{"title":"基于fpga的64通道读出电子器件用于高分辨率TOF-PET探测器","authors":"Xiang Zhang;Yonggang Wang;Mingchen Wang;Xiaoguang Kong","doi":"10.1109/TRPMS.2024.3443831","DOIUrl":null,"url":null,"abstract":"Field programmable logic array (FPGA)-based readout electronics has shown its capability of channel-by-channel signal readout for time-of-flight positron emission tomography (TOF-PET) detectors. However, for detectors that rely on light sharing to achieve subpixel resolution, the high-linear measurement dynamic range of the readout electronics is highly required. In this article, the problems with dynamic range in our previously proposed FPGA-based fast linear discharge circuit are investigated and corresponding methods are proposed to enhance its small signal measurement capability and improve the timing performance as well. A practical 64-channel TOF-PET detector module was constructed and evaluated. The readout electronics test results demonstrated a 240x measurement dynamic range with 99.5% conversion linearity. In the case that the \n<inline-formula> <tex-math>$8\\times 8$ </tex-math></inline-formula>\n silicon photomultiplier (SiPM) array in the detector combines with an \n<inline-formula> <tex-math>$8\\times 8$ </tex-math></inline-formula>\n LYSO crystal (each \n<inline-formula> <tex-math>$3.2\\times 3.2\\times 10$ </tex-math></inline-formula>\n mm3) array, the average energy and coincidence time resolution of the detector are measured as 10.68% (511 keV) and 364.9 ps, respectively. To demonstrate the benefit of large dynamic range to high-resolution detectors, the crystal array in the detector was replaced by a \n<inline-formula> <tex-math>$24\\times 24$ </tex-math></inline-formula>\n LYSO array (each \n<inline-formula> <tex-math>$1.04\\times 1.04\\times 15$ </tex-math></inline-formula>\n mm3) and achieved 1-mm resolution. The test results confirm that the proposed FPGA-based readout circuit is practical for laboratory instrumentation","PeriodicalId":46807,"journal":{"name":"IEEE Transactions on Radiation and Plasma Medical Sciences","volume":"9 1","pages":"11-19"},"PeriodicalIF":4.6000,"publicationDate":"2024-08-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"An FPGA-Based 64-Channel Readout Electronics for High-Resolution TOF-PET Detectors\",\"authors\":\"Xiang Zhang;Yonggang Wang;Mingchen Wang;Xiaoguang Kong\",\"doi\":\"10.1109/TRPMS.2024.3443831\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Field programmable logic array (FPGA)-based readout electronics has shown its capability of channel-by-channel signal readout for time-of-flight positron emission tomography (TOF-PET) detectors. However, for detectors that rely on light sharing to achieve subpixel resolution, the high-linear measurement dynamic range of the readout electronics is highly required. In this article, the problems with dynamic range in our previously proposed FPGA-based fast linear discharge circuit are investigated and corresponding methods are proposed to enhance its small signal measurement capability and improve the timing performance as well. A practical 64-channel TOF-PET detector module was constructed and evaluated. The readout electronics test results demonstrated a 240x measurement dynamic range with 99.5% conversion linearity. In the case that the \\n<inline-formula> <tex-math>$8\\\\times 8$ </tex-math></inline-formula>\\n silicon photomultiplier (SiPM) array in the detector combines with an \\n<inline-formula> <tex-math>$8\\\\times 8$ </tex-math></inline-formula>\\n LYSO crystal (each \\n<inline-formula> <tex-math>$3.2\\\\times 3.2\\\\times 10$ </tex-math></inline-formula>\\n mm3) array, the average energy and coincidence time resolution of the detector are measured as 10.68% (511 keV) and 364.9 ps, respectively. To demonstrate the benefit of large dynamic range to high-resolution detectors, the crystal array in the detector was replaced by a \\n<inline-formula> <tex-math>$24\\\\times 24$ </tex-math></inline-formula>\\n LYSO array (each \\n<inline-formula> <tex-math>$1.04\\\\times 1.04\\\\times 15$ </tex-math></inline-formula>\\n mm3) and achieved 1-mm resolution. The test results confirm that the proposed FPGA-based readout circuit is practical for laboratory instrumentation\",\"PeriodicalId\":46807,\"journal\":{\"name\":\"IEEE Transactions on Radiation and Plasma Medical Sciences\",\"volume\":\"9 1\",\"pages\":\"11-19\"},\"PeriodicalIF\":4.6000,\"publicationDate\":\"2024-08-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IEEE Transactions on Radiation and Plasma Medical Sciences\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://ieeexplore.ieee.org/document/10637268/\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Radiation and Plasma Medical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/10637268/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
An FPGA-Based 64-Channel Readout Electronics for High-Resolution TOF-PET Detectors
Field programmable logic array (FPGA)-based readout electronics has shown its capability of channel-by-channel signal readout for time-of-flight positron emission tomography (TOF-PET) detectors. However, for detectors that rely on light sharing to achieve subpixel resolution, the high-linear measurement dynamic range of the readout electronics is highly required. In this article, the problems with dynamic range in our previously proposed FPGA-based fast linear discharge circuit are investigated and corresponding methods are proposed to enhance its small signal measurement capability and improve the timing performance as well. A practical 64-channel TOF-PET detector module was constructed and evaluated. The readout electronics test results demonstrated a 240x measurement dynamic range with 99.5% conversion linearity. In the case that the
$8\times 8$
silicon photomultiplier (SiPM) array in the detector combines with an
$8\times 8$
LYSO crystal (each
$3.2\times 3.2\times 10$
mm3) array, the average energy and coincidence time resolution of the detector are measured as 10.68% (511 keV) and 364.9 ps, respectively. To demonstrate the benefit of large dynamic range to high-resolution detectors, the crystal array in the detector was replaced by a
$24\times 24$
LYSO array (each
$1.04\times 1.04\times 15$
mm3) and achieved 1-mm resolution. The test results confirm that the proposed FPGA-based readout circuit is practical for laboratory instrumentation