the DCA-T value exhibited only a moderate correlation with DCAV-P. DCA-P also showed the highest reliability, followed by DCA-A and then DCA-T. The reliability of DCA-A and DCA-T was not significantly affected by any of the assessed factors; in contrast, DCA-P's reliability was significantly influenced by arch location and specific quadrant position.</p><p><strong>Conclusions: </strong>With DCA-P as the reference, DCA-A demonstrated superior accuracy compared to DCA-T. Although DCA-P exhibited the highest reliability, its performance was significantly affected by the positional factors of the target implant.</p>\",\"PeriodicalId\":48795,\"journal\":{\"name\":\"Journal of Periodontal and Implant Science\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":3.2000,\"publicationDate\":\"2025-04-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Periodontal and Implant Science\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.5051/jpis.2404680234\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Periodontal and Implant Science","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5051/jpis.2404680234","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

目的:假体稳定性测试是假体装载前验证骨整合的关键。已经开发了几种评估骨整合的方法。其中,阻尼能力分析(DCA)设备提供了一种用户友好和非侵入性的方法。在这项研究中,我们评估了新出现的DCA装置在评估种植体稳定性方面的准确性和可靠性。方法:本研究包括37例患者的58个种植体,为期1个月。使用DCA-P (Periotest M, Medizintechnik Gulden)、DCA-A (Anycheck, Neobiotech)和DCA-T (the Trust, Dentium)设备,在愈合基台放置的情况下,对每个种植体进行三次测量。评估诸如愈合基台高度、放置时间、植骨、固定装置直径和固定装置长度等因素。以DCA-P作为参考评估准确性,并通过每个种植体3次测量来统计分析再现性。采用SPSS Statistics 23.0 (IBM Corp.)进行统计学分析。结果:所有使用DCA装置通过稳定性测试的种植体都用确定的假体修复,没有早期失效的迹象。DCA-A值与DCA-P值(DCAV-P)的相关性非常强,而DCA-T值与DCAV-P的相关性仅为中等。DCA-P的信度也最高,其次是DCA-A,然后是DCA-T。DCA-A和DCA-T的可靠性不受任何评估因素的显著影响;相比之下,DCA-P的可靠性受到弓位置和特定象限位置的显著影响。结论:以DCA-P为参照,与DCA-T相比,DCA-A具有更高的准确性。虽然DCA-P具有最高的可靠性,但其性能受到目标种植体位置因素的显著影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparative analysis of novel damping capacity analysis devices for assessing implant stability in clinical practice.

Purpose: Implant stability testing is crucial for verifying osseointegration before prosthetic loading. Several methods have been developed to assess osseointegration. Among these, damping capacity analysis (DCA) devices offer a user-friendly and non-invasive approach. In this study, we evaluated the accuracy and reliability of newly available DCA devices in assessing dental implant stability.

Methods: This study included 58 implants from 37 patients over a 1-month period. Three measurements per implant were obtained with healing abutments in place, following the manufacturers' guidelines, using the DCA-P (Periotest M, Medizintechnik Gulden), DCA-A (Anycheck, Neobiotech), and DCA-T (The Trust, Dentium) devices. Factors such as healing abutment height, time since placement, bone grafting, fixture diameter, and fixture length were evaluated. Accuracy was assessed using DCA-P as the reference, and reproducibility was statistically analyzed using 3 measurements per implant. Statistical analyses were performed with SPSS Statistics 23.0 (IBM Corp.).

Results: All implants that passed the stability tests using DCA devices were restored with definitive prostheses and showed no signs of early failure. The DCA-A value demonstrated a very strong correlation with the DCA-P value (DCAV-P), whereas the DCA-T value exhibited only a moderate correlation with DCAV-P. DCA-P also showed the highest reliability, followed by DCA-A and then DCA-T. The reliability of DCA-A and DCA-T was not significantly affected by any of the assessed factors; in contrast, DCA-P's reliability was significantly influenced by arch location and specific quadrant position.

Conclusions: With DCA-P as the reference, DCA-A demonstrated superior accuracy compared to DCA-T. Although DCA-P exhibited the highest reliability, its performance was significantly affected by the positional factors of the target implant.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Periodontal and Implant Science
Journal of Periodontal and Implant Science DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
3.30
自引率
5.30%
发文量
38
期刊介绍: Journal of Periodontal & Implant Science (JPIS) is a peer-reviewed and open-access journal providing up-to-date information relevant to professionalism of periodontology and dental implantology. JPIS is dedicated to global and extensive publication which includes evidence-based original articles, and fundamental reviews in order to cover a variety of interests in the field of periodontal as well as implant science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信