使用自动质谱系统区分多发性骨髓瘤患者血清中的Daratumumab和内源性单克隆蛋白

IF 1.8 Q3 MEDICAL LABORATORY TECHNOLOGY
David Barnidge, Dhananjay Sakrikar, Tadeusz Kubicki, Benjamin A Derman, Andrzej J Jakubowiak, Gabriella Lakos
{"title":"使用自动质谱系统区分多发性骨髓瘤患者血清中的Daratumumab和内源性单克隆蛋白","authors":"David Barnidge, Dhananjay Sakrikar, Tadeusz Kubicki, Benjamin A Derman, Andrzej J Jakubowiak, Gabriella Lakos","doi":"10.1093/jalm/jfae142","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Therapeutic monoclonal antibodies (t-mAbs) may interfere with electrophoresis-based methods used to monitor multiple myeloma (MM), which can create inaccurate results. Matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry is an alternative to gels distinguishing between endogenous M-proteins and t-mAbs based on molecular mass.</p><p><strong>Methods: </strong>Serum samples (n = 109) from 34 MM patients receiving Dara-KRd were collected 14 or 28 days postdaratumumab administration. Samples were analyzed using the EXENT® Analyzer that combines automated immunopurification and MALDI-TOF MS for the isotyping and quantification of monoclonal immunoglobulins.</p><p><strong>Results: </strong>Daratumumab was identified in 103 out of 109 samples (94.5%). In all IgGλ (n = 8), IgAκ (n = 8), and IgAλ (n = 2) patients, the M-protein and daratumumab were detected. Of the IgGκ patients (n = 18), 5 patients had a total of 6 samples where the M-protein was detected but daratumumab was not. There was no difference in the detection rate of daratumumab between samples taken 14 and 28 days postadministration with the median daratumumab concentration being 0.95 and 0.54 g/L, respectively. A precision study was also performed on 25 replicates containing 1 g/L daratumumab in serum where a coefficient of variation of 4.2% was observed as determined by the EXENT Analyzer.</p><p><strong>Conclusions: </strong>The Immunoglobulin Isotypes (GAM: IgG, IgA, and IgM) for the EXENT Analyzer detected and distinguished a daratumumab kappa light chain peak from an M-protein light chain peak in MM patient serum when resolved by the mass spectrometer.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2024-12-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Distinguishing Daratumumab from Endogenous Monoclonal Proteins in Serum from Multiple Myeloma Patients Using an Automated Mass Spectrometry System.\",\"authors\":\"David Barnidge, Dhananjay Sakrikar, Tadeusz Kubicki, Benjamin A Derman, Andrzej J Jakubowiak, Gabriella Lakos\",\"doi\":\"10.1093/jalm/jfae142\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Therapeutic monoclonal antibodies (t-mAbs) may interfere with electrophoresis-based methods used to monitor multiple myeloma (MM), which can create inaccurate results. Matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry is an alternative to gels distinguishing between endogenous M-proteins and t-mAbs based on molecular mass.</p><p><strong>Methods: </strong>Serum samples (n = 109) from 34 MM patients receiving Dara-KRd were collected 14 or 28 days postdaratumumab administration. Samples were analyzed using the EXENT® Analyzer that combines automated immunopurification and MALDI-TOF MS for the isotyping and quantification of monoclonal immunoglobulins.</p><p><strong>Results: </strong>Daratumumab was identified in 103 out of 109 samples (94.5%). In all IgGλ (n = 8), IgAκ (n = 8), and IgAλ (n = 2) patients, the M-protein and daratumumab were detected. Of the IgGκ patients (n = 18), 5 patients had a total of 6 samples where the M-protein was detected but daratumumab was not. There was no difference in the detection rate of daratumumab between samples taken 14 and 28 days postadministration with the median daratumumab concentration being 0.95 and 0.54 g/L, respectively. A precision study was also performed on 25 replicates containing 1 g/L daratumumab in serum where a coefficient of variation of 4.2% was observed as determined by the EXENT Analyzer.</p><p><strong>Conclusions: </strong>The Immunoglobulin Isotypes (GAM: IgG, IgA, and IgM) for the EXENT Analyzer detected and distinguished a daratumumab kappa light chain peak from an M-protein light chain peak in MM patient serum when resolved by the mass spectrometer.</p>\",\"PeriodicalId\":46361,\"journal\":{\"name\":\"Journal of Applied Laboratory Medicine\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2024-12-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Applied Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1093/jalm/jfae142\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/jalm/jfae142","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景:治疗性单克隆抗体(t- mab)可能会干扰用于监测多发性骨髓瘤(MM)的基于电泳的方法,从而产生不准确的结果。基质辅助激光解吸/电离飞行时间(MALDI-TOF)质谱法是基于分子质量区分内源性m蛋白和t单抗的一种替代方法。方法:34例接受达拉- krd治疗的MM患者的血清样本(n = 109)在给予达拉单抗后14或28天采集。使用event®分析仪对样品进行分析,该分析仪结合了自动免疫纯化和MALDI-TOF质谱,用于单克隆免疫球蛋白的等型和定量分析。结果:109个样本中有103个(94.5%)被鉴定出Daratumumab。在所有IgGλ (n = 8)、igakk (n = 8)和igab λ (n = 2)患者中,检测m蛋白和达拉单抗。在IgGκ患者(n = 18)中,5例患者总共有6个样品检测到m蛋白,但达拉单抗未检测到。给药后14天和28天的样品中,达拉图单抗的检出率无差异,达拉图单抗中位浓度分别为0.95和0.54 g/L。还对血清中含有1 g/L daratumumab的25个重复进行了精度研究,其中变异系数为4.2%,由event分析仪测定。结论:EXENT分析仪的免疫球蛋白同型(GAM: IgG、IgA和IgM)通过质谱分析,在MM患者血清中检测并区分了daratumumab kappa轻链峰和m蛋白轻链峰。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Distinguishing Daratumumab from Endogenous Monoclonal Proteins in Serum from Multiple Myeloma Patients Using an Automated Mass Spectrometry System.

Background: Therapeutic monoclonal antibodies (t-mAbs) may interfere with electrophoresis-based methods used to monitor multiple myeloma (MM), which can create inaccurate results. Matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry is an alternative to gels distinguishing between endogenous M-proteins and t-mAbs based on molecular mass.

Methods: Serum samples (n = 109) from 34 MM patients receiving Dara-KRd were collected 14 or 28 days postdaratumumab administration. Samples were analyzed using the EXENT® Analyzer that combines automated immunopurification and MALDI-TOF MS for the isotyping and quantification of monoclonal immunoglobulins.

Results: Daratumumab was identified in 103 out of 109 samples (94.5%). In all IgGλ (n = 8), IgAκ (n = 8), and IgAλ (n = 2) patients, the M-protein and daratumumab were detected. Of the IgGκ patients (n = 18), 5 patients had a total of 6 samples where the M-protein was detected but daratumumab was not. There was no difference in the detection rate of daratumumab between samples taken 14 and 28 days postadministration with the median daratumumab concentration being 0.95 and 0.54 g/L, respectively. A precision study was also performed on 25 replicates containing 1 g/L daratumumab in serum where a coefficient of variation of 4.2% was observed as determined by the EXENT Analyzer.

Conclusions: The Immunoglobulin Isotypes (GAM: IgG, IgA, and IgM) for the EXENT Analyzer detected and distinguished a daratumumab kappa light chain peak from an M-protein light chain peak in MM patient serum when resolved by the mass spectrometer.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Applied Laboratory Medicine
Journal of Applied Laboratory Medicine MEDICAL LABORATORY TECHNOLOGY-
CiteScore
3.70
自引率
5.00%
发文量
137
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信