{"title":"复合树脂基定制扭矩对舌托抗剪强度和黏合剂残余形态的影响。","authors":"Sekar Santhosh Kumar, Shivangi Ramteke, Bhaskar Nivethitha, Balasubramanian Madhan","doi":"10.1007/s00056-024-00559-6","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study evaluated the effect of torque customization of lingual brackets with resin-modified bases on their shear bond strength (SBS), adhesive remnant index (ARI), and bond failure patterns (BFP).</p><p><strong>Methods: </strong>The sample comprised 200 lingual lower incisor brackets (DTC® ORG, DTC Medical Apparatus, Hangzhou, China); 40 brackets were tested as-received and 4 groups with 40 brackets each were customized for -10, -20, +10 and +20° torque respectively with light-cured composite resin (Transbond XT™, 3M Unitek, Monrovia, CA, USA) bases. All brackets were bonded to bovine mandibular incisors. Their SBS were estimated in a universal testing machine (INSTRON®, Norwood, MA, USA) with a gingivally directed force applied at the incisal bracket-adhesive interface with 1 mm/min crosshead speed. Their ARI and BFP were evaluated under 20× magnification.</p><p><strong>Results: </strong>The SBS was 19.9 ± 7.6 MPa for noncustomized brackets, 20.1 ± 9.0 and 18.7 ± 8.2 MPa for brackets with 10 and 20° of negative torque, and 16.6 ± 5.68 and 19.45 ± 5.84 MPa for brackets with 10 and 20° of positive torque, respectively. The differences were not statistically significant (p = 0.097). Teeth with the -20° brackets exhibited higher median ARI scores than noncustomized brackets (1.5 vs 2, p = 0.018). Adhesive-cohesive bond failure with nearly axial split was more frequent in noncustomized brackets than customized ones, the reverse being true for adhesive-cohesive failure with nearly vertical split (p < 0.05). Truly cohesive bond failures were seen only in brackets with positive torque.</p><p><strong>Conclusions: </strong>Torque customization with a resin base is an acceptable strategy for metal orthodontic brackets as it does not affect their SBS. However, their BFP differed from noncustomized brackets, depending on the type of torque introduced.</p>","PeriodicalId":54776,"journal":{"name":"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie","volume":" ","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2024-12-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effect of torque customization with composite resin bases on the shear bond strength and adhesive remnant patterns of lingual brackets : An in vitro evaluation.\",\"authors\":\"Sekar Santhosh Kumar, Shivangi Ramteke, Bhaskar Nivethitha, Balasubramanian Madhan\",\"doi\":\"10.1007/s00056-024-00559-6\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>This study evaluated the effect of torque customization of lingual brackets with resin-modified bases on their shear bond strength (SBS), adhesive remnant index (ARI), and bond failure patterns (BFP).</p><p><strong>Methods: </strong>The sample comprised 200 lingual lower incisor brackets (DTC® ORG, DTC Medical Apparatus, Hangzhou, China); 40 brackets were tested as-received and 4 groups with 40 brackets each were customized for -10, -20, +10 and +20° torque respectively with light-cured composite resin (Transbond XT™, 3M Unitek, Monrovia, CA, USA) bases. All brackets were bonded to bovine mandibular incisors. Their SBS were estimated in a universal testing machine (INSTRON®, Norwood, MA, USA) with a gingivally directed force applied at the incisal bracket-adhesive interface with 1 mm/min crosshead speed. Their ARI and BFP were evaluated under 20× magnification.</p><p><strong>Results: </strong>The SBS was 19.9 ± 7.6 MPa for noncustomized brackets, 20.1 ± 9.0 and 18.7 ± 8.2 MPa for brackets with 10 and 20° of negative torque, and 16.6 ± 5.68 and 19.45 ± 5.84 MPa for brackets with 10 and 20° of positive torque, respectively. The differences were not statistically significant (p = 0.097). Teeth with the -20° brackets exhibited higher median ARI scores than noncustomized brackets (1.5 vs 2, p = 0.018). Adhesive-cohesive bond failure with nearly axial split was more frequent in noncustomized brackets than customized ones, the reverse being true for adhesive-cohesive failure with nearly vertical split (p < 0.05). Truly cohesive bond failures were seen only in brackets with positive torque.</p><p><strong>Conclusions: </strong>Torque customization with a resin base is an acceptable strategy for metal orthodontic brackets as it does not affect their SBS. However, their BFP differed from noncustomized brackets, depending on the type of torque introduced.</p>\",\"PeriodicalId\":54776,\"journal\":{\"name\":\"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-12-06\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s00056-024-00559-6\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00056-024-00559-6","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

目的:研究树脂基托槽的扭矩定制对其剪切粘结强度(SBS)、粘结残余指数(ARI)和粘结破坏模式(BFP)的影响。方法:样本包括200个舌下切牙托(DTC®ORG, DTC Medical Apparatus, Hangzhou, China);测试了40个托架,每组40个托架分别定制为-10、-20、+10和+20°扭矩,使用光固化复合树脂(Transbond XT™,3M Unitek, Monrovia, CA, USA)。所有托槽均与牛下颌门牙结合。在通用试验机(INSTRON®,Norwood, MA, USA)上估计他们的SBS,在切牙支架-粘合剂界面上以1 mm/min的十字头速度施加牙龈导向力。在20倍放大镜下评估ARI和BFP。结果:非定制托槽的SBS分别为19.9 ±7.6 MPa,负扭矩10°和20°托槽的SBS分别为20.1 ±9.0和18.7 ±8.2 MPa,正扭矩10°和20°托槽的SBS分别为16.6 ±5.68和19.45 ±5.84 MPa。差异无统计学意义(p = 0.097)。使用-20°托槽的牙齿ARI评分中位数高于非定制托槽(1.5 vs 2, p = 0.018)。非定制托槽中近轴向分裂的粘接性-粘接性破坏比定制托槽更频繁,而近垂直分裂的粘接性-粘接性破坏则相反(p 结论:树脂基托的扭矩定制是金属正畸托槽可接受的策略,因为它不影响其SBS。然而,根据引入的扭矩类型,它们的BFP与非定制支架不同。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Effect of torque customization with composite resin bases on the shear bond strength and adhesive remnant patterns of lingual brackets : An in vitro evaluation.

Purpose: This study evaluated the effect of torque customization of lingual brackets with resin-modified bases on their shear bond strength (SBS), adhesive remnant index (ARI), and bond failure patterns (BFP).

Methods: The sample comprised 200 lingual lower incisor brackets (DTC® ORG, DTC Medical Apparatus, Hangzhou, China); 40 brackets were tested as-received and 4 groups with 40 brackets each were customized for -10, -20, +10 and +20° torque respectively with light-cured composite resin (Transbond XT™, 3M Unitek, Monrovia, CA, USA) bases. All brackets were bonded to bovine mandibular incisors. Their SBS were estimated in a universal testing machine (INSTRON®, Norwood, MA, USA) with a gingivally directed force applied at the incisal bracket-adhesive interface with 1 mm/min crosshead speed. Their ARI and BFP were evaluated under 20× magnification.

Results: The SBS was 19.9 ± 7.6 MPa for noncustomized brackets, 20.1 ± 9.0 and 18.7 ± 8.2 MPa for brackets with 10 and 20° of negative torque, and 16.6 ± 5.68 and 19.45 ± 5.84 MPa for brackets with 10 and 20° of positive torque, respectively. The differences were not statistically significant (p = 0.097). Teeth with the -20° brackets exhibited higher median ARI scores than noncustomized brackets (1.5 vs 2, p = 0.018). Adhesive-cohesive bond failure with nearly axial split was more frequent in noncustomized brackets than customized ones, the reverse being true for adhesive-cohesive failure with nearly vertical split (p < 0.05). Truly cohesive bond failures were seen only in brackets with positive torque.

Conclusions: Torque customization with a resin base is an acceptable strategy for metal orthodontic brackets as it does not affect their SBS. However, their BFP differed from noncustomized brackets, depending on the type of torque introduced.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.90
自引率
0.00%
发文量
64
审稿时长
>12 weeks
期刊介绍: The Journal of Orofacial Orthopedics provides orthodontists and dentists who are also actively interested in orthodontics, whether in university clinics or private practice, with highly authoritative and up-to-date information based on experimental and clinical research. The journal is one of the leading publications for the promulgation of the results of original work both in the areas of scientific and clinical orthodontics and related areas. All articles undergo peer review before publication. The German Society of Orthodontics (DGKFO) also publishes in the journal important communications, statements and announcements.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信