Eun Kyoung Kim, Min-Ku Chon, Hyun-Sook Kim, Yong-Hyun Park, Sang-Hyun Lee, Ki Seok Choo, Hyung Gon Je, Dae-Hee Kim, Tae Oh Kim, Yoon Seok Koh, Jae-Hyeong Park, Jae-Hwan Lee, Young Jin Choi, Eun Seok Shin, Hyuck-Jun Yoon, Seung-Whan Lee, Joo-Yong Hahn
{"title":"枢轴球囊治疗严重三尖瓣反流的安全性和有效性:首例手术经验。","authors":"Eun Kyoung Kim, Min-Ku Chon, Hyun-Sook Kim, Yong-Hyun Park, Sang-Hyun Lee, Ki Seok Choo, Hyung Gon Je, Dae-Hee Kim, Tae Oh Kim, Yoon Seok Koh, Jae-Hyeong Park, Jae-Hwan Lee, Young Jin Choi, Eun Seok Shin, Hyuck-Jun Yoon, Seung-Whan Lee, Joo-Yong Hahn","doi":"10.4070/kcj.2024.0147","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and objectives: </strong>Among various emerging catheter-based treatments for severe tricuspid regurgitation (TR), the spacer device can reduce the regurgitation orifice without manipulating the valve leaflet. However, its clinical application has been hampered by traumatic anchoring to the myocardium and the coaxial alignment of the balloon resulting in insufficient TR reduction. This study aimed to evaluate the early-stage safety, technical feasibility, and preliminary efficacy of the novel atraumatic vertical spacer in patients with isolated severe TR.</p><p><strong>Methods: </strong>All procedures were guided by fluoroscopy and transthoracic echocardiography. The maximum device placement time with an inflated balloon was 24 hours. Changes in the amount of TR, right ventricular function, and patient hemodynamics were measured during balloon deployment.</p><p><strong>Results: </strong>A total of 7 patients (median age 74), underwent successful device implantation without procedure-related complications. During balloon inflation (median 25 minutes), there were no symptoms or signs indicative of TR intolerance. TR was reduced by 1 grade or greater in all patients, with 2 patients exhibiting a reduction of 3 grades, from torrential TR to a moderate degree. Mild TR after balloon inflation was achieved in 3 patients with baseline severe TR. The TR reduction observed during initial balloon deployment was sustained during the subsequent balloon maintenance period.</p><p><strong>Conclusions: </strong>The Pivot-balloon procedure was safe, technically feasible, and effective in reducing TR in patients with severe TR. No periprocedural complications or adverse cardiovascular events were reported during device placement with TR reduction observed in all patients. However, longer-term follow-up is needed to confirm safety and treatment effect.</p><p><strong>Trial registration: </strong>ClinicalTrials.gov Identifier: NCT05648838.</p>","PeriodicalId":17850,"journal":{"name":"Korean Circulation Journal","volume":" ","pages":"20-31"},"PeriodicalIF":3.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11735152/pdf/","citationCount":"0","resultStr":"{\"title\":\"Safety and Efficacy of Pivot-Balloon for Severe Tricuspid Regurgitation: The First-in-Man Experiences.\",\"authors\":\"Eun Kyoung Kim, Min-Ku Chon, Hyun-Sook Kim, Yong-Hyun Park, Sang-Hyun Lee, Ki Seok Choo, Hyung Gon Je, Dae-Hee Kim, Tae Oh Kim, Yoon Seok Koh, Jae-Hyeong Park, Jae-Hwan Lee, Young Jin Choi, Eun Seok Shin, Hyuck-Jun Yoon, Seung-Whan Lee, Joo-Yong Hahn\",\"doi\":\"10.4070/kcj.2024.0147\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background and objectives: </strong>Among various emerging catheter-based treatments for severe tricuspid regurgitation (TR), the spacer device can reduce the regurgitation orifice without manipulating the valve leaflet. However, its clinical application has been hampered by traumatic anchoring to the myocardium and the coaxial alignment of the balloon resulting in insufficient TR reduction. This study aimed to evaluate the early-stage safety, technical feasibility, and preliminary efficacy of the novel atraumatic vertical spacer in patients with isolated severe TR.</p><p><strong>Methods: </strong>All procedures were guided by fluoroscopy and transthoracic echocardiography. The maximum device placement time with an inflated balloon was 24 hours. Changes in the amount of TR, right ventricular function, and patient hemodynamics were measured during balloon deployment.</p><p><strong>Results: </strong>A total of 7 patients (median age 74), underwent successful device implantation without procedure-related complications. During balloon inflation (median 25 minutes), there were no symptoms or signs indicative of TR intolerance. TR was reduced by 1 grade or greater in all patients, with 2 patients exhibiting a reduction of 3 grades, from torrential TR to a moderate degree. Mild TR after balloon inflation was achieved in 3 patients with baseline severe TR. The TR reduction observed during initial balloon deployment was sustained during the subsequent balloon maintenance period.</p><p><strong>Conclusions: </strong>The Pivot-balloon procedure was safe, technically feasible, and effective in reducing TR in patients with severe TR. No periprocedural complications or adverse cardiovascular events were reported during device placement with TR reduction observed in all patients. However, longer-term follow-up is needed to confirm safety and treatment effect.</p><p><strong>Trial registration: </strong>ClinicalTrials.gov Identifier: NCT05648838.</p>\",\"PeriodicalId\":17850,\"journal\":{\"name\":\"Korean Circulation Journal\",\"volume\":\" \",\"pages\":\"20-31\"},\"PeriodicalIF\":3.0000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11735152/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Korean Circulation Journal\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.4070/kcj.2024.0147\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/10/14 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Korean Circulation Journal","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4070/kcj.2024.0147","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/14 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Safety and Efficacy of Pivot-Balloon for Severe Tricuspid Regurgitation: The First-in-Man Experiences.
Background and objectives: Among various emerging catheter-based treatments for severe tricuspid regurgitation (TR), the spacer device can reduce the regurgitation orifice without manipulating the valve leaflet. However, its clinical application has been hampered by traumatic anchoring to the myocardium and the coaxial alignment of the balloon resulting in insufficient TR reduction. This study aimed to evaluate the early-stage safety, technical feasibility, and preliminary efficacy of the novel atraumatic vertical spacer in patients with isolated severe TR.
Methods: All procedures were guided by fluoroscopy and transthoracic echocardiography. The maximum device placement time with an inflated balloon was 24 hours. Changes in the amount of TR, right ventricular function, and patient hemodynamics were measured during balloon deployment.
Results: A total of 7 patients (median age 74), underwent successful device implantation without procedure-related complications. During balloon inflation (median 25 minutes), there were no symptoms or signs indicative of TR intolerance. TR was reduced by 1 grade or greater in all patients, with 2 patients exhibiting a reduction of 3 grades, from torrential TR to a moderate degree. Mild TR after balloon inflation was achieved in 3 patients with baseline severe TR. The TR reduction observed during initial balloon deployment was sustained during the subsequent balloon maintenance period.
Conclusions: The Pivot-balloon procedure was safe, technically feasible, and effective in reducing TR in patients with severe TR. No periprocedural complications or adverse cardiovascular events were reported during device placement with TR reduction observed in all patients. However, longer-term follow-up is needed to confirm safety and treatment effect.
期刊介绍:
Korean Circulation Journal is the official journal of the Korean Society of Cardiology, the Korean Pediatric Heart Society, the Korean Society of Interventional Cardiology, and the Korean Society of Heart Failure. Abbreviated title is ''Korean Circ J''.
Korean Circulation Journal, established in 1971, is a professional, peer-reviewed journal covering all aspects of cardiovascular medicine, including original articles of basic research and clinical findings, review articles, editorials, images in cardiovascular medicine, and letters to the editor. Korean Circulation Journal is published monthly in English and publishes scientific and state-of-the-art clinical articles aimed at improving human health in general and contributing to the treatment and prevention of cardiovascular diseases in particular.
The journal is published on the official website (https://e-kcj.org). It is indexed in PubMed, PubMed Central, Science Citation Index Expanded (SCIE, Web of Science), Scopus, EMBASE, Chemical Abstracts Service (CAS), Google Scholar, KoreaMed, KoreaMed Synapse and KoMCI, and easily available to wide international researchers