根据国际标准化组织通用标准(ISO 81060-2:2018/AMD 1:2020)协议,Combei BP880W手腕装置在一般人群中自我血压测量的准确性。

IF 2.6 Q2 PERIPHERAL VASCULAR DISEASE
Vascular Health and Risk Management Pub Date : 2025-02-18 eCollection Date: 2025-01-01 DOI:10.2147/VHRM.S513013
Zoya Hakobyan, Parounak Zelveian, Heghine Gharibyan, Jirar Topouchian, Roland Asmar
{"title":"根据国际标准化组织通用标准(ISO 81060-2:2018/AMD 1:2020)协议,Combei BP880W手腕装置在一般人群中自我血压测量的准确性。","authors":"Zoya Hakobyan, Parounak Zelveian, Heghine Gharibyan, Jirar Topouchian, Roland Asmar","doi":"10.2147/VHRM.S513013","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Scientific societies universally recommend evaluating the accuracy of electronic devices designed for blood pressure (BP) measurement using established validation protocols.</p><p><strong>Objective: </strong>This study aimed to assess the accuracy of the Combei BP880W wrist device for BP measurement in the general population, according to the ISO 81060-2:2018/AMD 1:2020 Universal Standard.</p><p><strong>Methods: </strong>The Combei BP880W is an oscillometric device designed to measure BP at the wrist. This study adhered to the ISO 81060-2:2018/AMD 1:2020 protocol and employed the same-arm sequential BP measurement method. A total of 85 participants, meeting protocol-specified age, gender, BP, and cuff distribution criteria, were included. The accuracy analysis utilized Criterion 1 (differences and standard deviations between reference and test device measurements) and Criterion 2 (intra-individual standard deviation of BP differences).</p><p><strong>Results: </strong>Eighty-five participants were included. Mean BP differences between the simultaneous observer measurements were -0.2 ± 1.9 mmHg for systolic BP (SBP) and 0.1 ± 1.9 mmHg for diastolic BP (DBP). For Criterion 1, the mean difference ± standard deviation (SD) between the reference and test device measurements were -2.7 ± 5.9 mmHg (SBP) and -2.0 ± 3.9 mmHg (DBP), meeting the required threshold (≤ 5 ± 8 mmHg). For Criterion 2, intra-individual SDs were 4.6 mmHg (SBP) and 3.4 mmHg (DBP), both below the respective limits (≤ 6.39 mmHg for SBP and ≤ 6.65 mmHg for DBP).</p><p><strong>Conclusion: </strong>The Combei BP880W wrist device meets the accuracy requirements of the ISO 81060-2:2018/AMD 1:2020 protocol, supporting its use for home BP monitoring in the general population.</p>","PeriodicalId":23597,"journal":{"name":"Vascular Health and Risk Management","volume":"21 ","pages":"75-84"},"PeriodicalIF":2.6000,"publicationDate":"2025-02-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11846527/pdf/","citationCount":"0","resultStr":"{\"title\":\"Accuracy of the Combei BP880W Wrist Device for Self-Blood Pressure Measurements in General Population According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 1:2020) Protocol.\",\"authors\":\"Zoya Hakobyan, Parounak Zelveian, Heghine Gharibyan, Jirar Topouchian, Roland Asmar\",\"doi\":\"10.2147/VHRM.S513013\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Scientific societies universally recommend evaluating the accuracy of electronic devices designed for blood pressure (BP) measurement using established validation protocols.</p><p><strong>Objective: </strong>This study aimed to assess the accuracy of the Combei BP880W wrist device for BP measurement in the general population, according to the ISO 81060-2:2018/AMD 1:2020 Universal Standard.</p><p><strong>Methods: </strong>The Combei BP880W is an oscillometric device designed to measure BP at the wrist. This study adhered to the ISO 81060-2:2018/AMD 1:2020 protocol and employed the same-arm sequential BP measurement method. A total of 85 participants, meeting protocol-specified age, gender, BP, and cuff distribution criteria, were included. The accuracy analysis utilized Criterion 1 (differences and standard deviations between reference and test device measurements) and Criterion 2 (intra-individual standard deviation of BP differences).</p><p><strong>Results: </strong>Eighty-five participants were included. Mean BP differences between the simultaneous observer measurements were -0.2 ± 1.9 mmHg for systolic BP (SBP) and 0.1 ± 1.9 mmHg for diastolic BP (DBP). For Criterion 1, the mean difference ± standard deviation (SD) between the reference and test device measurements were -2.7 ± 5.9 mmHg (SBP) and -2.0 ± 3.9 mmHg (DBP), meeting the required threshold (≤ 5 ± 8 mmHg). For Criterion 2, intra-individual SDs were 4.6 mmHg (SBP) and 3.4 mmHg (DBP), both below the respective limits (≤ 6.39 mmHg for SBP and ≤ 6.65 mmHg for DBP).</p><p><strong>Conclusion: </strong>The Combei BP880W wrist device meets the accuracy requirements of the ISO 81060-2:2018/AMD 1:2020 protocol, supporting its use for home BP monitoring in the general population.</p>\",\"PeriodicalId\":23597,\"journal\":{\"name\":\"Vascular Health and Risk Management\",\"volume\":\"21 \",\"pages\":\"75-84\"},\"PeriodicalIF\":2.6000,\"publicationDate\":\"2025-02-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11846527/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Vascular Health and Risk Management\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2147/VHRM.S513013\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q2\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Vascular Health and Risk Management","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/VHRM.S513013","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

摘要

背景:科学协会普遍推荐使用已建立的验证方案评估用于血压测量的电子设备的准确性。目的:本研究旨在根据ISO 81060-2:2018/AMD 1:2020通用标准,评估康贝BP880W腕带血压测量仪在普通人群中的准确性。方法:Combei BP880W是一种用于测量腕部血压的振荡装置。本研究遵循ISO 81060-2:2018/AMD 1:2020协议,采用同臂顺序血压测量方法。共纳入85名符合方案规定的年龄、性别、血压和袖带分布标准的参与者。准确度分析采用标准1(参考和测试设备测量值之间的差异和标准差)和标准2(个体内血压差异的标准差)。结果:共纳入85名受试者。同时观察测量的平均血压差异为收缩压(SBP)为-0.2±1.9 mmHg,舒张压(DBP)为0.1±1.9 mmHg。对于标准1,参考文献和测试装置测量值之间的平均差值±标准差(SD)为-2.7±5.9 mmHg(收缩压)和-2.0±3.9 mmHg (DBP),满足要求的阈值(≤5±8 mmHg)。对于标准2,个体内SDs分别为4.6 mmHg(收缩压)和3.4 mmHg(舒张压),均低于各自的限值(收缩压≤6.39 mmHg,舒张压≤6.65 mmHg)。结论:康贝BP880W腕带装置满足ISO 81060-2:2018/AMD 1:2020协议的精度要求,支持其用于普通人群的家庭血压监测。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Accuracy of the Combei BP880W Wrist Device for Self-Blood Pressure Measurements in General Population According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 1:2020) Protocol.

Background: Scientific societies universally recommend evaluating the accuracy of electronic devices designed for blood pressure (BP) measurement using established validation protocols.

Objective: This study aimed to assess the accuracy of the Combei BP880W wrist device for BP measurement in the general population, according to the ISO 81060-2:2018/AMD 1:2020 Universal Standard.

Methods: The Combei BP880W is an oscillometric device designed to measure BP at the wrist. This study adhered to the ISO 81060-2:2018/AMD 1:2020 protocol and employed the same-arm sequential BP measurement method. A total of 85 participants, meeting protocol-specified age, gender, BP, and cuff distribution criteria, were included. The accuracy analysis utilized Criterion 1 (differences and standard deviations between reference and test device measurements) and Criterion 2 (intra-individual standard deviation of BP differences).

Results: Eighty-five participants were included. Mean BP differences between the simultaneous observer measurements were -0.2 ± 1.9 mmHg for systolic BP (SBP) and 0.1 ± 1.9 mmHg for diastolic BP (DBP). For Criterion 1, the mean difference ± standard deviation (SD) between the reference and test device measurements were -2.7 ± 5.9 mmHg (SBP) and -2.0 ± 3.9 mmHg (DBP), meeting the required threshold (≤ 5 ± 8 mmHg). For Criterion 2, intra-individual SDs were 4.6 mmHg (SBP) and 3.4 mmHg (DBP), both below the respective limits (≤ 6.39 mmHg for SBP and ≤ 6.65 mmHg for DBP).

Conclusion: The Combei BP880W wrist device meets the accuracy requirements of the ISO 81060-2:2018/AMD 1:2020 protocol, supporting its use for home BP monitoring in the general population.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Vascular Health and Risk Management
Vascular Health and Risk Management PERIPHERAL VASCULAR DISEASE-
CiteScore
4.20
自引率
3.40%
发文量
109
审稿时长
16 weeks
期刊介绍: An international, peer-reviewed journal of therapeutics and risk management, focusing on concise rapid reporting of clinical studies on the processes involved in the maintenance of vascular health; the monitoring, prevention, and treatment of vascular disease and its sequelae; and the involvement of metabolic disorders, particularly diabetes. In addition, the journal will also seek to define drug usage in terms of ultimate uptake and acceptance by the patient and healthcare professional.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信