{"title":"用Sysmex XT-2000iV和Sysmex XN-1000V分析仪评价雄性肉鸡(Ross 308)的红细胞参数,并测定手工和仪器方法获得的血液学参考区间","authors":"Damien Legroux, Laura Kersten, Gaelle Barral, Alexane Mauras, Thierry Buronfosse, Eve Ramery","doi":"10.1111/vcp.70009","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>The presence of both nucleated RBC and thrombocytes in birds makes the determination of hematological variables difficult. This leads to time-consuming manual evaluation. New-generation analyzers are hypothesized to be both time-saving and accurate.</p>\n </section>\n \n <section>\n \n <h3> Objectives</h3>\n \n <p>The objective was to evaluate the performance of two hematology analyzers in 27-day-old broilers and to establish reference intervals (RIs) for RBC and WBC variables obtained by automated and manual methods.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Automated HCT, WBC count, and differentials obtained using Sysmex XT-2000iV and XN-1000V on 309 and 56 EDTA blood samples from Ross 308 male broilers, respectively, were compared to manual PCV, manual WBC count, and differentials from November 2021 to November 2022. RIs for automated HCT and RBC variables (from impedance channel) on one hand, and manual PCV and WBC's variables on the other hand, were established.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The Sysmex XT-2000iV was accurate for HCT and outperformed the XN-1000V but, both automated methods and manual methods cannot be considered interchangeable. Constant error, proportional error, and bias were identified and calculated for each analyzer for HCT. Both analyzers were unable to enumerate WBC. RIs were established as follows: HCT-XT (L/L) [0.27–0.36]; HCT-XN (L/L) [0.24–0.38]; manual PCV (L/L) [0.27–0.37]; manual WBC (×10<sup>9</sup>/L) [10.0–38.4]; manual differential (×10<sup>9</sup>/L) with heterophils [1.5–16.2]; lymphocytes [5.1–18.9]; monocytes [0.0–2.1]; eosinophils [0.0–2.2]; basophils [0.0–3.1].</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Both analyzers are usable for the evaluation of HCT in Ross 308 chickens and in specific conditions for the other RBC variables. Manual methods are still preferred for the evaluation of WBC variables.</p>\n </section>\n </div>","PeriodicalId":23593,"journal":{"name":"Veterinary clinical pathology","volume":"54 2","pages":"106-119"},"PeriodicalIF":1.1000,"publicationDate":"2025-06-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/vcp.70009","citationCount":"0","resultStr":"{\"title\":\"Evaluation of blood erythroid parameters in male broiler chickens (Ross 308) with the Sysmex XT-2000iV and Sysmex XN-1000V analyzers and determination of hematological reference intervals obtained with manual and instrumental methods\",\"authors\":\"Damien Legroux, Laura Kersten, Gaelle Barral, Alexane Mauras, Thierry Buronfosse, Eve Ramery\",\"doi\":\"10.1111/vcp.70009\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>The presence of both nucleated RBC and thrombocytes in birds makes the determination of hematological variables difficult. This leads to time-consuming manual evaluation. New-generation analyzers are hypothesized to be both time-saving and accurate.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Objectives</h3>\\n \\n <p>The objective was to evaluate the performance of two hematology analyzers in 27-day-old broilers and to establish reference intervals (RIs) for RBC and WBC variables obtained by automated and manual methods.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>Automated HCT, WBC count, and differentials obtained using Sysmex XT-2000iV and XN-1000V on 309 and 56 EDTA blood samples from Ross 308 male broilers, respectively, were compared to manual PCV, manual WBC count, and differentials from November 2021 to November 2022. RIs for automated HCT and RBC variables (from impedance channel) on one hand, and manual PCV and WBC's variables on the other hand, were established.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>The Sysmex XT-2000iV was accurate for HCT and outperformed the XN-1000V but, both automated methods and manual methods cannot be considered interchangeable. Constant error, proportional error, and bias were identified and calculated for each analyzer for HCT. Both analyzers were unable to enumerate WBC. RIs were established as follows: HCT-XT (L/L) [0.27–0.36]; HCT-XN (L/L) [0.24–0.38]; manual PCV (L/L) [0.27–0.37]; manual WBC (×10<sup>9</sup>/L) [10.0–38.4]; manual differential (×10<sup>9</sup>/L) with heterophils [1.5–16.2]; lymphocytes [5.1–18.9]; monocytes [0.0–2.1]; eosinophils [0.0–2.2]; basophils [0.0–3.1].</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>Both analyzers are usable for the evaluation of HCT in Ross 308 chickens and in specific conditions for the other RBC variables. Manual methods are still preferred for the evaluation of WBC variables.</p>\\n </section>\\n </div>\",\"PeriodicalId\":23593,\"journal\":{\"name\":\"Veterinary clinical pathology\",\"volume\":\"54 2\",\"pages\":\"106-119\"},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2025-06-03\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1111/vcp.70009\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Veterinary clinical pathology\",\"FirstCategoryId\":\"97\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1111/vcp.70009\",\"RegionNum\":4,\"RegionCategory\":\"农林科学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"VETERINARY SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Veterinary clinical pathology","FirstCategoryId":"97","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/vcp.70009","RegionNum":4,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
Evaluation of blood erythroid parameters in male broiler chickens (Ross 308) with the Sysmex XT-2000iV and Sysmex XN-1000V analyzers and determination of hematological reference intervals obtained with manual and instrumental methods
Background
The presence of both nucleated RBC and thrombocytes in birds makes the determination of hematological variables difficult. This leads to time-consuming manual evaluation. New-generation analyzers are hypothesized to be both time-saving and accurate.
Objectives
The objective was to evaluate the performance of two hematology analyzers in 27-day-old broilers and to establish reference intervals (RIs) for RBC and WBC variables obtained by automated and manual methods.
Methods
Automated HCT, WBC count, and differentials obtained using Sysmex XT-2000iV and XN-1000V on 309 and 56 EDTA blood samples from Ross 308 male broilers, respectively, were compared to manual PCV, manual WBC count, and differentials from November 2021 to November 2022. RIs for automated HCT and RBC variables (from impedance channel) on one hand, and manual PCV and WBC's variables on the other hand, were established.
Results
The Sysmex XT-2000iV was accurate for HCT and outperformed the XN-1000V but, both automated methods and manual methods cannot be considered interchangeable. Constant error, proportional error, and bias were identified and calculated for each analyzer for HCT. Both analyzers were unable to enumerate WBC. RIs were established as follows: HCT-XT (L/L) [0.27–0.36]; HCT-XN (L/L) [0.24–0.38]; manual PCV (L/L) [0.27–0.37]; manual WBC (×109/L) [10.0–38.4]; manual differential (×109/L) with heterophils [1.5–16.2]; lymphocytes [5.1–18.9]; monocytes [0.0–2.1]; eosinophils [0.0–2.2]; basophils [0.0–3.1].
Conclusions
Both analyzers are usable for the evaluation of HCT in Ross 308 chickens and in specific conditions for the other RBC variables. Manual methods are still preferred for the evaluation of WBC variables.
期刊介绍:
Veterinary Clinical Pathology is the official journal of the American Society for Veterinary Clinical Pathology (ASVCP) and the European Society of Veterinary Clinical Pathology (ESVCP). The journal''s mission is to provide an international forum for communication and discussion of scientific investigations and new developments that advance the art and science of laboratory diagnosis in animals. Veterinary Clinical Pathology welcomes original experimental research and clinical contributions involving domestic, laboratory, avian, and wildlife species in the areas of hematology, hemostasis, immunopathology, clinical chemistry, cytopathology, surgical pathology, toxicology, endocrinology, laboratory and analytical techniques, instrumentation, quality assurance, and clinical pathology education.