用于辐射剂量计校准的荧光x射线场。

IF 0.8 4区 环境科学与生态学 Q4 ENVIRONMENTAL SCIENCES
Tomoya Tsuji, Hiroshi Yoshitomi, Fuminobu Sato, Yoshihiko Tanimura
{"title":"用于辐射剂量计校准的荧光x射线场。","authors":"Tomoya Tsuji, Hiroshi Yoshitomi, Fuminobu Sato, Yoshihiko Tanimura","doi":"10.1093/rpd/ncaf060","DOIUrl":null,"url":null,"abstract":"<p><p>An irradiation environment for dosemeters calibration in the fluorescence X-ray field was set up at the Facility of Radiation Standards of the Japan Atomic Energy Agency by re-establishing the fluorescence X-ray calibration field that was established in 1989. The reference air kerma rates were determined using reference ionization chambers calibrated using X-ray narrow-spectrum series calibration field of the national metrology standard in Japan. Since no X-ray calibration field using fluorescence X-rays is available in Japan, the measurements were corrected by factors that accounts for differences in the fluorescence X-ray calibration field energy and the X-ray narrow-spectrum series calibration field energy. The conversion coefficients and correction factors for air densities for the fluorescence X-ray calibration field, including those unavailable in the ISO 4037 standard series, were derived from the measured X-ray fluence rate spectra.</p>","PeriodicalId":20795,"journal":{"name":"Radiation protection dosimetry","volume":" ","pages":"648-662"},"PeriodicalIF":0.8000,"publicationDate":"2025-06-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Fluorescence X-ray field for radiation dosemeters calibration.\",\"authors\":\"Tomoya Tsuji, Hiroshi Yoshitomi, Fuminobu Sato, Yoshihiko Tanimura\",\"doi\":\"10.1093/rpd/ncaf060\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>An irradiation environment for dosemeters calibration in the fluorescence X-ray field was set up at the Facility of Radiation Standards of the Japan Atomic Energy Agency by re-establishing the fluorescence X-ray calibration field that was established in 1989. The reference air kerma rates were determined using reference ionization chambers calibrated using X-ray narrow-spectrum series calibration field of the national metrology standard in Japan. Since no X-ray calibration field using fluorescence X-rays is available in Japan, the measurements were corrected by factors that accounts for differences in the fluorescence X-ray calibration field energy and the X-ray narrow-spectrum series calibration field energy. The conversion coefficients and correction factors for air densities for the fluorescence X-ray calibration field, including those unavailable in the ISO 4037 standard series, were derived from the measured X-ray fluence rate spectra.</p>\",\"PeriodicalId\":20795,\"journal\":{\"name\":\"Radiation protection dosimetry\",\"volume\":\" \",\"pages\":\"648-662\"},\"PeriodicalIF\":0.8000,\"publicationDate\":\"2025-06-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Radiation protection dosimetry\",\"FirstCategoryId\":\"93\",\"ListUrlMain\":\"https://doi.org/10.1093/rpd/ncaf060\",\"RegionNum\":4,\"RegionCategory\":\"环境科学与生态学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENVIRONMENTAL SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation protection dosimetry","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1093/rpd/ncaf060","RegionNum":4,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENVIRONMENTAL SCIENCES","Score":null,"Total":0}
引用次数: 0

摘要

通过重建1989年建立的荧光x射线校准场,在日本原子能机构辐射标准设施建立了荧光x射线场剂量计校准的辐照环境。参考电离室采用日本国家计量标准的x射线窄谱串联校准场校准,确定了参考空气热率。由于日本没有使用荧光x射线的x射线校准场,因此根据考虑荧光x射线校准场能量和x射线窄谱系列校准场能量差异的因素对测量结果进行了校正。荧光x射线校准场的空气密度转换系数和校正系数,包括ISO 4037标准系列中没有的转换系数和校正系数,是从测量的x射线通量率光谱中导出的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Fluorescence X-ray field for radiation dosemeters calibration.

An irradiation environment for dosemeters calibration in the fluorescence X-ray field was set up at the Facility of Radiation Standards of the Japan Atomic Energy Agency by re-establishing the fluorescence X-ray calibration field that was established in 1989. The reference air kerma rates were determined using reference ionization chambers calibrated using X-ray narrow-spectrum series calibration field of the national metrology standard in Japan. Since no X-ray calibration field using fluorescence X-rays is available in Japan, the measurements were corrected by factors that accounts for differences in the fluorescence X-ray calibration field energy and the X-ray narrow-spectrum series calibration field energy. The conversion coefficients and correction factors for air densities for the fluorescence X-ray calibration field, including those unavailable in the ISO 4037 standard series, were derived from the measured X-ray fluence rate spectra.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Radiation protection dosimetry
Radiation protection dosimetry 环境科学-公共卫生、环境卫生与职业卫生
CiteScore
1.40
自引率
10.00%
发文量
223
审稿时长
6-12 weeks
期刊介绍: Radiation Protection Dosimetry covers all aspects of personal and environmental dosimetry and monitoring, for both ionising and non-ionising radiations. This includes biological aspects, physical concepts, biophysical dosimetry, external and internal personal dosimetry and monitoring, environmental and workplace monitoring, accident dosimetry, and dosimetry related to the protection of patients. Particular emphasis is placed on papers covering the fundamentals of dosimetry; units, radiation quantities and conversion factors. Papers covering archaeological dating are included only if the fundamental measurement method or technique, such as thermoluminescence, has direct application to personal dosimetry measurements. Papers covering the dosimetric aspects of radon or other naturally occurring radioactive materials and low level radiation are included. Animal experiments and ecological sample measurements are not included unless there is a significant relevant content reason.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信