{"title":"自动降钙素原测定的临床和分析性能评价。","authors":"Godwin Ogbonna, Jodiann R Atienza, David W Singleton, Andrea Ott-Vasconi, Stacey A Alvey","doi":"10.1093/jalm/jfae114","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Procalcitonin (PCT) measurement is useful for guiding antibiotic therapy and risk assessment in lower respiratory infections and/or sepsis. This study evaluated clinical and analytical performance of the Vitros® Immunodiagnostic Products B·R·A·H·M·S PCT assay (Vitros PCT).</p><p><strong>Methods: </strong>Precision, limits of blank (LoB), detection (LoD), and quantitation (LoQ) were determined for Vitros PCT, along with method comparison and clinical concordance with the B·R·A·H·M·S PCT™-sensitive KRYPTOR™ assay (KRYPTOR PCT). All-cause 28-day mortality was evaluated according to the change in PCT values (ΔPCT) from day 0 through day 4 in samples from 598 intensive care unit patients with sepsis.</p><p><strong>Results: </strong>Comparison of Vitros PCT and KRYPTOR PCT results yielded a Deming regression slope of 1.057, intercept of -0.010, and correlation coefficient (r) of 0.994. Precision analysis demonstrated within-laboratory coefficients of variation for Vitros PCT ranging from 3.1% to 6.4%. The LoD and observed LoQ were determined as 0.007 and 0.013 ng/mL, respectively. Overall agreement between assay methods was 98.5%, 98.0%, 97.4%, and 97.8%, at PCT clinical decision cutoffs of 0.100, 0.250, 0.500, and 2.00 ng/mL, respectively, with Cohen's Kappa coefficients (κ) > 0.91. ΔPCT values ≤80% vs >80% were associated with increased 28-day-all-cause mortality (P = 0.006).</p><p><strong>Conclusions: </strong>Vitros PCT compares well with KRYPTOR PCT, showing excellent agreement at relevant clinical decision cutoffs that have been used for antibiotic decision-making and assessment of risk for sepsis progression. ΔPCT values determined with Vitros PCT were useful for evaluation of 28-day mortality risk in patients with severe sepsis.</p>","PeriodicalId":46361,"journal":{"name":"Journal of Applied Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2024-12-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Clinical and Analytical Performance Evaluation of an Automated Procalcitonin Assay.\",\"authors\":\"Godwin Ogbonna, Jodiann R Atienza, David W Singleton, Andrea Ott-Vasconi, Stacey A Alvey\",\"doi\":\"10.1093/jalm/jfae114\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Procalcitonin (PCT) measurement is useful for guiding antibiotic therapy and risk assessment in lower respiratory infections and/or sepsis. This study evaluated clinical and analytical performance of the Vitros® Immunodiagnostic Products B·R·A·H·M·S PCT assay (Vitros PCT).</p><p><strong>Methods: </strong>Precision, limits of blank (LoB), detection (LoD), and quantitation (LoQ) were determined for Vitros PCT, along with method comparison and clinical concordance with the B·R·A·H·M·S PCT™-sensitive KRYPTOR™ assay (KRYPTOR PCT). All-cause 28-day mortality was evaluated according to the change in PCT values (ΔPCT) from day 0 through day 4 in samples from 598 intensive care unit patients with sepsis.</p><p><strong>Results: </strong>Comparison of Vitros PCT and KRYPTOR PCT results yielded a Deming regression slope of 1.057, intercept of -0.010, and correlation coefficient (r) of 0.994. Precision analysis demonstrated within-laboratory coefficients of variation for Vitros PCT ranging from 3.1% to 6.4%. The LoD and observed LoQ were determined as 0.007 and 0.013 ng/mL, respectively. Overall agreement between assay methods was 98.5%, 98.0%, 97.4%, and 97.8%, at PCT clinical decision cutoffs of 0.100, 0.250, 0.500, and 2.00 ng/mL, respectively, with Cohen's Kappa coefficients (κ) > 0.91. ΔPCT values ≤80% vs >80% were associated with increased 28-day-all-cause mortality (P = 0.006).</p><p><strong>Conclusions: </strong>Vitros PCT compares well with KRYPTOR PCT, showing excellent agreement at relevant clinical decision cutoffs that have been used for antibiotic decision-making and assessment of risk for sepsis progression. ΔPCT values determined with Vitros PCT were useful for evaluation of 28-day mortality risk in patients with severe sepsis.</p>\",\"PeriodicalId\":46361,\"journal\":{\"name\":\"Journal of Applied Laboratory Medicine\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2024-12-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Applied Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1093/jalm/jfae114\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/jalm/jfae114","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
Clinical and Analytical Performance Evaluation of an Automated Procalcitonin Assay.
Background: Procalcitonin (PCT) measurement is useful for guiding antibiotic therapy and risk assessment in lower respiratory infections and/or sepsis. This study evaluated clinical and analytical performance of the Vitros® Immunodiagnostic Products B·R·A·H·M·S PCT assay (Vitros PCT).
Methods: Precision, limits of blank (LoB), detection (LoD), and quantitation (LoQ) were determined for Vitros PCT, along with method comparison and clinical concordance with the B·R·A·H·M·S PCT™-sensitive KRYPTOR™ assay (KRYPTOR PCT). All-cause 28-day mortality was evaluated according to the change in PCT values (ΔPCT) from day 0 through day 4 in samples from 598 intensive care unit patients with sepsis.
Results: Comparison of Vitros PCT and KRYPTOR PCT results yielded a Deming regression slope of 1.057, intercept of -0.010, and correlation coefficient (r) of 0.994. Precision analysis demonstrated within-laboratory coefficients of variation for Vitros PCT ranging from 3.1% to 6.4%. The LoD and observed LoQ were determined as 0.007 and 0.013 ng/mL, respectively. Overall agreement between assay methods was 98.5%, 98.0%, 97.4%, and 97.8%, at PCT clinical decision cutoffs of 0.100, 0.250, 0.500, and 2.00 ng/mL, respectively, with Cohen's Kappa coefficients (κ) > 0.91. ΔPCT values ≤80% vs >80% were associated with increased 28-day-all-cause mortality (P = 0.006).
Conclusions: Vitros PCT compares well with KRYPTOR PCT, showing excellent agreement at relevant clinical decision cutoffs that have been used for antibiotic decision-making and assessment of risk for sepsis progression. ΔPCT values determined with Vitros PCT were useful for evaluation of 28-day mortality risk in patients with severe sepsis.