对AlphaGUARD PQ2000PRO的简单改进,可准确测量土壤中的氡呼出率

IF 1.5 3区 化学 Q3 CHEMISTRY, ANALYTICAL
Feng Xiao, Shuaibin Liu, Ruomei Xie, Huiying Li, Haibo Yi, Jiale Sun, Zhipeng Liu, Jiulin Wu, Xinyue Yang, Shicheng Luo, Xianfa Mao, Hongbo Xu, Chenxi Zu, Hongzhi Yuan, Yanliang Tan
{"title":"对AlphaGUARD PQ2000PRO的简单改进,可准确测量土壤中的氡呼出率","authors":"Feng Xiao,&nbsp;Shuaibin Liu,&nbsp;Ruomei Xie,&nbsp;Huiying Li,&nbsp;Haibo Yi,&nbsp;Jiale Sun,&nbsp;Zhipeng Liu,&nbsp;Jiulin Wu,&nbsp;Xinyue Yang,&nbsp;Shicheng Luo,&nbsp;Xianfa Mao,&nbsp;Hongbo Xu,&nbsp;Chenxi Zu,&nbsp;Hongzhi Yuan,&nbsp;Yanliang Tan","doi":"10.1007/s10967-025-10018-0","DOIUrl":null,"url":null,"abstract":"<div><p>When using the AlphaGUARD PQ2000PRO to measure the radon exhalation rate from soil, its high sensitivity to thoron can lead to an overestimation of the radon level. To address this issue, a simple improvement was made to the AlphaGUARD PQ2000PRO in this study. A comparative experiment was conducted using the RAD7. The radon exhalation rate measured by the improved AlphaGUARD PQ2000PRO is in good agreement with the RAD7 exhalation rate within one standard deviation. The results show that this simple improvement effectively reduces interference from thoron, thus improving the accuracy of the AlphaGUARD PQ2000PRO in soil radon exhalation rate measurements.</p></div>","PeriodicalId":661,"journal":{"name":"Journal of Radioanalytical and Nuclear Chemistry","volume":"334 4","pages":"2839 - 2849"},"PeriodicalIF":1.5000,"publicationDate":"2025-02-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A simple improvement on the AlphaGUARD PQ2000PRO to accurately measure the radon exhalation rate from soil\",\"authors\":\"Feng Xiao,&nbsp;Shuaibin Liu,&nbsp;Ruomei Xie,&nbsp;Huiying Li,&nbsp;Haibo Yi,&nbsp;Jiale Sun,&nbsp;Zhipeng Liu,&nbsp;Jiulin Wu,&nbsp;Xinyue Yang,&nbsp;Shicheng Luo,&nbsp;Xianfa Mao,&nbsp;Hongbo Xu,&nbsp;Chenxi Zu,&nbsp;Hongzhi Yuan,&nbsp;Yanliang Tan\",\"doi\":\"10.1007/s10967-025-10018-0\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>When using the AlphaGUARD PQ2000PRO to measure the radon exhalation rate from soil, its high sensitivity to thoron can lead to an overestimation of the radon level. To address this issue, a simple improvement was made to the AlphaGUARD PQ2000PRO in this study. A comparative experiment was conducted using the RAD7. The radon exhalation rate measured by the improved AlphaGUARD PQ2000PRO is in good agreement with the RAD7 exhalation rate within one standard deviation. The results show that this simple improvement effectively reduces interference from thoron, thus improving the accuracy of the AlphaGUARD PQ2000PRO in soil radon exhalation rate measurements.</p></div>\",\"PeriodicalId\":661,\"journal\":{\"name\":\"Journal of Radioanalytical and Nuclear Chemistry\",\"volume\":\"334 4\",\"pages\":\"2839 - 2849\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2025-02-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Radioanalytical and Nuclear Chemistry\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://link.springer.com/article/10.1007/s10967-025-10018-0\",\"RegionNum\":3,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CHEMISTRY, ANALYTICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radioanalytical and Nuclear Chemistry","FirstCategoryId":"92","ListUrlMain":"https://link.springer.com/article/10.1007/s10967-025-10018-0","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

摘要

当使用AlphaGUARD PQ2000PRO测量土壤中的氡呼出率时,其对氡的高灵敏度可能导致对氡水平的高估。为了解决这个问题,本研究对AlphaGUARD PQ2000PRO进行了简单的改进。采用RAD7进行了对比试验。改进后的AlphaGUARD PQ2000PRO测得的氡呼出率与RAD7呼出率在一个标准差内吻合良好。结果表明,这一简单的改进有效地减少了来自钍的干扰,从而提高了AlphaGUARD PQ2000PRO在土壤氡呼出率测量中的准确性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A simple improvement on the AlphaGUARD PQ2000PRO to accurately measure the radon exhalation rate from soil

When using the AlphaGUARD PQ2000PRO to measure the radon exhalation rate from soil, its high sensitivity to thoron can lead to an overestimation of the radon level. To address this issue, a simple improvement was made to the AlphaGUARD PQ2000PRO in this study. A comparative experiment was conducted using the RAD7. The radon exhalation rate measured by the improved AlphaGUARD PQ2000PRO is in good agreement with the RAD7 exhalation rate within one standard deviation. The results show that this simple improvement effectively reduces interference from thoron, thus improving the accuracy of the AlphaGUARD PQ2000PRO in soil radon exhalation rate measurements.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
2.80
自引率
18.80%
发文量
504
审稿时长
2.2 months
期刊介绍: An international periodical publishing original papers, letters, review papers and short communications on nuclear chemistry. The subjects covered include: Nuclear chemistry, Radiochemistry, Radiation chemistry, Radiobiological chemistry, Environmental radiochemistry, Production and control of radioisotopes and labelled compounds, Nuclear power plant chemistry, Nuclear fuel chemistry, Radioanalytical chemistry, Radiation detection and measurement, Nuclear instrumentation and automation, etc.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信