{"title":"分析病人等深点位置精度与门户成像获得的在不同的龙门角度及其dosimetrical影响剂量交付头部和颈部癌症的。","authors":"Bharath Pandu, D Khanna, Mohandass Palanisamy, Saro Jacob, Tatarao Maddipati","doi":"10.1002/pro6.1234","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study aimed to analyze the effect of different gantry angles on portal imaging in terms of isocenter shifts and their dosimetric impact on dose delivery.</p><p><strong>Methods: </strong>Thirty patients with head and neck cancers were prospectively selected for this study. The reference anterior to posterior (AP) digitally reconstructed reference radiograph image was obtained at a gantry angle of 0°. The AP portal images were acquired at angles of 0°, 0.5°, 1°, 1.5°, and 2.0°. The average deviation of the isocenter shift with respect to the zero-gantry angle was calculated. To check the dosimetric effects, the reference fluence was compared with different fluences measured at different isocenter shifts.</p><p><strong>Results: </strong>The average isocenter shift differences in the lateral direction were 0.7 mm, 1.3 mm, 1.9 mm, and 2.5 mm. The average difference was <±0.1 mm for isocenter shift in the longitudinal direction. The results of the statistical analysis showed that the average isocenter shift and gamma pass rate with respect to the different isocenter position errors were significant in the lateral direction.</p><p><strong>Conclusion: </strong>The results of this study showed that as the gantry angle increased, the isocenter shifted and the percentage of deviation in the lateral direction increased.</p>","PeriodicalId":32406,"journal":{"name":"Precision Radiation Oncology","volume":"8 2","pages":"77-84"},"PeriodicalIF":0.0000,"publicationDate":"2024-06-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11935235/pdf/","citationCount":"0","resultStr":"{\"title\":\"Analyzing the patient isocenter positional accuracy with portal imaging acquired at different gantry angles and its dosimetrical impact on the dose delivery for head and neck cancers.\",\"authors\":\"Bharath Pandu, D Khanna, Mohandass Palanisamy, Saro Jacob, Tatarao Maddipati\",\"doi\":\"10.1002/pro6.1234\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>This study aimed to analyze the effect of different gantry angles on portal imaging in terms of isocenter shifts and their dosimetric impact on dose delivery.</p><p><strong>Methods: </strong>Thirty patients with head and neck cancers were prospectively selected for this study. The reference anterior to posterior (AP) digitally reconstructed reference radiograph image was obtained at a gantry angle of 0°. The AP portal images were acquired at angles of 0°, 0.5°, 1°, 1.5°, and 2.0°. The average deviation of the isocenter shift with respect to the zero-gantry angle was calculated. To check the dosimetric effects, the reference fluence was compared with different fluences measured at different isocenter shifts.</p><p><strong>Results: </strong>The average isocenter shift differences in the lateral direction were 0.7 mm, 1.3 mm, 1.9 mm, and 2.5 mm. The average difference was <±0.1 mm for isocenter shift in the longitudinal direction. The results of the statistical analysis showed that the average isocenter shift and gamma pass rate with respect to the different isocenter position errors were significant in the lateral direction.</p><p><strong>Conclusion: </strong>The results of this study showed that as the gantry angle increased, the isocenter shifted and the percentage of deviation in the lateral direction increased.</p>\",\"PeriodicalId\":32406,\"journal\":{\"name\":\"Precision Radiation Oncology\",\"volume\":\"8 2\",\"pages\":\"77-84\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-06-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11935235/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Precision Radiation Oncology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1002/pro6.1234\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/6/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Precision Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/pro6.1234","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

目的:分析不同门框角度对门静脉成像等中心偏移的影响及其对剂量传递的剂量学影响。方法:前瞻性选择30例头颈部肿瘤患者作为研究对象。以0°龙门角获得参考前后(AP)数字重建参考x线片图像。分别在0°、0.5°、1°、1.5°和2.0°角度获取AP门静脉图像。计算了等心位移相对于零门角的平均偏差。为了检查剂量学效应,将参考通量与在不同等中心位移下测量的不同通量进行了比较。结果:侧向等心位移差值分别为0.7 mm、1.3 mm、1.9 mm、2.5 mm。结论:本研究结果表明,随着龙门角度的增大,等心发生偏移,横向偏移比例增大。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Analyzing the patient isocenter positional accuracy with portal imaging acquired at different gantry angles and its dosimetrical impact on the dose delivery for head and neck cancers.

Objective: This study aimed to analyze the effect of different gantry angles on portal imaging in terms of isocenter shifts and their dosimetric impact on dose delivery.

Methods: Thirty patients with head and neck cancers were prospectively selected for this study. The reference anterior to posterior (AP) digitally reconstructed reference radiograph image was obtained at a gantry angle of 0°. The AP portal images were acquired at angles of 0°, 0.5°, 1°, 1.5°, and 2.0°. The average deviation of the isocenter shift with respect to the zero-gantry angle was calculated. To check the dosimetric effects, the reference fluence was compared with different fluences measured at different isocenter shifts.

Results: The average isocenter shift differences in the lateral direction were 0.7 mm, 1.3 mm, 1.9 mm, and 2.5 mm. The average difference was <±0.1 mm for isocenter shift in the longitudinal direction. The results of the statistical analysis showed that the average isocenter shift and gamma pass rate with respect to the different isocenter position errors were significant in the lateral direction.

Conclusion: The results of this study showed that as the gantry angle increased, the isocenter shifted and the percentage of deviation in the lateral direction increased.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Precision Radiation Oncology
Precision Radiation Oncology Medicine-Oncology
CiteScore
1.20
自引率
0.00%
发文量
32
审稿时长
13 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信