{"title":"不同能量和场大小下千伏x射线百分比深度剂量的薄膜测量。","authors":"Tze Yee Lim, Dershan Luo, Jamshid Moradi Kurdestany, Xin Wang, Ramesh Tailor","doi":"10.1093/bjr/tqaf107","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Kilovoltage x-ray therapy, which includes superficial x-rays and orthovoltage x-rays, is particularly suited for treatments near the skin surface. Radiation dose deposition at depth is typically characterized by percentage depth dose (PDD) curves. However, PDD data are difficult to obtain in kilovoltage x-ray therapy, hence clinics often rely on British Journal of Radiology Supplement 25 (BJR-25) archival data. This study measured PDD for different energies and field sizes on the Xstrahl 300 kilovoltage x-ray therapy unit for comparison with BJR-25 data.</p><p><strong>Methods: </strong>We irradiated EBT3 film in water using different energy and field size combinations, then compared the acquired PDD curves with BJR-25 data. We used 75 kVp, 125 kVp, and 250 kVp beams, for 10 × 10 cm2 open field, 4 × 4 cm2 closed-ended square applicator, and open-ended circular applicators of 1.5 cm, 2.0 cm, 2.8 cm, and 4.7 cm diameter.</p><p><strong>Results: </strong>PDD for different energies and field sizes on the Xstrahl 300 kilovoltage x-ray therapy unit agreed well with BJR-25 data. The average difference between measured and BJR-25 data was -0.3 ± 1.8%; 85.9% of the individual local differences were within ±3%, 94.8% within ±4%, and 99.5% within ±5%.</p><p><strong>Conclusions: </strong>The PDD curves presented here, together with BJR-25 data, can serve as useful comparison datasets for the commissioning of kilovoltage x-ray machines.</p><p><strong>Advances in knowledge: </strong>Our film-based measurement technique provides high spatial resolution PDD datapoints for a modern kilovoltage x-ray unit, adding to the limited published data in the kilovoltage energy range.</p>","PeriodicalId":9306,"journal":{"name":"British Journal of Radiology","volume":" ","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2025-05-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Film measurements of kilovoltage x-ray percentage depth doses for different energies and field sizes.\",\"authors\":\"Tze Yee Lim, Dershan Luo, Jamshid Moradi Kurdestany, Xin Wang, Ramesh Tailor\",\"doi\":\"10.1093/bjr/tqaf107\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>Kilovoltage x-ray therapy, which includes superficial x-rays and orthovoltage x-rays, is particularly suited for treatments near the skin surface. Radiation dose deposition at depth is typically characterized by percentage depth dose (PDD) curves. However, PDD data are difficult to obtain in kilovoltage x-ray therapy, hence clinics often rely on British Journal of Radiology Supplement 25 (BJR-25) archival data. This study measured PDD for different energies and field sizes on the Xstrahl 300 kilovoltage x-ray therapy unit for comparison with BJR-25 data.</p><p><strong>Methods: </strong>We irradiated EBT3 film in water using different energy and field size combinations, then compared the acquired PDD curves with BJR-25 data. We used 75 kVp, 125 kVp, and 250 kVp beams, for 10 × 10 cm2 open field, 4 × 4 cm2 closed-ended square applicator, and open-ended circular applicators of 1.5 cm, 2.0 cm, 2.8 cm, and 4.7 cm diameter.</p><p><strong>Results: </strong>PDD for different energies and field sizes on the Xstrahl 300 kilovoltage x-ray therapy unit agreed well with BJR-25 data. The average difference between measured and BJR-25 data was -0.3 ± 1.8%; 85.9% of the individual local differences were within ±3%, 94.8% within ±4%, and 99.5% within ±5%.</p><p><strong>Conclusions: </strong>The PDD curves presented here, together with BJR-25 data, can serve as useful comparison datasets for the commissioning of kilovoltage x-ray machines.</p><p><strong>Advances in knowledge: </strong>Our film-based measurement technique provides high spatial resolution PDD datapoints for a modern kilovoltage x-ray unit, adding to the limited published data in the kilovoltage energy range.</p>\",\"PeriodicalId\":9306,\"journal\":{\"name\":\"British Journal of Radiology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2025-05-27\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"British Journal of Radiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/bjr/tqaf107\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"British Journal of Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/bjr/tqaf107","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:千伏x线治疗,包括浅表x线和正压x线,特别适用于皮肤表面附近的治疗。深度辐射剂量沉积的典型特征是百分比深度剂量(PDD)曲线。然而,在千伏x线治疗中很难获得PDD数据,因此诊所通常依赖于英国放射学杂志增刊25 (BJR-25)档案数据。本研究在Xstrahl 300千伏x射线治疗机上测量了不同能量和场大小的PDD,并与BJR-25数据进行了比较。方法:采用不同能量和场大小组合辐照EBT3薄膜,并将所得PDD曲线与BJR-25数据进行比较。我们使用75 kVp、125 kVp和250 kVp的光束,用于10 × 10 cm2的开场,4 × 4 cm2的封闭式方形涂抹器,以及直径为1.5 cm、2.0 cm、2.8 cm和4.7 cm的开放式圆形涂抹器。结果:在Xstrahl 300千伏x射线治疗机上不同能量和场大小的PDD与BJR-25的数据吻合较好。测量值与BJR-25数据的平均差异为-0.3±1.8%;85.9%的个体局部差异在±3%以内,94.8%在±4%以内,99.5%在±5%以内。结论:本文给出的PDD曲线与BJR-25数据可以作为千伏x线机调试的有用比较数据集。知识的进步:我们基于胶片的测量技术为现代千伏x射线装置提供了高空间分辨率的PDD数据点,增加了千伏能量范围内有限的已发表数据。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Film measurements of kilovoltage x-ray percentage depth doses for different energies and field sizes.

Objectives: Kilovoltage x-ray therapy, which includes superficial x-rays and orthovoltage x-rays, is particularly suited for treatments near the skin surface. Radiation dose deposition at depth is typically characterized by percentage depth dose (PDD) curves. However, PDD data are difficult to obtain in kilovoltage x-ray therapy, hence clinics often rely on British Journal of Radiology Supplement 25 (BJR-25) archival data. This study measured PDD for different energies and field sizes on the Xstrahl 300 kilovoltage x-ray therapy unit for comparison with BJR-25 data.

Methods: We irradiated EBT3 film in water using different energy and field size combinations, then compared the acquired PDD curves with BJR-25 data. We used 75 kVp, 125 kVp, and 250 kVp beams, for 10 × 10 cm2 open field, 4 × 4 cm2 closed-ended square applicator, and open-ended circular applicators of 1.5 cm, 2.0 cm, 2.8 cm, and 4.7 cm diameter.

Results: PDD for different energies and field sizes on the Xstrahl 300 kilovoltage x-ray therapy unit agreed well with BJR-25 data. The average difference between measured and BJR-25 data was -0.3 ± 1.8%; 85.9% of the individual local differences were within ±3%, 94.8% within ±4%, and 99.5% within ±5%.

Conclusions: The PDD curves presented here, together with BJR-25 data, can serve as useful comparison datasets for the commissioning of kilovoltage x-ray machines.

Advances in knowledge: Our film-based measurement technique provides high spatial resolution PDD datapoints for a modern kilovoltage x-ray unit, adding to the limited published data in the kilovoltage energy range.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
British Journal of Radiology
British Journal of Radiology 医学-核医学
CiteScore
5.30
自引率
3.80%
发文量
330
审稿时长
2-4 weeks
期刊介绍: BJR is the international research journal of the British Institute of Radiology and is the oldest scientific journal in the field of radiology and related sciences. Dating back to 1896, BJR’s history is radiology’s history, and the journal has featured some landmark papers such as the first description of Computed Tomography "Computerized transverse axial tomography" by Godfrey Hounsfield in 1973. A valuable historical resource, the complete BJR archive has been digitized from 1896. Quick Facts: - 2015 Impact Factor – 1.840 - Receipt to first decision – average of 6 weeks - Acceptance to online publication – average of 3 weeks - ISSN: 0007-1285 - eISSN: 1748-880X Open Access option
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信