科华北极星c2000全自动生化分析仪内部质量控制的功能评价。

IF 2 4区 医学 Q4 BIOCHEMISTRY & MOLECULAR BIOLOGY
Yiyi Wang, Lei Shi, Jue Zhang, Chen Chen
{"title":"科华北极星c2000全自动生化分析仪内部质量控制的功能评价。","authors":"Yiyi Wang, Lei Shi, Jue Zhang, Chen Chen","doi":"10.5937/jomb0-47151","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The present study aims to (1) evaluate the autoanalyser quality control (QC) module of the Kehua Polaris c2000 automatic modularised biochemical analyser and (2) verify the impact of the analyser's automatic implementation of internal QC (IQC) testing at a set time point on the results of IQC and turnaround time (TAT).</p><p><strong>Methods: </strong>For 5 consecutive days, three different methods were used to conduct IQC. Method 1: Internal QC was carried out at 8:00 every day. Method 2: The QC products were placed in the instrument calibration QC plate in the afternoon, and the instrument was set to measure the products at 7:00 the next day automatically. Method 3: The QC products were placed in the instrument calibration QC plate, and the instrument was set up for automatic measurement at 7:30 every day. All three methods were compared and evaluated. The effect of IQC on the TAT was monitored using method 2.</p><p><strong>Results: </strong>There were no statistical differences in the IQC results between methods 2 and 1. However, there were statistical differences in some items between methods 3 and 1; thus, the IQC results of method 2 can be adopted. Implementing method 2 for IQC can help achieve a significant TAT-saving of 35.23 min during the automatic retest process after the IQC indicates an out-of-control situation. This time reduction is highly valuable for improving efficiency and streamlining the testing workflow.</p><p><strong>Conclusions: </strong>Using the autoanalyser QC module of the Kehua Polaris c2000 automatic modular biochemical analysis system to perform IQC has no impact on the IQC test results and can save TAT, as well as automatically correct most out-of-control occurrences.</p>","PeriodicalId":16175,"journal":{"name":"Journal of Medical Biochemistry","volume":"43 6","pages":"853-859"},"PeriodicalIF":2.0000,"publicationDate":"2024-11-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11773182/pdf/","citationCount":"0","resultStr":"{\"title\":\"Function evaluation of the autoanalyser internal quality control of the Kehua Polaris c2000 automatic biochemical analyser.\",\"authors\":\"Yiyi Wang, Lei Shi, Jue Zhang, Chen Chen\",\"doi\":\"10.5937/jomb0-47151\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The present study aims to (1) evaluate the autoanalyser quality control (QC) module of the Kehua Polaris c2000 automatic modularised biochemical analyser and (2) verify the impact of the analyser's automatic implementation of internal QC (IQC) testing at a set time point on the results of IQC and turnaround time (TAT).</p><p><strong>Methods: </strong>For 5 consecutive days, three different methods were used to conduct IQC. Method 1: Internal QC was carried out at 8:00 every day. Method 2: The QC products were placed in the instrument calibration QC plate in the afternoon, and the instrument was set to measure the products at 7:00 the next day automatically. Method 3: The QC products were placed in the instrument calibration QC plate, and the instrument was set up for automatic measurement at 7:30 every day. All three methods were compared and evaluated. The effect of IQC on the TAT was monitored using method 2.</p><p><strong>Results: </strong>There were no statistical differences in the IQC results between methods 2 and 1. However, there were statistical differences in some items between methods 3 and 1; thus, the IQC results of method 2 can be adopted. Implementing method 2 for IQC can help achieve a significant TAT-saving of 35.23 min during the automatic retest process after the IQC indicates an out-of-control situation. This time reduction is highly valuable for improving efficiency and streamlining the testing workflow.</p><p><strong>Conclusions: </strong>Using the autoanalyser QC module of the Kehua Polaris c2000 automatic modular biochemical analysis system to perform IQC has no impact on the IQC test results and can save TAT, as well as automatically correct most out-of-control occurrences.</p>\",\"PeriodicalId\":16175,\"journal\":{\"name\":\"Journal of Medical Biochemistry\",\"volume\":\"43 6\",\"pages\":\"853-859\"},\"PeriodicalIF\":2.0000,\"publicationDate\":\"2024-11-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11773182/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Biochemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.5937/jomb0-47151\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"BIOCHEMISTRY & MOLECULAR BIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Biochemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5937/jomb0-47151","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOCHEMISTRY & MOLECULAR BIOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景:本研究旨在(1)评估科华北极星c2000自动模块化生化分析仪的自动分析仪质量控制(QC)模块,(2)验证分析仪在设定时间点自动实施内部QC (IQC)测试对IQC结果和周转期(TAT)的影响。方法:连续5天,采用3种不同方法进行IQC。方法一:每天8:00进行内部QC。方法二:下午将QC产品放置在仪器校准QC板上,设置仪器次日7:00自动测量产品。方法三:将QC产品置于仪器校准QC板中,仪器设置每天7:30自动测量。对三种方法进行比较和评价。采用方法2监测IQC对TAT的影响。结果:方法2与方法1的IQC结果无统计学差异。但是,方法3和方法1在一些项目上有统计学差异;因此,可以采用方法2的IQC结果。在IQC中实施方法2可以帮助在IQC发现失控情况后的自动重测过程中节省35.23分钟的tat。这种时间的减少对于提高效率和简化测试工作流程是非常有价值的。结论:使用科华北极星c2000全自动模块化生化分析系统的自动分析仪QC模块进行IQC,对IQC检测结果无影响,可节省TAT,并可自动纠正大部分失控现象。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Function evaluation of the autoanalyser internal quality control of the Kehua Polaris c2000 automatic biochemical analyser.

Background: The present study aims to (1) evaluate the autoanalyser quality control (QC) module of the Kehua Polaris c2000 automatic modularised biochemical analyser and (2) verify the impact of the analyser's automatic implementation of internal QC (IQC) testing at a set time point on the results of IQC and turnaround time (TAT).

Methods: For 5 consecutive days, three different methods were used to conduct IQC. Method 1: Internal QC was carried out at 8:00 every day. Method 2: The QC products were placed in the instrument calibration QC plate in the afternoon, and the instrument was set to measure the products at 7:00 the next day automatically. Method 3: The QC products were placed in the instrument calibration QC plate, and the instrument was set up for automatic measurement at 7:30 every day. All three methods were compared and evaluated. The effect of IQC on the TAT was monitored using method 2.

Results: There were no statistical differences in the IQC results between methods 2 and 1. However, there were statistical differences in some items between methods 3 and 1; thus, the IQC results of method 2 can be adopted. Implementing method 2 for IQC can help achieve a significant TAT-saving of 35.23 min during the automatic retest process after the IQC indicates an out-of-control situation. This time reduction is highly valuable for improving efficiency and streamlining the testing workflow.

Conclusions: Using the autoanalyser QC module of the Kehua Polaris c2000 automatic modular biochemical analysis system to perform IQC has no impact on the IQC test results and can save TAT, as well as automatically correct most out-of-control occurrences.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Biochemistry
Journal of Medical Biochemistry BIOCHEMISTRY & MOLECULAR BIOLOGY-
CiteScore
3.00
自引率
12.00%
发文量
60
审稿时长
>12 weeks
期刊介绍: The JOURNAL OF MEDICAL BIOCHEMISTRY (J MED BIOCHEM) is the official journal of the Society of Medical Biochemists of Serbia with international peer-review. Papers are independently reviewed by at least two reviewers selected by the Editors as Blind Peer Reviews. The Journal of Medical Biochemistry is published quarterly. The Journal publishes original scientific and specialized articles on all aspects of clinical and medical biochemistry, molecular medicine, clinical hematology and coagulation, clinical immunology and autoimmunity, clinical microbiology, virology, clinical genomics and molecular biology, genetic epidemiology, drug measurement, evaluation of diagnostic markers, new reagents and laboratory equipment, reference materials and methods, reference values, laboratory organization, automation, quality control, clinical metrology, all related scientific disciplines where chemistry, biochemistry, molecular biology and immunochemistry deal with the study of normal and pathologic processes in human beings.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信