{"title":"Dissolution of Bovine Palatal Tissue with Degassed Sodium Hypochlorite.","authors":"Justin Fang, Zhen Shen, Ransome van der Hoeven, David E Jaramillo","doi":"10.3390/dj13030110","DOIUrl":null,"url":null,"abstract":"<p><p><b>Objective:</b> To investigate the tissue dissolution efficacy of different concentrations of sodium hypochlorite (NaOCl) using the GentleWave<sup>®</sup> technology (GW) and compare it with that of passive ultrasonic irrigation. This study will provide a novel in vitro model used to evaluate the tissue dissolution efficacy of different concentrations of NaOCl using the GentleWave<sup>®</sup> procedure. <b>Materials and Methods:</b> Palatal bovine tissue was standardized by weight at 70-75 mg (average of 74.4 mg). The samples were divided into three groups of <i>n</i> = 10, Group 1: GW + 0.5% NaOCl, Group 2: GW + 3% NaOCl, and Group 3: ultrasonic group (US) + 6% NaOCl, and a control group of <i>n</i> = 2 with GW + H<sub>2</sub>O. A specialized CleanFlow instrument was manufactured for the GW groups. For the US group, an apparatus was developed to allow the tissue to be in close contact with a freely operating ultrasonic tip. Each group was operated with its specific irrigants and dissolution times were obtained unless the samples were not visually dissolved within 5 min, and the dissolution rates were calculated. Dissolution times and rates were analyzed using univariate analysis of variance followed by a <i>t</i>-test. <b>Results:</b> The GW groups with 0.5% and 3% NaOCl dissolved entire tissue samples within 5 min and had average dissolution times of 227.5 s (Group 1) and 81.5 s (Group 2). GW with water had a higher tissue dissolution rate than ultrasonics with 6% NaOCl. <b>Conclusions:</b> The GentleWave<sup>®</sup> using a lower concentration of NaOCl showed a similar clinical efficacy of dissolving tissues but displayed a significantly faster rate when compared with passive ultrasonic agitation.</p>","PeriodicalId":11269,"journal":{"name":"Dentistry Journal","volume":"13 3","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2025-02-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11941429/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dentistry Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/dj13030110","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

目的利用 GentleWave® 技术(GW)研究不同浓度次氯酸钠(NaOCl)的组织溶解效果,并与被动超声波冲洗进行比较。本研究将提供一种新型体外模型,用于评估使用 GentleWave® 程序的不同浓度次氯酸钠(NaOCl)的组织溶解效果。材料和方法:腭牛组织的标准重量为 70-75 毫克(平均 74.4 毫克)。样本分为三组,每组 n = 10,第一组:GW + 0.5% NaOCl,第二组:GW + 3% NaOCl,第三组:超声波组(US)+ 6% NaOCl,对照组 n = 2(GW + H2O)。为 GW 组制造了专门的 CleanFlow 仪器。对于 US 组,则开发了一种仪器,使组织与可自由操作的超声波尖端紧密接触。每组都使用特定的冲洗剂进行操作,除非样品在 5 分钟内肉眼无法看到溶解,否则都会获得溶解时间,并计算溶解速率。溶解时间和溶解率采用单变量方差分析,然后进行 t 检验。结果含 0.5% 和 3% NaOCl 的 GW 组在 5 分钟内溶解了整个组织样本,平均溶解时间分别为 227.5 秒(第 1 组)和 81.5 秒(第 2 组)。与使用 6% NaOCl 的超声波相比,使用水的 GW 的组织溶解率更高。结论:使用较低浓度 NaOCl 的 GentleWave® 在溶解组织方面具有相似的临床疗效,但与被动超声波搅拌相比,溶解速度明显更快。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Dissolution of Bovine Palatal Tissue with Degassed Sodium Hypochlorite.

Objective: To investigate the tissue dissolution efficacy of different concentrations of sodium hypochlorite (NaOCl) using the GentleWave® technology (GW) and compare it with that of passive ultrasonic irrigation. This study will provide a novel in vitro model used to evaluate the tissue dissolution efficacy of different concentrations of NaOCl using the GentleWave® procedure. Materials and Methods: Palatal bovine tissue was standardized by weight at 70-75 mg (average of 74.4 mg). The samples were divided into three groups of n = 10, Group 1: GW + 0.5% NaOCl, Group 2: GW + 3% NaOCl, and Group 3: ultrasonic group (US) + 6% NaOCl, and a control group of n = 2 with GW + H2O. A specialized CleanFlow instrument was manufactured for the GW groups. For the US group, an apparatus was developed to allow the tissue to be in close contact with a freely operating ultrasonic tip. Each group was operated with its specific irrigants and dissolution times were obtained unless the samples were not visually dissolved within 5 min, and the dissolution rates were calculated. Dissolution times and rates were analyzed using univariate analysis of variance followed by a t-test. Results: The GW groups with 0.5% and 3% NaOCl dissolved entire tissue samples within 5 min and had average dissolution times of 227.5 s (Group 1) and 81.5 s (Group 2). GW with water had a higher tissue dissolution rate than ultrasonics with 6% NaOCl. Conclusions: The GentleWave® using a lower concentration of NaOCl showed a similar clinical efficacy of dissolving tissues but displayed a significantly faster rate when compared with passive ultrasonic agitation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Dentistry Journal
Dentistry Journal Dentistry-Dentistry (all)
CiteScore
3.70
自引率
7.70%
发文量
213
审稿时长
11 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信