Rouzbeh Kotaki, Ravi Shastri, Mohammad Ghasemi-Rad
{"title":"被光击碎:股动脉导管断裂及血管密封装置失效的隐患。","authors":"Rouzbeh Kotaki, Ravi Shastri, Mohammad Ghasemi-Rad","doi":"10.1177/15385744251332765","DOIUrl":null,"url":null,"abstract":"<p><p>We report a case of a 77-year-old male with pancreatic cancer and thrombocytopenia who presented with acute stroke symptoms and underwent successful endovascular thrombectomy. During femoral artery closure, an 8-F Angio-Seal device fractured, leaving catheter fragments in the right femoral artery, confirmed on imaging. Device analysis revealed oxidation-induced brittleness, leading to sheath fragmentation. This case highlights a rare complication of Angio-Seal devices, suggesting potential material vulnerabilities that warrant further investigation.</p>","PeriodicalId":94265,"journal":{"name":"Vascular and endovascular surgery","volume":" ","pages":"15385744251332765"},"PeriodicalIF":0.0000,"publicationDate":"2025-05-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Shattered by Light: Catheter Fractures and the Hidden Danger of Angio-Seal Device Failure in the Femoral Artery.\",\"authors\":\"Rouzbeh Kotaki, Ravi Shastri, Mohammad Ghasemi-Rad\",\"doi\":\"10.1177/15385744251332765\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>We report a case of a 77-year-old male with pancreatic cancer and thrombocytopenia who presented with acute stroke symptoms and underwent successful endovascular thrombectomy. During femoral artery closure, an 8-F Angio-Seal device fractured, leaving catheter fragments in the right femoral artery, confirmed on imaging. Device analysis revealed oxidation-induced brittleness, leading to sheath fragmentation. This case highlights a rare complication of Angio-Seal devices, suggesting potential material vulnerabilities that warrant further investigation.</p>\",\"PeriodicalId\":94265,\"journal\":{\"name\":\"Vascular and endovascular surgery\",\"volume\":\" \",\"pages\":\"15385744251332765\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-05-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Vascular and endovascular surgery\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1177/15385744251332765\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Vascular and endovascular surgery","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/15385744251332765","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Shattered by Light: Catheter Fractures and the Hidden Danger of Angio-Seal Device Failure in the Femoral Artery.
We report a case of a 77-year-old male with pancreatic cancer and thrombocytopenia who presented with acute stroke symptoms and underwent successful endovascular thrombectomy. During femoral artery closure, an 8-F Angio-Seal device fractured, leaving catheter fragments in the right femoral artery, confirmed on imaging. Device analysis revealed oxidation-induced brittleness, leading to sheath fragmentation. This case highlights a rare complication of Angio-Seal devices, suggesting potential material vulnerabilities that warrant further investigation.