用单蠕动泵管进行多通道稀释分析。

IF 2.2 3区 化学 Q2 INSTRUMENTS & INSTRUMENTATION
Robbie M Huff, Willis B Jones, Bradley T Jones
{"title":"用单蠕动泵管进行多通道稀释分析。","authors":"Robbie M Huff, Willis B Jones, Bradley T Jones","doi":"10.1177/00037028251315753","DOIUrl":null,"url":null,"abstract":"<p><p>Multi-channel dilution analysis (MCDA) is a novel, matrix-matched calibration method that automatically dilutes a standard solution using a tubing manifold to split a solution stream traveling between an autosampler and an analytical instrument. The manifold consists of separate tubing channels of various lengths, resulting in a stairstep of signal levels as different portions of the solution reach the instrument to be measured at different points in time. Multi-channel dilution analysis was initially developed on an inductively coupled plasma optical emission spectrometer (ICP-OES) equipped with a peristaltic pump with five channels, with three used to drive solution through individual tubes in the manifold and a fourth reserved for waste. The setup limited MCDA measurements from being realized on instruments that have fewer pump channels without employing an external pump. The use of a single peristaltic pump line to drive solution through the entire tubing manifold is proposed as a feasible way to overcome this hardware limitation and is applied to both ICP-OES and ICP mass spectrometry (ICP-MS). The method is validated through the analysis of certified reference materials, with recoveries for a suite of analytes ranging from 83% to 117%, with relative standard deviations on the order of 1%. The matrix matching capabilities of the method are demonstrated through the analysis of spiked matrices containing high concentrations of ethanol and calcium, and a comparison to traditional calibration techniques. In addition, the improvements in calibration statistics for ICP-MS measurements using MCDA are discussed.</p>","PeriodicalId":8253,"journal":{"name":"Applied Spectroscopy","volume":" ","pages":"37028251315753"},"PeriodicalIF":2.2000,"publicationDate":"2025-02-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Multichannel Dilution Analysis Using a Single Peristaltic Pump Tube.\",\"authors\":\"Robbie M Huff, Willis B Jones, Bradley T Jones\",\"doi\":\"10.1177/00037028251315753\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Multi-channel dilution analysis (MCDA) is a novel, matrix-matched calibration method that automatically dilutes a standard solution using a tubing manifold to split a solution stream traveling between an autosampler and an analytical instrument. The manifold consists of separate tubing channels of various lengths, resulting in a stairstep of signal levels as different portions of the solution reach the instrument to be measured at different points in time. Multi-channel dilution analysis was initially developed on an inductively coupled plasma optical emission spectrometer (ICP-OES) equipped with a peristaltic pump with five channels, with three used to drive solution through individual tubes in the manifold and a fourth reserved for waste. The setup limited MCDA measurements from being realized on instruments that have fewer pump channels without employing an external pump. The use of a single peristaltic pump line to drive solution through the entire tubing manifold is proposed as a feasible way to overcome this hardware limitation and is applied to both ICP-OES and ICP mass spectrometry (ICP-MS). The method is validated through the analysis of certified reference materials, with recoveries for a suite of analytes ranging from 83% to 117%, with relative standard deviations on the order of 1%. The matrix matching capabilities of the method are demonstrated through the analysis of spiked matrices containing high concentrations of ethanol and calcium, and a comparison to traditional calibration techniques. In addition, the improvements in calibration statistics for ICP-MS measurements using MCDA are discussed.</p>\",\"PeriodicalId\":8253,\"journal\":{\"name\":\"Applied Spectroscopy\",\"volume\":\" \",\"pages\":\"37028251315753\"},\"PeriodicalIF\":2.2000,\"publicationDate\":\"2025-02-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Applied Spectroscopy\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://doi.org/10.1177/00037028251315753\",\"RegionNum\":3,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"INSTRUMENTS & INSTRUMENTATION\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Applied Spectroscopy","FirstCategoryId":"92","ListUrlMain":"https://doi.org/10.1177/00037028251315753","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

摘要

多通道稀释分析(MCDA)是一种新型的、矩阵匹配的校准方法,它使用一个管汇来分离在自动进样器和分析仪器之间流动的溶液流,自动稀释标准溶液。歧管由不同长度的独立管道通道组成,当溶液的不同部分在不同时间点到达待测仪器时,导致信号电平的阶梯。多通道稀释分析最初是在电感耦合等离子体光学发射光谱仪(ICP-OES)上开发的,该光谱仪配备了一个带有五个通道的蠕动泵,其中三个用于驱动溶液通过歧管中的单个管,第四个用于废物处理。这种设置限制了在没有外部泵的情况下,在泵通道较少的仪器上实现MCDA测量。使用单一蠕动泵管线驱动溶液通过整个油管歧管是一种可行的方法,可以克服这种硬件限制,并适用于ICP- oes和ICP质谱(ICP- ms)。通过对标准物质的分析验证了该方法的有效性,分析物的回收率为83% ~ 117%,相对标准偏差为1%。通过分析含有高浓度乙醇和钙的加标基质,并与传统校准技术进行比较,证明了该方法的基质匹配能力。此外,还讨论了使用MCDA对ICP-MS测量的校准统计的改进。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Multichannel Dilution Analysis Using a Single Peristaltic Pump Tube.

Multi-channel dilution analysis (MCDA) is a novel, matrix-matched calibration method that automatically dilutes a standard solution using a tubing manifold to split a solution stream traveling between an autosampler and an analytical instrument. The manifold consists of separate tubing channels of various lengths, resulting in a stairstep of signal levels as different portions of the solution reach the instrument to be measured at different points in time. Multi-channel dilution analysis was initially developed on an inductively coupled plasma optical emission spectrometer (ICP-OES) equipped with a peristaltic pump with five channels, with three used to drive solution through individual tubes in the manifold and a fourth reserved for waste. The setup limited MCDA measurements from being realized on instruments that have fewer pump channels without employing an external pump. The use of a single peristaltic pump line to drive solution through the entire tubing manifold is proposed as a feasible way to overcome this hardware limitation and is applied to both ICP-OES and ICP mass spectrometry (ICP-MS). The method is validated through the analysis of certified reference materials, with recoveries for a suite of analytes ranging from 83% to 117%, with relative standard deviations on the order of 1%. The matrix matching capabilities of the method are demonstrated through the analysis of spiked matrices containing high concentrations of ethanol and calcium, and a comparison to traditional calibration techniques. In addition, the improvements in calibration statistics for ICP-MS measurements using MCDA are discussed.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Applied Spectroscopy
Applied Spectroscopy 工程技术-光谱学
CiteScore
6.60
自引率
5.70%
发文量
139
审稿时长
3.5 months
期刊介绍: Applied Spectroscopy is one of the world''s leading spectroscopy journals, publishing high-quality peer-reviewed articles, both fundamental and applied, covering all aspects of spectroscopy. Established in 1951, the journal is owned by the Society for Applied Spectroscopy and is published monthly. The journal is dedicated to fulfilling the mission of the Society to “…advance and disseminate knowledge and information concerning the art and science of spectroscopy and other allied sciences.”
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信