{"title":"选择性激光熔融法制备上颌全口义齿金属框架,构建方位对内凹面真实性和适应性的影响。","authors":"Wenrui Qu, Atsushi Takaichi, Yuka Kajima, Shingo Kamijo, Noriyuki Wakabayashi","doi":"10.2186/jpr.JPR_D_24_00195","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>We aimed to evaluate the trueness and adaptation of the intaglio surfaces of complete denture metal frameworks fabricated using selective laser melting (SLM) with different building orientations (0°, 45°, and 90°).</p><p><strong>Methods: </strong>Cobalt-chromium maxillary complete denture metal frameworks were fabricated at 0°, 45°, and 90° using SLM. Trueness measurements were conducted using the best-fit alignment method of three-dimensional inspection software, and adaptation was evaluated using the silicone replica technique. Nonparametric Kruskal-Wallis and Bonferroni tests were performed to determine differences among the groups (P = 0.05).</p><p><strong>Results: </strong>No significant differences were observed in the root mean square (RMS), positive, and negative deviation values between 0° and 45° group (P > 0.05), and the median, first quartile, and third quartile of each were considered clinically acceptable. For the 90° group, the RMS, positive deviation, and negative deviation values were significantly higher than those of the other groups (P < 0.05). The overall adaptation of the 45° group was better than that of the 0° and 90° groups (P < 0.05), whereas the 0° and 90° groups showed no significant difference (P > 0.05).</p><p><strong>Conclusions: </strong>The trueness and adaptation of the intaglio surface of cobalt-chromium maxillary complete denture metal frameworks were affected by the building orientation through SLM. Considering both trueness and adaptation, 45° is recommended as the most efficient build orientation for clinical applications.</p>","PeriodicalId":16887,"journal":{"name":"Journal of prosthodontic research","volume":" ","pages":""},"PeriodicalIF":3.2000,"publicationDate":"2025-02-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effect of build orientation on trueness and adaptation of intaglio surface of maxillary complete denture metal frameworks fabricated by selective laser melting.\",\"authors\":\"Wenrui Qu, Atsushi Takaichi, Yuka Kajima, Shingo Kamijo, Noriyuki Wakabayashi\",\"doi\":\"10.2186/jpr.JPR_D_24_00195\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>We aimed to evaluate the trueness and adaptation of the intaglio surfaces of complete denture metal frameworks fabricated using selective laser melting (SLM) with different building orientations (0°, 45°, and 90°).</p><p><strong>Methods: </strong>Cobalt-chromium maxillary complete denture metal frameworks were fabricated at 0°, 45°, and 90° using SLM. Trueness measurements were conducted using the best-fit alignment method of three-dimensional inspection software, and adaptation was evaluated using the silicone replica technique. Nonparametric Kruskal-Wallis and Bonferroni tests were performed to determine differences among the groups (P = 0.05).</p><p><strong>Results: </strong>No significant differences were observed in the root mean square (RMS), positive, and negative deviation values between 0° and 45° group (P > 0.05), and the median, first quartile, and third quartile of each were considered clinically acceptable. For the 90° group, the RMS, positive deviation, and negative deviation values were significantly higher than those of the other groups (P < 0.05). The overall adaptation of the 45° group was better than that of the 0° and 90° groups (P < 0.05), whereas the 0° and 90° groups showed no significant difference (P > 0.05).</p><p><strong>Conclusions: </strong>The trueness and adaptation of the intaglio surface of cobalt-chromium maxillary complete denture metal frameworks were affected by the building orientation through SLM. Considering both trueness and adaptation, 45° is recommended as the most efficient build orientation for clinical applications.</p>\",\"PeriodicalId\":16887,\"journal\":{\"name\":\"Journal of prosthodontic research\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":3.2000,\"publicationDate\":\"2025-02-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of prosthodontic research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.2186/jpr.JPR_D_24_00195\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of prosthodontic research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2186/jpr.JPR_D_24_00195","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

目的:评价选择性激光熔化(SLM)制备的全口义齿金属框架在不同方位(0°、45°和90°)下凹形表面的真实度和适应性。方法:采用SLM在0°、45°和90°位置制作钴铬上颌全口义齿金属框架。使用三维检测软件的最佳拟合对齐方法进行正确度测量,并使用硅胶复制技术评估适应性。采用非参数Kruskal-Wallis和Bonferroni检验确定组间差异(P = 0.05)。结果:0°组和45°组的均方根(RMS)、正、负偏差值无显著差异(P < 0.05),中位数、第一四分位数和第三四分位数均为临床可接受。90°组的均方根值、正偏差值、负偏差值均显著高于其他组(P < 0.05)。45°组的整体适应性优于0°和90°组(P < 0.05),而0°和90°组差异不显著(P < 0.05)。结论:SLM对钴铬上颌全口义齿金属框架内凹面的真实性和适应性有影响。考虑到准确性和适应性,推荐45°为临床应用中最有效的构建方向。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Effect of build orientation on trueness and adaptation of intaglio surface of maxillary complete denture metal frameworks fabricated by selective laser melting.

Purpose: We aimed to evaluate the trueness and adaptation of the intaglio surfaces of complete denture metal frameworks fabricated using selective laser melting (SLM) with different building orientations (0°, 45°, and 90°).

Methods: Cobalt-chromium maxillary complete denture metal frameworks were fabricated at 0°, 45°, and 90° using SLM. Trueness measurements were conducted using the best-fit alignment method of three-dimensional inspection software, and adaptation was evaluated using the silicone replica technique. Nonparametric Kruskal-Wallis and Bonferroni tests were performed to determine differences among the groups (P = 0.05).

Results: No significant differences were observed in the root mean square (RMS), positive, and negative deviation values between 0° and 45° group (P > 0.05), and the median, first quartile, and third quartile of each were considered clinically acceptable. For the 90° group, the RMS, positive deviation, and negative deviation values were significantly higher than those of the other groups (P < 0.05). The overall adaptation of the 45° group was better than that of the 0° and 90° groups (P < 0.05), whereas the 0° and 90° groups showed no significant difference (P > 0.05).

Conclusions: The trueness and adaptation of the intaglio surface of cobalt-chromium maxillary complete denture metal frameworks were affected by the building orientation through SLM. Considering both trueness and adaptation, 45° is recommended as the most efficient build orientation for clinical applications.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of prosthodontic research
Journal of prosthodontic research DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
6.90
自引率
11.10%
发文量
161
期刊介绍: Journal of Prosthodontic Research is published 4 times annually, in January, April, July, and October, under supervision by the Editorial Board of Japan Prosthodontic Society, which selects all materials submitted for publication. Journal of Prosthodontic Research originated as an official journal of Japan Prosthodontic Society. It has recently developed a long-range plan to become the most prestigious Asian journal of dental research regarding all aspects of oral and occlusal rehabilitation, fixed/removable prosthodontics, oral implantology and applied oral biology and physiology. The Journal will cover all diagnostic and clinical management aspects necessary to reestablish subjective and objective harmonious oral aesthetics and function. The most-targeted topics: 1) Clinical Epidemiology and Prosthodontics 2) Fixed/Removable Prosthodontics 3) Oral Implantology 4) Prosthodontics-Related Biosciences (Regenerative Medicine, Bone Biology, Mechanobiology, Microbiology/Immunology) 5) Oral Physiology and Biomechanics (Masticating and Swallowing Function, Parafunction, e.g., bruxism) 6) Orofacial Pain and Temporomandibular Disorders (TMDs) 7) Adhesive Dentistry / Dental Materials / Aesthetic Dentistry 8) Maxillofacial Prosthodontics and Dysphagia Rehabilitation 9) Digital Dentistry Prosthodontic treatment may become necessary as a result of developmental or acquired disturbances in the orofacial region, of orofacial trauma, or of a variety of dental and oral diseases and orofacial pain conditions. Reviews, Original articles, technical procedure and case reports can be submitted. Letters to the Editor commenting on papers or any aspect of Journal of Prosthodontic Research are welcomed.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信