左颈内静脉入路至右心房附件基底植入Aveir AR无铅起搏器。

Q3 Medicine
Howard Liu, Daniel Cortez
{"title":"左颈内静脉入路至右心房附件基底植入Aveir AR无铅起搏器。","authors":"Howard Liu,&nbsp;Daniel Cortez","doi":"10.1016/j.ipej.2024.11.004","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><div>There is no prior report of an Aveir leadless pacemaker implantation into the atrial appendage via the internal jugular vein.</div></div><div><h3>Case</h3><div>A 44-year-old female patient with history of multiple ablations for sinus node dysfunction presented with symptomatic bradycardia. The patient had femoral veins &lt;9mm, chronic pain at the femoral vein insertion sites, as well as a recent car accident with persistent leg pain due to femoral fractures. Placement of a leadless pacemaker was decided based on patient discretion. An Abbott Aveir AR leadless pacemaker was implanted via left internal jugular vein access without complication. Post device implantation showed threshold of 0.75V @0.4 ms, impedance of 340 Ω, Pwave of 4.4 mV. Six-month follow-up demonstrated a threshold of 0.5 V @0.2 milliseconds, impedance of 300 Ω and Pwave of 7.2 mV with 92 % pacing and predicted longevity of 12.7 years.</div></div><div><h3>Discussion</h3><div>The follow-up showed no complications in the patient. A similar approach may be feasible for other patients needing atrial leadless pacing, in which the transfemoral approach is not preferred.</div></div><div><h3>Conclusion</h3><div>Implantation of the Aveir AR leadless pacemaker into the right atrial appendage is feasible via the left internal jugular vein.</div></div>","PeriodicalId":35900,"journal":{"name":"Indian Pacing and Electrophysiology Journal","volume":"25 1","pages":"Pages 25-27"},"PeriodicalIF":0.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Left internal jugular vein approach to right atrial appendage base implantation of the Aveir AR leadless pacemaker\",\"authors\":\"Howard Liu,&nbsp;Daniel Cortez\",\"doi\":\"10.1016/j.ipej.2024.11.004\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><div>There is no prior report of an Aveir leadless pacemaker implantation into the atrial appendage via the internal jugular vein.</div></div><div><h3>Case</h3><div>A 44-year-old female patient with history of multiple ablations for sinus node dysfunction presented with symptomatic bradycardia. The patient had femoral veins &lt;9mm, chronic pain at the femoral vein insertion sites, as well as a recent car accident with persistent leg pain due to femoral fractures. Placement of a leadless pacemaker was decided based on patient discretion. An Abbott Aveir AR leadless pacemaker was implanted via left internal jugular vein access without complication. Post device implantation showed threshold of 0.75V @0.4 ms, impedance of 340 Ω, Pwave of 4.4 mV. Six-month follow-up demonstrated a threshold of 0.5 V @0.2 milliseconds, impedance of 300 Ω and Pwave of 7.2 mV with 92 % pacing and predicted longevity of 12.7 years.</div></div><div><h3>Discussion</h3><div>The follow-up showed no complications in the patient. A similar approach may be feasible for other patients needing atrial leadless pacing, in which the transfemoral approach is not preferred.</div></div><div><h3>Conclusion</h3><div>Implantation of the Aveir AR leadless pacemaker into the right atrial appendage is feasible via the left internal jugular vein.</div></div>\",\"PeriodicalId\":35900,\"journal\":{\"name\":\"Indian Pacing and Electrophysiology Journal\",\"volume\":\"25 1\",\"pages\":\"Pages 25-27\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Indian Pacing and Electrophysiology Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0972629224001621\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian Pacing and Electrophysiology Journal","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0972629224001621","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

导读:目前尚无Aveir无导线起搏器经颈内静脉植入心房附件的报道。病例:一名44岁女性,因窦房结功能障碍多次消融史,表现为症状性心动过缓。讨论:随访无并发症发生。类似的入路可能适用于其他需要心房无导线起搏的患者,其中不推荐经股动脉入路。结论:Aveir AR无铅起搏器经左颈内静脉置入右心耳是可行的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Left internal jugular vein approach to right atrial appendage base implantation of the Aveir AR leadless pacemaker

Introduction

There is no prior report of an Aveir leadless pacemaker implantation into the atrial appendage via the internal jugular vein.

Case

A 44-year-old female patient with history of multiple ablations for sinus node dysfunction presented with symptomatic bradycardia. The patient had femoral veins <9mm, chronic pain at the femoral vein insertion sites, as well as a recent car accident with persistent leg pain due to femoral fractures. Placement of a leadless pacemaker was decided based on patient discretion. An Abbott Aveir AR leadless pacemaker was implanted via left internal jugular vein access without complication. Post device implantation showed threshold of 0.75V @0.4 ms, impedance of 340 Ω, Pwave of 4.4 mV. Six-month follow-up demonstrated a threshold of 0.5 V @0.2 milliseconds, impedance of 300 Ω and Pwave of 7.2 mV with 92 % pacing and predicted longevity of 12.7 years.

Discussion

The follow-up showed no complications in the patient. A similar approach may be feasible for other patients needing atrial leadless pacing, in which the transfemoral approach is not preferred.

Conclusion

Implantation of the Aveir AR leadless pacemaker into the right atrial appendage is feasible via the left internal jugular vein.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Indian Pacing and Electrophysiology Journal
Indian Pacing and Electrophysiology Journal Medicine-Cardiology and Cardiovascular Medicine
CiteScore
2.20
自引率
0.00%
发文量
91
审稿时长
61 days
期刊介绍: Indian Pacing and Electrophysiology Journal is a peer reviewed online journal devoted to cardiac pacing and electrophysiology. Editorial Advisory Board includes eminent personalities in the field of cardiac pacing and electrophysiology from Asia, Australia, Europe and North America.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信