高剂量率近距离放疗192Ir源参考空气kerma率的测量:单个机构结果的回顾性分析

IF 1.8 3区 工程技术 Q3 CHEMISTRY, INORGANIC & NUCLEAR
Tomislav Bokulić , Mirjana Budanec , Iva Mrčela , Marin Gregov , Ante Matanić , Vera Vujasinović , Mihaela Mlinarić
{"title":"高剂量率近距离放疗192Ir源参考空气kerma率的测量:单个机构结果的回顾性分析","authors":"Tomislav Bokulić ,&nbsp;Mirjana Budanec ,&nbsp;Iva Mrčela ,&nbsp;Marin Gregov ,&nbsp;Ante Matanić ,&nbsp;Vera Vujasinović ,&nbsp;Mihaela Mlinarić","doi":"10.1016/j.apradiso.2025.111959","DOIUrl":null,"url":null,"abstract":"<div><div>This work reports on the experiences and results of a single radiotherapy centre in over twenty years of well-type ionisation chamber reference air kerma rate (<em>RAKR</em>) measurements of two models <sup>192</sup>Ir high dose rate brachytherapy (HDRBT) sources. Quality assurance (QA) programmes worldwide recommend determining the <em>RAKR</em> of sources before being used in patient treatment. The observed discrepancies between the measured <em>RAKR</em> and the manufacturer's stated <em>RAKR</em> were analysed. A systematic analysis of the data was used to calculate the uncertainty budget. Measurement procedures and conditions were adjusted during the time to warrant accurate and consistent results. The <em>RAKR</em> uncertainty budget was estimated by examining all contributions to the model equation for calculating <em>RAKR</em> from measured signals. The mean percentage difference between the measured <em>RAKR</em> and the manufacturer's certificate value of 0.32 % ± 0.97 % was obtained. An analysis of these differences revealed a potential drift in chamber sensitivity over time and a shift in results following the introduction of a new chamber. Nevertheless, all observed differences were within the 3 % tolerance level and 85 % within the 1.5 % level. The overall expanded uncertainty of the measured <em>RAKR</em> of the HDRBT <sup>192</sup>Ir source of 3.04 % (k = 2) was estimated. Independent determination of <em>RAKR</em> of an HDRBT source is crucial for accurate and reliable brachytherapy treatments. A detailed analysis of the uncertainties associated with <em>RAKR</em> measurements is indispensable for an appropriate understanding of the sources of uncertainty, means of their reduction, and any adjustments needed for clinical QA programmes.</div></div>","PeriodicalId":8096,"journal":{"name":"Applied Radiation and Isotopes","volume":"225 ","pages":"Article 111959"},"PeriodicalIF":1.8000,"publicationDate":"2025-06-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Measurements of the high dose rate brachytherapy 192Ir source reference air kerma rate: a retrospective analysis of the single institution results\",\"authors\":\"Tomislav Bokulić ,&nbsp;Mirjana Budanec ,&nbsp;Iva Mrčela ,&nbsp;Marin Gregov ,&nbsp;Ante Matanić ,&nbsp;Vera Vujasinović ,&nbsp;Mihaela Mlinarić\",\"doi\":\"10.1016/j.apradiso.2025.111959\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><div>This work reports on the experiences and results of a single radiotherapy centre in over twenty years of well-type ionisation chamber reference air kerma rate (<em>RAKR</em>) measurements of two models <sup>192</sup>Ir high dose rate brachytherapy (HDRBT) sources. Quality assurance (QA) programmes worldwide recommend determining the <em>RAKR</em> of sources before being used in patient treatment. The observed discrepancies between the measured <em>RAKR</em> and the manufacturer's stated <em>RAKR</em> were analysed. A systematic analysis of the data was used to calculate the uncertainty budget. Measurement procedures and conditions were adjusted during the time to warrant accurate and consistent results. The <em>RAKR</em> uncertainty budget was estimated by examining all contributions to the model equation for calculating <em>RAKR</em> from measured signals. The mean percentage difference between the measured <em>RAKR</em> and the manufacturer's certificate value of 0.32 % ± 0.97 % was obtained. An analysis of these differences revealed a potential drift in chamber sensitivity over time and a shift in results following the introduction of a new chamber. Nevertheless, all observed differences were within the 3 % tolerance level and 85 % within the 1.5 % level. The overall expanded uncertainty of the measured <em>RAKR</em> of the HDRBT <sup>192</sup>Ir source of 3.04 % (k = 2) was estimated. Independent determination of <em>RAKR</em> of an HDRBT source is crucial for accurate and reliable brachytherapy treatments. A detailed analysis of the uncertainties associated with <em>RAKR</em> measurements is indispensable for an appropriate understanding of the sources of uncertainty, means of their reduction, and any adjustments needed for clinical QA programmes.</div></div>\",\"PeriodicalId\":8096,\"journal\":{\"name\":\"Applied Radiation and Isotopes\",\"volume\":\"225 \",\"pages\":\"Article 111959\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2025-06-06\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Applied Radiation and Isotopes\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0969804325003045\",\"RegionNum\":3,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CHEMISTRY, INORGANIC & NUCLEAR\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Applied Radiation and Isotopes","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0969804325003045","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, INORGANIC & NUCLEAR","Score":null,"Total":0}
引用次数: 0

摘要

本工作报告了单个放射治疗中心在20多年中对两种型号192Ir高剂量率近距离放射治疗(HDRBT)源进行井型电离室参考空气克氏率(RAKR)测量的经验和结果。世界各地的质量保证(QA)规划建议在用于患者治疗之前确定来源的RAKR。观察到的差异之间的测量RAKR和制造商的声明RAKR进行了分析。通过对数据的系统分析,计算了不确定性预算。测量程序和条件在此期间进行了调整,以保证准确和一致的结果。RAKR不确定性预算是通过检查从测量信号计算RAKR的模型方程的所有贡献来估计的。测量的RAKR与制造商证书值之间的平均百分比差为0.32%±0.97%。对这些差异的分析揭示了室灵敏度随时间的潜在漂移和引入新室后结果的变化。然而,所有观察到的差异都在3%的耐受性水平内,85%的差异在1.5%的耐受性水平内。HDRBT 192Ir源的测量RAKR的总体扩展不确定度估计为3.04% (k = 2)。独立测定HDRBT源的RAKR对于准确可靠的近距离治疗至关重要。对与RAKR测量相关的不确定度的详细分析对于适当理解不确定度的来源、减少不确定度的方法以及临床QA计划所需的任何调整是必不可少的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Measurements of the high dose rate brachytherapy 192Ir source reference air kerma rate: a retrospective analysis of the single institution results
This work reports on the experiences and results of a single radiotherapy centre in over twenty years of well-type ionisation chamber reference air kerma rate (RAKR) measurements of two models 192Ir high dose rate brachytherapy (HDRBT) sources. Quality assurance (QA) programmes worldwide recommend determining the RAKR of sources before being used in patient treatment. The observed discrepancies between the measured RAKR and the manufacturer's stated RAKR were analysed. A systematic analysis of the data was used to calculate the uncertainty budget. Measurement procedures and conditions were adjusted during the time to warrant accurate and consistent results. The RAKR uncertainty budget was estimated by examining all contributions to the model equation for calculating RAKR from measured signals. The mean percentage difference between the measured RAKR and the manufacturer's certificate value of 0.32 % ± 0.97 % was obtained. An analysis of these differences revealed a potential drift in chamber sensitivity over time and a shift in results following the introduction of a new chamber. Nevertheless, all observed differences were within the 3 % tolerance level and 85 % within the 1.5 % level. The overall expanded uncertainty of the measured RAKR of the HDRBT 192Ir source of 3.04 % (k = 2) was estimated. Independent determination of RAKR of an HDRBT source is crucial for accurate and reliable brachytherapy treatments. A detailed analysis of the uncertainties associated with RAKR measurements is indispensable for an appropriate understanding of the sources of uncertainty, means of their reduction, and any adjustments needed for clinical QA programmes.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Applied Radiation and Isotopes
Applied Radiation and Isotopes 工程技术-核科学技术
CiteScore
3.00
自引率
12.50%
发文量
406
审稿时长
13.5 months
期刊介绍: Applied Radiation and Isotopes provides a high quality medium for the publication of substantial, original and scientific and technological papers on the development and peaceful application of nuclear, radiation and radionuclide techniques in chemistry, physics, biochemistry, biology, medicine, security, engineering and in the earth, planetary and environmental sciences, all including dosimetry. Nuclear techniques are defined in the broadest sense and both experimental and theoretical papers are welcome. They include the development and use of α- and β-particles, X-rays and γ-rays, neutrons and other nuclear particles and radiations from all sources, including radionuclides, synchrotron sources, cyclotrons and reactors and from the natural environment. The journal aims to publish papers with significance to an international audience, containing substantial novelty and scientific impact. The Editors reserve the rights to reject, with or without external review, papers that do not meet these criteria. Papers dealing with radiation processing, i.e., where radiation is used to bring about a biological, chemical or physical change in a material, should be directed to our sister journal Radiation Physics and Chemistry.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信