{"title":"消化道内窥镜再处理过程中污染率的比较评估:消毒与高水平消毒","authors":"Tanyaporn Chantarojanasiri, Rachanikorn Rungrueangmaitree, Siriporn Thongsri, Urasa Jampa-ngern, Thawee Ratanachu-Ek","doi":"10.1002/deo2.70093","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objectives</h3>\n \n <p>This study aimed to evaluate the disparity in culture results between sterilization and high-level disinfection (HLD) for duodenoscopes and linear endoscopic ultrasound (EUS), and to assess the effectiveness of different bacterial contamination detection methods.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>This is a prospective randomized study, including duodenoscopes and linear EUS with adenosine triphosphate bioluminescence assay values below 200 relative light units after manual cleaning which were randomly assigned to undergo either sterilization or HLD in a 1:1 ratio. Following disinfection, all endoscopes were subjected to adenosine triphosphate bioluminescence assay testing and cultures using both swab and liquid samples from endoscope channels.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Totally 752 endoscopes (444 duodenoscopes and 308 linear EUS) were studied. After disinfection, the positive culture rates for the sterilization and HLD groups were 5.9% and 7.2%, respectively (<i>p</i> = 0.460). No significant difference in contamination rates was observed between duodenoscopes and linear EUS (5.9% and 7.5%, respectively; <i>p</i> = 0.379), and no significant association between contamination rates and the presence of biliary stones was seen (7.3% vs. 6.9%; <i>p</i> = 0.613). The detection rate of bacteria from liquid samples taken from endoscope channels was 0.5%, which was significantly lower than the swabbing method (6.0%, <i>p</i> < 0.001).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>This study found no statistically significant difference in contamination rates between sterilization and HLD methods for gastrointestinal endoscope reprocessing. The type of endoscope and the presence of biliary stones did not influence the positive culture rate. The swabbing method showed significantly higher bacterial detection when compared with liquid samples.</p>\n </section>\n </div>","PeriodicalId":93973,"journal":{"name":"DEN open","volume":"5 1","pages":""},"PeriodicalIF":1.4000,"publicationDate":"2025-03-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/deo2.70093","citationCount":"0","resultStr":"{\"title\":\"A comparative assessment of contamination rates in gastrointestinal endoscope reprocessing: sterilization versus high-level disinfection\",\"authors\":\"Tanyaporn Chantarojanasiri, Rachanikorn Rungrueangmaitree, Siriporn Thongsri, Urasa Jampa-ngern, Thawee Ratanachu-Ek\",\"doi\":\"10.1002/deo2.70093\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Objectives</h3>\\n \\n <p>This study aimed to evaluate the disparity in culture results between sterilization and high-level disinfection (HLD) for duodenoscopes and linear endoscopic ultrasound (EUS), and to assess the effectiveness of different bacterial contamination detection methods.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>This is a prospective randomized study, including duodenoscopes and linear EUS with adenosine triphosphate bioluminescence assay values below 200 relative light units after manual cleaning which were randomly assigned to undergo either sterilization or HLD in a 1:1 ratio. Following disinfection, all endoscopes were subjected to adenosine triphosphate bioluminescence assay testing and cultures using both swab and liquid samples from endoscope channels.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>Totally 752 endoscopes (444 duodenoscopes and 308 linear EUS) were studied. After disinfection, the positive culture rates for the sterilization and HLD groups were 5.9% and 7.2%, respectively (<i>p</i> = 0.460). No significant difference in contamination rates was observed between duodenoscopes and linear EUS (5.9% and 7.5%, respectively; <i>p</i> = 0.379), and no significant association between contamination rates and the presence of biliary stones was seen (7.3% vs. 6.9%; <i>p</i> = 0.613). The detection rate of bacteria from liquid samples taken from endoscope channels was 0.5%, which was significantly lower than the swabbing method (6.0%, <i>p</i> < 0.001).</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>This study found no statistically significant difference in contamination rates between sterilization and HLD methods for gastrointestinal endoscope reprocessing. The type of endoscope and the presence of biliary stones did not influence the positive culture rate. The swabbing method showed significantly higher bacterial detection when compared with liquid samples.</p>\\n </section>\\n </div>\",\"PeriodicalId\":93973,\"journal\":{\"name\":\"DEN open\",\"volume\":\"5 1\",\"pages\":\"\"},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2025-03-09\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1002/deo2.70093\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"DEN open\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/deo2.70093\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"GASTROENTEROLOGY & HEPATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"DEN open","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/deo2.70093","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"GASTROENTEROLOGY & HEPATOLOGY","Score":null,"Total":0}
A comparative assessment of contamination rates in gastrointestinal endoscope reprocessing: sterilization versus high-level disinfection
Objectives
This study aimed to evaluate the disparity in culture results between sterilization and high-level disinfection (HLD) for duodenoscopes and linear endoscopic ultrasound (EUS), and to assess the effectiveness of different bacterial contamination detection methods.
Methods
This is a prospective randomized study, including duodenoscopes and linear EUS with adenosine triphosphate bioluminescence assay values below 200 relative light units after manual cleaning which were randomly assigned to undergo either sterilization or HLD in a 1:1 ratio. Following disinfection, all endoscopes were subjected to adenosine triphosphate bioluminescence assay testing and cultures using both swab and liquid samples from endoscope channels.
Results
Totally 752 endoscopes (444 duodenoscopes and 308 linear EUS) were studied. After disinfection, the positive culture rates for the sterilization and HLD groups were 5.9% and 7.2%, respectively (p = 0.460). No significant difference in contamination rates was observed between duodenoscopes and linear EUS (5.9% and 7.5%, respectively; p = 0.379), and no significant association between contamination rates and the presence of biliary stones was seen (7.3% vs. 6.9%; p = 0.613). The detection rate of bacteria from liquid samples taken from endoscope channels was 0.5%, which was significantly lower than the swabbing method (6.0%, p < 0.001).
Conclusions
This study found no statistically significant difference in contamination rates between sterilization and HLD methods for gastrointestinal endoscope reprocessing. The type of endoscope and the presence of biliary stones did not influence the positive culture rate. The swabbing method showed significantly higher bacterial detection when compared with liquid samples.