采用适应性心脏再同步化治疗算法编程的患者的电非同步化映射和无反应优化。

IF 5.6 2区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS
Alan J Bank, Kevin V Burns, Christopher D Brown, Evan Walser-Kuntz, Madeline A Czeck, Robert G Hauser, Jay D Sengupta
{"title":"采用适应性心脏再同步化治疗算法编程的患者的电非同步化映射和无反应优化。","authors":"Alan J Bank, Kevin V Burns, Christopher D Brown, Evan Walser-Kuntz, Madeline A Czeck, Robert G Hauser, Jay D Sengupta","doi":"10.1016/j.hrthm.2024.12.012","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The adaptive cardiac resynchronization therapy (CRT) (aCRT) algorithm provides an important clinical benefit. However, a significant number of patients are nonresponders.</p><p><strong>Objectives: </strong>The goals of this study were to quantify electrical synchrony in patients programmed with aCRT and to assess the echocardiographic effects of optimization in CRT nonresponders and incomplete responders.</p><p><strong>Methods: </strong>We studied 125 patients programmed with aCRT and measured electrical synchrony at multiple device settings using novel electrical dyssynchrony mapping (EDM) technology. Electrical synchrony was quantified as cardiac resynchronization index (CRI), a measure that analyzes areas between multiple pairs of anterior and posterior electrograms and calculates synchrony normalized to native rhythm.</p><p><strong>Results: </strong>CRI improved from baseline aCRT settings to optimal settings on the basis of EDM (56%±29% vs 92%±12%; P<.001). Patients programmed with left ventricle (LV)-only aCRT (group 1, n=68 [54%]) had a higher CRI (62%±25% vs 48%±31%; P=.014) than did patients programmed with biventricular aCRT (group 2, n=57 [46%]). In group 1 and group 2, optimal CRI during sequential biventricular (92%±13% and 93%±9%, respectively) and LV-only (92%±6% and 91%±7%, respectively) pacing was significantly (P<.001) higher than CRI at baseline aCRT setting. In a subset of 53 nonresponders optimized using EDM, there were significant improvements in CRI (37%±25%; P<.0001), LV ejection fraction (6.2%±6.6%; P<.0001), end-diastolic volume (9.5±28.2 mL; P=.015), end-systolic volume (13.4±24.9 mL; P<.001), and transverse (1.5%±4.4%; P=.014), longitudinal (1.0%±2.5%; P=.003), and circumferential (2.6%±8.5%; P=.047) strain.</p><p><strong>Conclusion: </strong>Electrical synchrony improves 56% with CRT using aCRT programming and 92% with EDM optimization. Optimization of aCRT-programmed nonresponders results in significant improvements in LV size and systolic function, offering the possibility of converting CRT nonresponders into responders.</p>","PeriodicalId":12886,"journal":{"name":"Heart rhythm","volume":" ","pages":""},"PeriodicalIF":5.6000,"publicationDate":"2024-12-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Electrical dyssynchrony mapping and optimization of nonresponders in patients programmed with the adaptive cardiac resynchronization therapy algorithm.\",\"authors\":\"Alan J Bank, Kevin V Burns, Christopher D Brown, Evan Walser-Kuntz, Madeline A Czeck, Robert G Hauser, Jay D Sengupta\",\"doi\":\"10.1016/j.hrthm.2024.12.012\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The adaptive cardiac resynchronization therapy (CRT) (aCRT) algorithm provides an important clinical benefit. However, a significant number of patients are nonresponders.</p><p><strong>Objectives: </strong>The goals of this study were to quantify electrical synchrony in patients programmed with aCRT and to assess the echocardiographic effects of optimization in CRT nonresponders and incomplete responders.</p><p><strong>Methods: </strong>We studied 125 patients programmed with aCRT and measured electrical synchrony at multiple device settings using novel electrical dyssynchrony mapping (EDM) technology. Electrical synchrony was quantified as cardiac resynchronization index (CRI), a measure that analyzes areas between multiple pairs of anterior and posterior electrograms and calculates synchrony normalized to native rhythm.</p><p><strong>Results: </strong>CRI improved from baseline aCRT settings to optimal settings on the basis of EDM (56%±29% vs 92%±12%; P<.001). Patients programmed with left ventricle (LV)-only aCRT (group 1, n=68 [54%]) had a higher CRI (62%±25% vs 48%±31%; P=.014) than did patients programmed with biventricular aCRT (group 2, n=57 [46%]). In group 1 and group 2, optimal CRI during sequential biventricular (92%±13% and 93%±9%, respectively) and LV-only (92%±6% and 91%±7%, respectively) pacing was significantly (P<.001) higher than CRI at baseline aCRT setting. In a subset of 53 nonresponders optimized using EDM, there were significant improvements in CRI (37%±25%; P<.0001), LV ejection fraction (6.2%±6.6%; P<.0001), end-diastolic volume (9.5±28.2 mL; P=.015), end-systolic volume (13.4±24.9 mL; P<.001), and transverse (1.5%±4.4%; P=.014), longitudinal (1.0%±2.5%; P=.003), and circumferential (2.6%±8.5%; P=.047) strain.</p><p><strong>Conclusion: </strong>Electrical synchrony improves 56% with CRT using aCRT programming and 92% with EDM optimization. Optimization of aCRT-programmed nonresponders results in significant improvements in LV size and systolic function, offering the possibility of converting CRT nonresponders into responders.</p>\",\"PeriodicalId\":12886,\"journal\":{\"name\":\"Heart rhythm\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":5.6000,\"publicationDate\":\"2024-12-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Heart rhythm\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1016/j.hrthm.2024.12.012\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Heart rhythm","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.hrthm.2024.12.012","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

背景:适应性心脏再同步化治疗(CRT)算法(aCRT)提供了重要的临床效益。然而,相当数量的患者无反应。目的:本研究的目的是量化aCRT患者的电同步,并评估优化对无反应者的影响。方法:我们研究了125例使用aCRT编程的患者,并使用新型电非同步映射(EDM)技术在多个设备设置下测量电同步。电同步被量化为心脏再同步指数(CRI),这是一种分析多对前后电图之间区域的测量方法,并计算同步归一化到自然节律。结果:CRI从基线aCRT设置提高到基于EDM的最佳设置(56±29% vs 92±12%)。结论:使用aCRT编程的CRT提高56%,EDM优化提高92%。优化acrt编程无应答者可显著改善左室大小和收缩功能,为将CRT无应答者转化为应答者提供了可能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Electrical dyssynchrony mapping and optimization of nonresponders in patients programmed with the adaptive cardiac resynchronization therapy algorithm.

Background: The adaptive cardiac resynchronization therapy (CRT) (aCRT) algorithm provides an important clinical benefit. However, a significant number of patients are nonresponders.

Objectives: The goals of this study were to quantify electrical synchrony in patients programmed with aCRT and to assess the echocardiographic effects of optimization in CRT nonresponders and incomplete responders.

Methods: We studied 125 patients programmed with aCRT and measured electrical synchrony at multiple device settings using novel electrical dyssynchrony mapping (EDM) technology. Electrical synchrony was quantified as cardiac resynchronization index (CRI), a measure that analyzes areas between multiple pairs of anterior and posterior electrograms and calculates synchrony normalized to native rhythm.

Results: CRI improved from baseline aCRT settings to optimal settings on the basis of EDM (56%±29% vs 92%±12%; P<.001). Patients programmed with left ventricle (LV)-only aCRT (group 1, n=68 [54%]) had a higher CRI (62%±25% vs 48%±31%; P=.014) than did patients programmed with biventricular aCRT (group 2, n=57 [46%]). In group 1 and group 2, optimal CRI during sequential biventricular (92%±13% and 93%±9%, respectively) and LV-only (92%±6% and 91%±7%, respectively) pacing was significantly (P<.001) higher than CRI at baseline aCRT setting. In a subset of 53 nonresponders optimized using EDM, there were significant improvements in CRI (37%±25%; P<.0001), LV ejection fraction (6.2%±6.6%; P<.0001), end-diastolic volume (9.5±28.2 mL; P=.015), end-systolic volume (13.4±24.9 mL; P<.001), and transverse (1.5%±4.4%; P=.014), longitudinal (1.0%±2.5%; P=.003), and circumferential (2.6%±8.5%; P=.047) strain.

Conclusion: Electrical synchrony improves 56% with CRT using aCRT programming and 92% with EDM optimization. Optimization of aCRT-programmed nonresponders results in significant improvements in LV size and systolic function, offering the possibility of converting CRT nonresponders into responders.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Heart rhythm
Heart rhythm 医学-心血管系统
CiteScore
10.50
自引率
5.50%
发文量
1465
审稿时长
24 days
期刊介绍: HeartRhythm, the official Journal of the Heart Rhythm Society and the Cardiac Electrophysiology Society, is a unique journal for fundamental discovery and clinical applicability. HeartRhythm integrates the entire cardiac electrophysiology (EP) community from basic and clinical academic researchers, private practitioners, engineers, allied professionals, industry, and trainees, all of whom are vital and interdependent members of our EP community. The Heart Rhythm Society is the international leader in science, education, and advocacy for cardiac arrhythmia professionals and patients, and the primary information resource on heart rhythm disorders. Its mission is to improve the care of patients by promoting research, education, and optimal health care policies and standards.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信