单能量金属伪影还原算法在CT中的有效性:应用于EVAR和线圈栓塞的增强CT。

Junji Mochizuki, Noriko Oyama-Manabe, Fumi Kato, Hideki Takahashi, Osamu Manabe, Akihiro Sawada, Homare Okamura, Mitsunori Nakano, Atsushi Yamaguchi
{"title":"单能量金属伪影还原算法在CT中的有效性:应用于EVAR和线圈栓塞的增强CT。","authors":"Junji Mochizuki, Noriko Oyama-Manabe, Fumi Kato, Hideki Takahashi, Osamu Manabe, Akihiro Sawada, Homare Okamura, Mitsunori Nakano, Atsushi Yamaguchi","doi":"10.1007/s10554-024-03318-1","DOIUrl":null,"url":null,"abstract":"<p><p>This study aimed to evaluate the efficacy of the single-energy metal artifact reduction (SEMAR) algorithm in reducing metal artifacts and enhancing image quality in contrast-enhanced computed tomography (CT) for patients undergoing endovascular aneurysm repair (EVAR) with coil embolization. Thirty-eight patients (mean age 81.0 ± 6 years; 31 men, 7 women) who underwent contrast-enhanced CT following EVAR and internal iliac artery coil embolization between September 2022 and May 2023 were retrospectively analyzed. The Artifact Index (AI) quantified metallic artifacts from internal iliac artery aneurysm coils in CT images, calculated from the standard deviation of the artifact-containing region relative to a reference region. CT values of the external iliac artery at the same slice were also evaluated and compared. Two radiologists independently performed qualitative assessments of SEMAR and non-SEMAR images. SEMAR significantly reduced metal artifacts, decreasing the AI from 171.9 ± 74.5 HU to 35.8 ± 16.9 HU (p < 0.001). The mean CT values of the external iliac artery were similar for SEMAR (259.4 ± 63.7 HU) and non-SEMAR (257.1 ± 63.6 HU, indicating no significant difference. Qualitative assessment scores improved significantly with SEMAR (from 1.0 ± 0.0 to 2.5 ± 0.5; p < 0.001), enhancing visualization of internal iliac artery aneurysms. Interobserver agreement was high (κ = 0.83). The SEMAR algorithm effectively reduces metal artifacts in contrast-enhanced CT, significantly enhancing image quality without altering adjacent artery CT values. These improvements enhance the image quality of post-operative assessments in patients undergoing EVAR with coil embolization.</p>","PeriodicalId":94227,"journal":{"name":"The international journal of cardiovascular imaging","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-01-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effectiveness of single-energy metal artifact reduction algorithm in CT: application to contrast-enhanced CT with EVAR and coil embolization.\",\"authors\":\"Junji Mochizuki, Noriko Oyama-Manabe, Fumi Kato, Hideki Takahashi, Osamu Manabe, Akihiro Sawada, Homare Okamura, Mitsunori Nakano, Atsushi Yamaguchi\",\"doi\":\"10.1007/s10554-024-03318-1\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>This study aimed to evaluate the efficacy of the single-energy metal artifact reduction (SEMAR) algorithm in reducing metal artifacts and enhancing image quality in contrast-enhanced computed tomography (CT) for patients undergoing endovascular aneurysm repair (EVAR) with coil embolization. Thirty-eight patients (mean age 81.0 ± 6 years; 31 men, 7 women) who underwent contrast-enhanced CT following EVAR and internal iliac artery coil embolization between September 2022 and May 2023 were retrospectively analyzed. The Artifact Index (AI) quantified metallic artifacts from internal iliac artery aneurysm coils in CT images, calculated from the standard deviation of the artifact-containing region relative to a reference region. CT values of the external iliac artery at the same slice were also evaluated and compared. Two radiologists independently performed qualitative assessments of SEMAR and non-SEMAR images. SEMAR significantly reduced metal artifacts, decreasing the AI from 171.9 ± 74.5 HU to 35.8 ± 16.9 HU (p < 0.001). The mean CT values of the external iliac artery were similar for SEMAR (259.4 ± 63.7 HU) and non-SEMAR (257.1 ± 63.6 HU, indicating no significant difference. Qualitative assessment scores improved significantly with SEMAR (from 1.0 ± 0.0 to 2.5 ± 0.5; p < 0.001), enhancing visualization of internal iliac artery aneurysms. Interobserver agreement was high (κ = 0.83). The SEMAR algorithm effectively reduces metal artifacts in contrast-enhanced CT, significantly enhancing image quality without altering adjacent artery CT values. These improvements enhance the image quality of post-operative assessments in patients undergoing EVAR with coil embolization.</p>\",\"PeriodicalId\":94227,\"journal\":{\"name\":\"The international journal of cardiovascular imaging\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-01-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The international journal of cardiovascular imaging\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s10554-024-03318-1\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The international journal of cardiovascular imaging","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s10554-024-03318-1","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本研究旨在评估单能量金属伪影减少(SEMAR)算法在血管内动脉瘤修复(EVAR)合并线圈栓塞患者中减少金属伪影和提高对比度增强计算机断层扫描(CT)图像质量的效果。38例患者(平均年龄81.0±6岁;回顾性分析了2022年9月至2023年5月期间,31名男性(7名女性)在EVAR和髂内动脉圈栓塞后接受对比增强CT检查的患者。伪影指数(artificial Index, AI)量化了CT图像中髂内动脉瘤线圈的金属伪影,由伪影包含区域相对于参考区域的标准差计算得出。并对同一层的髂外动脉CT值进行评价和比较。两名放射科医生独立地对SEMAR和非SEMAR图像进行定性评估。SEMAR显著降低了金属伪影,将AI从171.9±74.5 HU降低到35.8±16.9 HU (p
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Effectiveness of single-energy metal artifact reduction algorithm in CT: application to contrast-enhanced CT with EVAR and coil embolization.

This study aimed to evaluate the efficacy of the single-energy metal artifact reduction (SEMAR) algorithm in reducing metal artifacts and enhancing image quality in contrast-enhanced computed tomography (CT) for patients undergoing endovascular aneurysm repair (EVAR) with coil embolization. Thirty-eight patients (mean age 81.0 ± 6 years; 31 men, 7 women) who underwent contrast-enhanced CT following EVAR and internal iliac artery coil embolization between September 2022 and May 2023 were retrospectively analyzed. The Artifact Index (AI) quantified metallic artifacts from internal iliac artery aneurysm coils in CT images, calculated from the standard deviation of the artifact-containing region relative to a reference region. CT values of the external iliac artery at the same slice were also evaluated and compared. Two radiologists independently performed qualitative assessments of SEMAR and non-SEMAR images. SEMAR significantly reduced metal artifacts, decreasing the AI from 171.9 ± 74.5 HU to 35.8 ± 16.9 HU (p < 0.001). The mean CT values of the external iliac artery were similar for SEMAR (259.4 ± 63.7 HU) and non-SEMAR (257.1 ± 63.6 HU, indicating no significant difference. Qualitative assessment scores improved significantly with SEMAR (from 1.0 ± 0.0 to 2.5 ± 0.5; p < 0.001), enhancing visualization of internal iliac artery aneurysms. Interobserver agreement was high (κ = 0.83). The SEMAR algorithm effectively reduces metal artifacts in contrast-enhanced CT, significantly enhancing image quality without altering adjacent artery CT values. These improvements enhance the image quality of post-operative assessments in patients undergoing EVAR with coil embolization.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信