中国血浆儿茶酚胺和肾上腺素分析的一致性:三种主流市售液相色谱串联质谱试剂盒方法的比较

IF 1 4区 医学 Q4 MEDICAL LABORATORY TECHNOLOGY
Nijiao Qin, Qingxiang Liu, Zhenni Liu, Ziyang Li, Binghui Li, Zhe Guo, Zhiyang Chen, Weiming Wu, Min Hu, Qichen Long
{"title":"中国血浆儿茶酚胺和肾上腺素分析的一致性:三种主流市售液相色谱串联质谱试剂盒方法的比较","authors":"Nijiao Qin, Qingxiang Liu, Zhenni Liu, Ziyang Li, Binghui Li, Zhe Guo, Zhiyang Chen, Weiming Wu, Min Hu, Qichen Long","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The consistency of plasma catecholamines and metanephrines measurements is crucial for the screening and diagnosis in pheochromocytomas and paragangliomas (PPGLs). There are many commercially available kits for plasma catecholamines and metanephrines testing, but their performance and consistency information are unknown. This study evaluated the performance of three mainstream commercially available LC-MS/MS kit methods and evaluated their consistency.</p><p><strong>Methods: </strong>Performance validation of three kit methods was performed according to Clinical Laboratory Standards Institute (CLSI) document C-62A. 40 plasma samples from patients tested for PPGLs were analyzed in three mainstream commercially available kit methods. Agreement of test results from different kit methods was assessed by Passing-Bablok regression and Bland-Altman analysis. Recalibration using uniform calibrations was performed to identify the possible calibration difference.</p><p><strong>Results: </strong>The performance of three mainstream commercially available LC-MS/MS kit methods was satisfied. The mean deviations ranges were as follows: epinephrine(E), -41.1% ~ -2.4%; norepinephrine (NE), -34.3% ~ -19.1%; metanephrine (MN), -6.5% ~ 0.3% and normetanephrine (NMN), -2.3% to 0.0%. After recalibration, the mean deviation ranges of the E and NE were reduced remarkably (E: -18% to -2.4%, NE: -19.1% to 8.0%).</p><p><strong>Conclusions: </strong>For MN and NMN, the consistency between the three commercially available kit methods is acceptable. The results for the E and NE showed unsatisfying comparability for almost all method pairs. The application of uniform calibrators could improve consistency among the mainstream systems for the measurement of plasma catecholamines.</p>","PeriodicalId":8228,"journal":{"name":"Annals of clinical and laboratory science","volume":"55 1","pages":"115-127"},"PeriodicalIF":1.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The Consistency of Plasma Catecholamines and Metanephrines Analysis in China: A Comparison of Three Mainstream Commercially Available Liquid Chromatography Tandem Mass Spectrometry Kit Methods.\",\"authors\":\"Nijiao Qin, Qingxiang Liu, Zhenni Liu, Ziyang Li, Binghui Li, Zhe Guo, Zhiyang Chen, Weiming Wu, Min Hu, Qichen Long\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>The consistency of plasma catecholamines and metanephrines measurements is crucial for the screening and diagnosis in pheochromocytomas and paragangliomas (PPGLs). There are many commercially available kits for plasma catecholamines and metanephrines testing, but their performance and consistency information are unknown. This study evaluated the performance of three mainstream commercially available LC-MS/MS kit methods and evaluated their consistency.</p><p><strong>Methods: </strong>Performance validation of three kit methods was performed according to Clinical Laboratory Standards Institute (CLSI) document C-62A. 40 plasma samples from patients tested for PPGLs were analyzed in three mainstream commercially available kit methods. Agreement of test results from different kit methods was assessed by Passing-Bablok regression and Bland-Altman analysis. Recalibration using uniform calibrations was performed to identify the possible calibration difference.</p><p><strong>Results: </strong>The performance of three mainstream commercially available LC-MS/MS kit methods was satisfied. The mean deviations ranges were as follows: epinephrine(E), -41.1% ~ -2.4%; norepinephrine (NE), -34.3% ~ -19.1%; metanephrine (MN), -6.5% ~ 0.3% and normetanephrine (NMN), -2.3% to 0.0%. After recalibration, the mean deviation ranges of the E and NE were reduced remarkably (E: -18% to -2.4%, NE: -19.1% to 8.0%).</p><p><strong>Conclusions: </strong>For MN and NMN, the consistency between the three commercially available kit methods is acceptable. The results for the E and NE showed unsatisfying comparability for almost all method pairs. The application of uniform calibrators could improve consistency among the mainstream systems for the measurement of plasma catecholamines.</p>\",\"PeriodicalId\":8228,\"journal\":{\"name\":\"Annals of clinical and laboratory science\",\"volume\":\"55 1\",\"pages\":\"115-127\"},\"PeriodicalIF\":1.0000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annals of clinical and laboratory science\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of clinical and laboratory science","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:血浆儿茶酚胺和肾上腺素测定的一致性对嗜铬细胞瘤和副神经节瘤(PPGLs)的筛查和诊断至关重要。市面上有许多血浆儿茶酚胺和肾上腺素检测试剂盒,但其性能和一致性信息尚不清楚。本研究评估了三种主流市售LC-MS/MS试剂盒方法的性能,并评估了它们的一致性。方法:根据临床实验室标准协会(CLSI) C-62A文件对三种试剂盒方法进行性能验证。采用三种主流的市售试剂盒方法对40份PPGLs检测患者血浆样本进行分析。通过pass - bablok回归和Bland-Altman分析评估不同试剂盒方法检测结果的一致性。使用均匀校准进行重新校准,以确定可能的校准差异。结果:3种主流市售LC-MS/MS试剂盒检测方法均能满足要求。平均偏差范围为:肾上腺素(E), -41.1% ~ -2.4%;去甲肾上腺素(NE), -34.3% ~ -19.1%;去甲肾上腺素(MN) -6.5% ~ 0.3%,去甲肾上腺素(NMN) -2.3% ~ 0.0%。重新校准后,E和NE的平均偏差范围明显减小(E: -18%至-2.4%,NE: -19.1%至8.0%)。结论:对于MN和NMN,三种市售试剂盒方法之间的一致性是可以接受的。E和NE的结果显示几乎所有方法对的可比性都不令人满意。统一校准器的应用可以提高血浆儿茶酚胺测量主流系统之间的一致性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The Consistency of Plasma Catecholamines and Metanephrines Analysis in China: A Comparison of Three Mainstream Commercially Available Liquid Chromatography Tandem Mass Spectrometry Kit Methods.

Objective: The consistency of plasma catecholamines and metanephrines measurements is crucial for the screening and diagnosis in pheochromocytomas and paragangliomas (PPGLs). There are many commercially available kits for plasma catecholamines and metanephrines testing, but their performance and consistency information are unknown. This study evaluated the performance of three mainstream commercially available LC-MS/MS kit methods and evaluated their consistency.

Methods: Performance validation of three kit methods was performed according to Clinical Laboratory Standards Institute (CLSI) document C-62A. 40 plasma samples from patients tested for PPGLs were analyzed in three mainstream commercially available kit methods. Agreement of test results from different kit methods was assessed by Passing-Bablok regression and Bland-Altman analysis. Recalibration using uniform calibrations was performed to identify the possible calibration difference.

Results: The performance of three mainstream commercially available LC-MS/MS kit methods was satisfied. The mean deviations ranges were as follows: epinephrine(E), -41.1% ~ -2.4%; norepinephrine (NE), -34.3% ~ -19.1%; metanephrine (MN), -6.5% ~ 0.3% and normetanephrine (NMN), -2.3% to 0.0%. After recalibration, the mean deviation ranges of the E and NE were reduced remarkably (E: -18% to -2.4%, NE: -19.1% to 8.0%).

Conclusions: For MN and NMN, the consistency between the three commercially available kit methods is acceptable. The results for the E and NE showed unsatisfying comparability for almost all method pairs. The application of uniform calibrators could improve consistency among the mainstream systems for the measurement of plasma catecholamines.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Annals of clinical and laboratory science
Annals of clinical and laboratory science 医学-医学实验技术
CiteScore
1.60
自引率
0.00%
发文量
112
审稿时长
6-12 weeks
期刊介绍: The Annals of Clinical & Laboratory Science welcomes manuscripts that report research in clinical science, including pathology, clinical chemistry, biotechnology, molecular biology, cytogenetics, microbiology, immunology, hematology, transfusion medicine, organ and tissue transplantation, therapeutics, toxicology, and clinical informatics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信