根据 AAMI/ESH/ISO (ISO 81060-2:2018) + AMD1:2020 验证欧姆龙 J760 电子血压计在普通人群中的准确性。

IF 1.8 4区 医学
Blood Pressure Pub Date : 2025-12-01 Epub Date: 2025-02-25 DOI:10.1080/08037051.2025.2469264
Bihua Chen, Guorong Chen, Hong Jiang, Ya Sun, Jie Chen, Dan Wang, Mingzhi Chen, Jia Shi, Yan Peng, Yimin Cheng, Cheng Yang, Yuanyuan Ding, Jin Su, Ming Liu, Fulai Shen, Yicheng Qiu, Yi Shen, Qiyun Cao, Jiayu Su
{"title":"根据 AAMI/ESH/ISO (ISO 81060-2:2018) + AMD1:2020 验证欧姆龙 J760 电子血压计在普通人群中的准确性。","authors":"Bihua Chen, Guorong Chen, Hong Jiang, Ya Sun, Jie Chen, Dan Wang, Mingzhi Chen, Jia Shi, Yan Peng, Yimin Cheng, Cheng Yang, Yuanyuan Ding, Jin Su, Ming Liu, Fulai Shen, Yicheng Qiu, Yi Shen, Qiyun Cao, Jiayu Su","doi":"10.1080/08037051.2025.2469264","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The accuracy of the Omron J760 electronic blood pressure (BP) monitor for upper arm BP measurement in the adult general population was validated following the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 protocol. We expect that this device can be used for home blood pressure monitoring, with its measurements serving as a clinical reference.</p><p><strong>Methods: </strong>Subjects meeting the age, gender, BP, and cuff distribution criteria as specified by the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 were recruited. BP measurements were conducted using the same-arm sequential method. The test device cuff was suitable for arm circumferences ranging from 22.0 cm to 42.0 cm.</p><p><strong>Results: </strong>Eighty-nine participants were initially recruited, 85 were evaluated after excluding 4 participants. The mean age of the participants was 48.5 ± 15.17 years. For validation criterion 1, the mean ± standard deviation (SD) of the differences between the test device and the reference BP measurements were 0.2 ± 5.74 mmHg for systolic BP and -0.9 ± 4.69 mmHg for diastolic BP. These results met the AAMI/ESH/ISO (ISO 81060-2:2018)+Amd1:2020 standard, which requires differences of ≤5 ± ≤8 mmHg. For validation criterion 2, the mean differences between the test device and the reference device were 0.2 ± 5.10 mmHg for systolic BP and -0.9 ± 4.30 mmHg for diastolic BP. This criterion was satisfied with criterion 2 of ≤6.95 mmHg for systolic BP and ≤6.88 mmHg for diastolic BP.</p><p><strong>Conclusion: </strong>The Omron J760 electronic BP monitor meets the requirements of the AAMI/ESH/ISO (ISO 81060-2:2018)+Amd 1:2020 validation standard.</p>","PeriodicalId":9000,"journal":{"name":"Blood Pressure","volume":" ","pages":"2469264"},"PeriodicalIF":1.8000,"publicationDate":"2025-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Validating the accuracy of Omron J760 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018) + AMD1:2020.\",\"authors\":\"Bihua Chen, Guorong Chen, Hong Jiang, Ya Sun, Jie Chen, Dan Wang, Mingzhi Chen, Jia Shi, Yan Peng, Yimin Cheng, Cheng Yang, Yuanyuan Ding, Jin Su, Ming Liu, Fulai Shen, Yicheng Qiu, Yi Shen, Qiyun Cao, Jiayu Su\",\"doi\":\"10.1080/08037051.2025.2469264\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>The accuracy of the Omron J760 electronic blood pressure (BP) monitor for upper arm BP measurement in the adult general population was validated following the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 protocol. We expect that this device can be used for home blood pressure monitoring, with its measurements serving as a clinical reference.</p><p><strong>Methods: </strong>Subjects meeting the age, gender, BP, and cuff distribution criteria as specified by the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 were recruited. BP measurements were conducted using the same-arm sequential method. The test device cuff was suitable for arm circumferences ranging from 22.0 cm to 42.0 cm.</p><p><strong>Results: </strong>Eighty-nine participants were initially recruited, 85 were evaluated after excluding 4 participants. The mean age of the participants was 48.5 ± 15.17 years. For validation criterion 1, the mean ± standard deviation (SD) of the differences between the test device and the reference BP measurements were 0.2 ± 5.74 mmHg for systolic BP and -0.9 ± 4.69 mmHg for diastolic BP. These results met the AAMI/ESH/ISO (ISO 81060-2:2018)+Amd1:2020 standard, which requires differences of ≤5 ± ≤8 mmHg. For validation criterion 2, the mean differences between the test device and the reference device were 0.2 ± 5.10 mmHg for systolic BP and -0.9 ± 4.30 mmHg for diastolic BP. This criterion was satisfied with criterion 2 of ≤6.95 mmHg for systolic BP and ≤6.88 mmHg for diastolic BP.</p><p><strong>Conclusion: </strong>The Omron J760 electronic BP monitor meets the requirements of the AAMI/ESH/ISO (ISO 81060-2:2018)+Amd 1:2020 validation standard.</p>\",\"PeriodicalId\":9000,\"journal\":{\"name\":\"Blood Pressure\",\"volume\":\" \",\"pages\":\"2469264\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2025-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1080/08037051.2025.2469264\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/2/25 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/08037051.2025.2469264","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/2/25 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本文章由计算机程序翻译,如有差异,请以英文原文为准。
Validating the accuracy of Omron J760 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018) + AMD1:2020.

Purpose: The accuracy of the Omron J760 electronic blood pressure (BP) monitor for upper arm BP measurement in the adult general population was validated following the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 protocol. We expect that this device can be used for home blood pressure monitoring, with its measurements serving as a clinical reference.

Methods: Subjects meeting the age, gender, BP, and cuff distribution criteria as specified by the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 were recruited. BP measurements were conducted using the same-arm sequential method. The test device cuff was suitable for arm circumferences ranging from 22.0 cm to 42.0 cm.

Results: Eighty-nine participants were initially recruited, 85 were evaluated after excluding 4 participants. The mean age of the participants was 48.5 ± 15.17 years. For validation criterion 1, the mean ± standard deviation (SD) of the differences between the test device and the reference BP measurements were 0.2 ± 5.74 mmHg for systolic BP and -0.9 ± 4.69 mmHg for diastolic BP. These results met the AAMI/ESH/ISO (ISO 81060-2:2018)+Amd1:2020 standard, which requires differences of ≤5 ± ≤8 mmHg. For validation criterion 2, the mean differences between the test device and the reference device were 0.2 ± 5.10 mmHg for systolic BP and -0.9 ± 4.30 mmHg for diastolic BP. This criterion was satisfied with criterion 2 of ≤6.95 mmHg for systolic BP and ≤6.88 mmHg for diastolic BP.

Conclusion: The Omron J760 electronic BP monitor meets the requirements of the AAMI/ESH/ISO (ISO 81060-2:2018)+Amd 1:2020 validation standard.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Blood Pressure
Blood Pressure Medicine-Cardiology and Cardiovascular Medicine
CiteScore
3.20
自引率
5.60%
发文量
41
期刊介绍: For outstanding coverage of the latest advances in hypertension research, turn to Blood Pressure, a primary source for authoritative and timely information on all aspects of hypertension research and management. Features include: • Physiology and pathophysiology of blood pressure regulation • Primary and secondary hypertension • Cerebrovascular and cardiovascular complications of hypertension • Detection, treatment and follow-up of hypertension • Non pharmacological and pharmacological management • Large outcome trials in hypertension.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信