用于同时测定人血浆基质中钠离子和钾离子的便携式纸基传感器

IF 2.7 3区 化学 Q2 CHEMISTRY, ANALYTICAL
Mohammed Alqarni
{"title":"用于同时测定人血浆基质中钠离子和钾离子的便携式纸基传感器","authors":"Mohammed Alqarni","doi":"10.1039/D5AY00234F","DOIUrl":null,"url":null,"abstract":"<p >Accurate determination of sodium and potassium ions in biological fluids is crucial for managing various health conditions, as these electrolytes play a vital role in regulating physiological processes and maintaining overall well-being. A low-cost portable paper-based sensor has been developed for the simultaneous determination of sodium and potassium in the human plasma matrix. It comprises three electrodes on a single paper substrate: a solid-state polyvinyl butyral reference electrode, a sodium selective electrode and a potassium selective electrode. The sensor utilizes conductive functionalized multi-walled carbon nanotubes, selective ionophores and dyes to enhance sensing capabilities. It demonstrates high selectivity for sodium and potassium ions, with a linear range of 10<small><sup>−6</sup></small> to 10<small><sup>−2</sup></small> M and a Nernstian response (slopes of 59.08 ± 0.32 and 59.15 ± 0.28 mV per decade for sodium and potassium, respectively). The sensor provides a reproducible stable potential over 180 days and a dynamic response time of 10 s. The described paper-based sensor is an excellent tool for point-of-care analysis of sodium and potassium imbalances due to its simplicity, cost-effectiveness and disposability. It provides fast and reliable determination of sodium and potassium ions in the human plasma matrix, making it a valuable tool for managing their imbalances in clinical settings.</p>","PeriodicalId":64,"journal":{"name":"Analytical Methods","volume":" 17","pages":" 3499-3508"},"PeriodicalIF":2.7000,"publicationDate":"2025-04-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A portable paper-based sensor for simultaneous determination of sodium and potassium ions in the human plasma matrix\",\"authors\":\"Mohammed Alqarni\",\"doi\":\"10.1039/D5AY00234F\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p >Accurate determination of sodium and potassium ions in biological fluids is crucial for managing various health conditions, as these electrolytes play a vital role in regulating physiological processes and maintaining overall well-being. A low-cost portable paper-based sensor has been developed for the simultaneous determination of sodium and potassium in the human plasma matrix. It comprises three electrodes on a single paper substrate: a solid-state polyvinyl butyral reference electrode, a sodium selective electrode and a potassium selective electrode. The sensor utilizes conductive functionalized multi-walled carbon nanotubes, selective ionophores and dyes to enhance sensing capabilities. It demonstrates high selectivity for sodium and potassium ions, with a linear range of 10<small><sup>−6</sup></small> to 10<small><sup>−2</sup></small> M and a Nernstian response (slopes of 59.08 ± 0.32 and 59.15 ± 0.28 mV per decade for sodium and potassium, respectively). The sensor provides a reproducible stable potential over 180 days and a dynamic response time of 10 s. The described paper-based sensor is an excellent tool for point-of-care analysis of sodium and potassium imbalances due to its simplicity, cost-effectiveness and disposability. It provides fast and reliable determination of sodium and potassium ions in the human plasma matrix, making it a valuable tool for managing their imbalances in clinical settings.</p>\",\"PeriodicalId\":64,\"journal\":{\"name\":\"Analytical Methods\",\"volume\":\" 17\",\"pages\":\" 3499-3508\"},\"PeriodicalIF\":2.7000,\"publicationDate\":\"2025-04-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Analytical Methods\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://pubs.rsc.org/en/content/articlelanding/2025/ay/d5ay00234f\",\"RegionNum\":3,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"CHEMISTRY, ANALYTICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Analytical Methods","FirstCategoryId":"92","ListUrlMain":"https://pubs.rsc.org/en/content/articlelanding/2025/ay/d5ay00234f","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

摘要

准确测定生物体液中的钠离子和钾离子对于管理各种健康状况至关重要,因为这些电解质在调节生理过程和维持整体健康方面起着至关重要的作用。研制了一种低成本便携式纸基传感器,用于同时测定人血浆基质中的钠和钾。它包括三个电极在一个单一的纸衬底:一个固态聚乙烯醇丁醛基准电极,一个钠选择电极和一个钾选择电极。该传感器利用导电功能化多壁碳纳米管、选择性离子载体和染料来增强传感能力。它对钠离子和钾离子具有很高的选择性,线性范围为10−6 ~ 10−2 M,并具有Nernstian响应(钠离子和钾离子的斜率分别为59.08±0.32和59.15±0.28 mV / 10年)。该传感器可在180天内提供可重复的稳定电位和10秒的动态响应时间。由于其简单,成本效益和一次性,所描述的基于纸张的传感器是钠和钾失衡的护理点分析的极好工具。它提供了快速可靠的测定人血浆基质中的钠离子和钾离子,使其成为在临床环境中管理其不平衡的有价值的工具。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A portable paper-based sensor for simultaneous determination of sodium and potassium ions in the human plasma matrix

Accurate determination of sodium and potassium ions in biological fluids is crucial for managing various health conditions, as these electrolytes play a vital role in regulating physiological processes and maintaining overall well-being. A low-cost portable paper-based sensor has been developed for the simultaneous determination of sodium and potassium in the human plasma matrix. It comprises three electrodes on a single paper substrate: a solid-state polyvinyl butyral reference electrode, a sodium selective electrode and a potassium selective electrode. The sensor utilizes conductive functionalized multi-walled carbon nanotubes, selective ionophores and dyes to enhance sensing capabilities. It demonstrates high selectivity for sodium and potassium ions, with a linear range of 10−6 to 10−2 M and a Nernstian response (slopes of 59.08 ± 0.32 and 59.15 ± 0.28 mV per decade for sodium and potassium, respectively). The sensor provides a reproducible stable potential over 180 days and a dynamic response time of 10 s. The described paper-based sensor is an excellent tool for point-of-care analysis of sodium and potassium imbalances due to its simplicity, cost-effectiveness and disposability. It provides fast and reliable determination of sodium and potassium ions in the human plasma matrix, making it a valuable tool for managing their imbalances in clinical settings.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Analytical Methods
Analytical Methods CHEMISTRY, ANALYTICAL-FOOD SCIENCE & TECHNOLOGY
CiteScore
5.10
自引率
3.20%
发文量
569
审稿时长
1.8 months
期刊介绍: Early applied demonstrations of new analytical methods with clear societal impact
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信