便携式12导联心电图仪在心脏病患者中的临床价值:一项验证性研究。

IF 1.3 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS
Xiuyan Lu , Qiuhuan Wang , Xiujie Sun , Yibing Shao , Wenbo Jiang
{"title":"便携式12导联心电图仪在心脏病患者中的临床价值:一项验证性研究。","authors":"Xiuyan Lu ,&nbsp;Qiuhuan Wang ,&nbsp;Xiujie Sun ,&nbsp;Yibing Shao ,&nbsp;Wenbo Jiang","doi":"10.1016/j.jelectrocard.2024.153835","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>The present study was conducted to assess the accuracy and reliability of portable 12‑lead electrocardiography (ECG) devices in patients with heart disease.</div></div><div><h3>Materials and methods</h3><div>This single-center, prospective, blinded study enrolled 62 patients between September and October 2023 from the Heart Center of a Class III hospital. In sequential tests on each patient, heart rate (HR) and the PR, QT, QTc and QRS intervals of ECG recordings obtained with a portable 12‑lead device (Weheal, CN) were compared with those obtained via conventional 12‑lead ECG. ECG parameters were read in batches by 3 blinded electrophysiologists. Two-tailed paired <em>t-</em>tests were used to compare the continuous variables. Agreement was evaluated via Bland–Altman plots.</div></div><div><h3>Results</h3><div>Sixty-two patients were included. HR and the QT, QTc and QRS intervals from the portable 12‑lead electrocardiogram recordings were essentially the same as those obtained via conventional ECG. Bland–Altman analysis revealed no significant differences in these values, indicating suitable agreement between the 2 measurements. The PR interval was 176.89 ± 29.53 ms in the portable group and 161.56 ± 17.78 ms in the standard group, which was statistically (<em>p</em> &lt; 0.001) but not clinically significant.</div></div><div><h3>Conclusions</h3><div>ECG recordings obtained with a portable 12‑lead device (Weheal, CN) allow for accurate HR, PR, QT, QTc and QRS assessments. Considering its simplicity, this approach has advantages over conventional ECG and can provide an alternative for evaluating patients outside the hospital. How to improve patients' acceptance of portable ECG machines still needs further research.</div></div>","PeriodicalId":15606,"journal":{"name":"Journal of electrocardiology","volume":"88 ","pages":"Article 153835"},"PeriodicalIF":1.3000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Clinical value of portable 12‐lead electrocardiography devices in patients with heart disease: A validation study\",\"authors\":\"Xiuyan Lu ,&nbsp;Qiuhuan Wang ,&nbsp;Xiujie Sun ,&nbsp;Yibing Shao ,&nbsp;Wenbo Jiang\",\"doi\":\"10.1016/j.jelectrocard.2024.153835\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objective</h3><div>The present study was conducted to assess the accuracy and reliability of portable 12‑lead electrocardiography (ECG) devices in patients with heart disease.</div></div><div><h3>Materials and methods</h3><div>This single-center, prospective, blinded study enrolled 62 patients between September and October 2023 from the Heart Center of a Class III hospital. In sequential tests on each patient, heart rate (HR) and the PR, QT, QTc and QRS intervals of ECG recordings obtained with a portable 12‑lead device (Weheal, CN) were compared with those obtained via conventional 12‑lead ECG. ECG parameters were read in batches by 3 blinded electrophysiologists. Two-tailed paired <em>t-</em>tests were used to compare the continuous variables. Agreement was evaluated via Bland–Altman plots.</div></div><div><h3>Results</h3><div>Sixty-two patients were included. HR and the QT, QTc and QRS intervals from the portable 12‑lead electrocardiogram recordings were essentially the same as those obtained via conventional ECG. Bland–Altman analysis revealed no significant differences in these values, indicating suitable agreement between the 2 measurements. The PR interval was 176.89 ± 29.53 ms in the portable group and 161.56 ± 17.78 ms in the standard group, which was statistically (<em>p</em> &lt; 0.001) but not clinically significant.</div></div><div><h3>Conclusions</h3><div>ECG recordings obtained with a portable 12‑lead device (Weheal, CN) allow for accurate HR, PR, QT, QTc and QRS assessments. Considering its simplicity, this approach has advantages over conventional ECG and can provide an alternative for evaluating patients outside the hospital. How to improve patients' acceptance of portable ECG machines still needs further research.</div></div>\",\"PeriodicalId\":15606,\"journal\":{\"name\":\"Journal of electrocardiology\",\"volume\":\"88 \",\"pages\":\"Article 153835\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of electrocardiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0022073624003054\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of electrocardiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0022073624003054","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究旨在评估便携式12导联心电图(ECG)设备在心脏病患者中的准确性和可靠性。材料和方法:这项单中心、前瞻性、盲法研究于2023年9月至10月从某三甲医院心脏中心招募了62例患者。在对每位患者进行的顺序试验中,将使用便携式12导联仪(Weheal, CN)获得的心电图记录的心率(HR)、PR、QT、QTc和QRS间期与常规12导联心电图记录的心电图进行比较。3名盲法电生理学家分批读取心电参数。采用双尾配对t检验比较连续变量。通过Bland-Altman图评估一致性。结果:纳入62例患者。便携式12导联心电图记录的HR、QT、QTc和QRS间期与常规心电图记录基本相同。Bland-Altman分析显示这些值没有显著差异,表明2个测量值之间存在合适的一致性。便携式组的PR间期为176.89±29.53 ms,标准组的PR间期为161.56±17.78 ms,差异有统计学意义(p)。结论:使用便携式12导联仪(Weheal, CN)获得的心电图记录可以准确评估HR、PR、QT、QTc和QRS。考虑到它的简单性,这种方法比传统的心电图有优势,可以为医院外的患者提供评估的替代方法。如何提高患者对便携式心电机的接受程度还有待进一步研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Clinical value of portable 12‐lead electrocardiography devices in patients with heart disease: A validation study

Objective

The present study was conducted to assess the accuracy and reliability of portable 12‑lead electrocardiography (ECG) devices in patients with heart disease.

Materials and methods

This single-center, prospective, blinded study enrolled 62 patients between September and October 2023 from the Heart Center of a Class III hospital. In sequential tests on each patient, heart rate (HR) and the PR, QT, QTc and QRS intervals of ECG recordings obtained with a portable 12‑lead device (Weheal, CN) were compared with those obtained via conventional 12‑lead ECG. ECG parameters were read in batches by 3 blinded electrophysiologists. Two-tailed paired t-tests were used to compare the continuous variables. Agreement was evaluated via Bland–Altman plots.

Results

Sixty-two patients were included. HR and the QT, QTc and QRS intervals from the portable 12‑lead electrocardiogram recordings were essentially the same as those obtained via conventional ECG. Bland–Altman analysis revealed no significant differences in these values, indicating suitable agreement between the 2 measurements. The PR interval was 176.89 ± 29.53 ms in the portable group and 161.56 ± 17.78 ms in the standard group, which was statistically (p < 0.001) but not clinically significant.

Conclusions

ECG recordings obtained with a portable 12‑lead device (Weheal, CN) allow for accurate HR, PR, QT, QTc and QRS assessments. Considering its simplicity, this approach has advantages over conventional ECG and can provide an alternative for evaluating patients outside the hospital. How to improve patients' acceptance of portable ECG machines still needs further research.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of electrocardiology
Journal of electrocardiology 医学-心血管系统
CiteScore
2.70
自引率
7.70%
发文量
152
审稿时长
38 days
期刊介绍: The Journal of Electrocardiology is devoted exclusively to clinical and experimental studies of the electrical activities of the heart. It seeks to contribute significantly to the accuracy of diagnosis and prognosis and the effective treatment, prevention, or delay of heart disease. Editorial contents include electrocardiography, vectorcardiography, arrhythmias, membrane action potential, cardiac pacing, monitoring defibrillation, instrumentation, drug effects, and computer applications.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信