通过眼压调节泵施加负压,持续降低眼压。

Clinical ophthalmology (Auckland, N.Z.) Pub Date : 2024-11-06 eCollection Date: 2024-01-01 DOI:10.2147/OPTH.S492530
Tanner J Ferguson, Leon W Herndon, Daniel C Terveen, Manjool Shah, Thomas W Samuelson, Paul Yoo, John P Berdahl
{"title":"通过眼压调节泵施加负压,持续降低眼压。","authors":"Tanner J Ferguson, Leon W Herndon, Daniel C Terveen, Manjool Shah, Thomas W Samuelson, Paul Yoo, John P Berdahl","doi":"10.2147/OPTH.S492530","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the sustainability of IOP reduction with continuous negative pressure application over an extended duration with use of the Ocular Pressure Adjusting Pump.</p><p><strong>Methods: </strong>Prospective, controlled, open-label, randomized, single site, pilot study. Subjects with primary open-angle glaucoma (OAG) were enrolled. One eye of each subject was randomized to receive negative pressure application; the fellow eye served as a control. The study eye negative pressure setting was programmed for 60% of the baseline IOP. Subjects wore the Ocular Pressure Adjusting Pump for 8 consecutive hours and IOP measurements occurred at 2-hour intervals for a total of 5 IOP measurements (08:00, 10:00, 12:00, 14:00, 16:00).</p><p><strong>Results: </strong>Nine subjects successfully enrolled and completed the study. The mean programmed negative pressure setting was -12.0 mmHg. At baseline, the mean IOP in the study eye was 21.4 ± 4.3 mmHg. The mean IOP reduction in the study eye at hours 0, 2, 4, 6, and 8 was 8.1 (37%), 6.4 (28%), 6.3 (29%), 7.3 (34%) and 6.7 (31%), respectively. All IOP measurements during negative pressure application were reduced from baseline. There were no serious adverse events.</p><p><strong>Conclusion: </strong>The Ocular Pressure Adjusting Pump provides a sustained reduction in IOP while the device is worn with negative pressure applied with an IOP reduction exceeding 25% across 8 hours of continuous wear.</p>","PeriodicalId":93945,"journal":{"name":"Clinical ophthalmology (Auckland, N.Z.)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-11-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11550690/pdf/","citationCount":"0","resultStr":"{\"title\":\"Application of Negative Pressure by the Ocular Pressure Adjusting Pump to Provide a Sustained Reduction in IOP.\",\"authors\":\"Tanner J Ferguson, Leon W Herndon, Daniel C Terveen, Manjool Shah, Thomas W Samuelson, Paul Yoo, John P Berdahl\",\"doi\":\"10.2147/OPTH.S492530\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To evaluate the sustainability of IOP reduction with continuous negative pressure application over an extended duration with use of the Ocular Pressure Adjusting Pump.</p><p><strong>Methods: </strong>Prospective, controlled, open-label, randomized, single site, pilot study. Subjects with primary open-angle glaucoma (OAG) were enrolled. One eye of each subject was randomized to receive negative pressure application; the fellow eye served as a control. The study eye negative pressure setting was programmed for 60% of the baseline IOP. Subjects wore the Ocular Pressure Adjusting Pump for 8 consecutive hours and IOP measurements occurred at 2-hour intervals for a total of 5 IOP measurements (08:00, 10:00, 12:00, 14:00, 16:00).</p><p><strong>Results: </strong>Nine subjects successfully enrolled and completed the study. The mean programmed negative pressure setting was -12.0 mmHg. At baseline, the mean IOP in the study eye was 21.4 ± 4.3 mmHg. The mean IOP reduction in the study eye at hours 0, 2, 4, 6, and 8 was 8.1 (37%), 6.4 (28%), 6.3 (29%), 7.3 (34%) and 6.7 (31%), respectively. All IOP measurements during negative pressure application were reduced from baseline. There were no serious adverse events.</p><p><strong>Conclusion: </strong>The Ocular Pressure Adjusting Pump provides a sustained reduction in IOP while the device is worn with negative pressure applied with an IOP reduction exceeding 25% across 8 hours of continuous wear.</p>\",\"PeriodicalId\":93945,\"journal\":{\"name\":\"Clinical ophthalmology (Auckland, N.Z.)\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-11-06\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11550690/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical ophthalmology (Auckland, N.Z.)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2147/OPTH.S492530\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical ophthalmology (Auckland, N.Z.)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/OPTH.S492530","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:评估使用眼压调节泵长时间持续负压治疗降低眼压的可持续性:前瞻性、对照、开放标签、随机、单点试验研究。研究对象为原发性开角型青光眼(OAG)患者。每名受试者的一只眼被随机分配接受负压治疗,另一只眼作为对照组。研究用眼的负压设置为基线眼压的 60%。受试者连续 8 小时佩戴眼压调节泵,每隔 2 小时测量一次眼压,共测量 5 次眼压(08:00、10:00、12:00、14:00、16:00):结果:9 名受试者成功注册并完成了研究。平均程序负压设置为 -12.0 mmHg。基线时,研究用眼的平均眼压为 21.4 ± 4.3 mmHg。在第 0、2、4、6 和 8 小时,研究用眼的平均眼压降低率分别为 8.1 (37%)、6.4 (28%)、6.3 (29%)、7.3 (34%) 和 6.7 (31%)。负压治疗期间的所有眼压测量值均比基线值有所下降。无严重不良事件发生:结论:眼压调节泵可在佩戴设备并施加负压的情况下持续降低眼压,连续佩戴 8 小时后眼压降低幅度超过 25%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Application of Negative Pressure by the Ocular Pressure Adjusting Pump to Provide a Sustained Reduction in IOP.

Purpose: To evaluate the sustainability of IOP reduction with continuous negative pressure application over an extended duration with use of the Ocular Pressure Adjusting Pump.

Methods: Prospective, controlled, open-label, randomized, single site, pilot study. Subjects with primary open-angle glaucoma (OAG) were enrolled. One eye of each subject was randomized to receive negative pressure application; the fellow eye served as a control. The study eye negative pressure setting was programmed for 60% of the baseline IOP. Subjects wore the Ocular Pressure Adjusting Pump for 8 consecutive hours and IOP measurements occurred at 2-hour intervals for a total of 5 IOP measurements (08:00, 10:00, 12:00, 14:00, 16:00).

Results: Nine subjects successfully enrolled and completed the study. The mean programmed negative pressure setting was -12.0 mmHg. At baseline, the mean IOP in the study eye was 21.4 ± 4.3 mmHg. The mean IOP reduction in the study eye at hours 0, 2, 4, 6, and 8 was 8.1 (37%), 6.4 (28%), 6.3 (29%), 7.3 (34%) and 6.7 (31%), respectively. All IOP measurements during negative pressure application were reduced from baseline. There were no serious adverse events.

Conclusion: The Ocular Pressure Adjusting Pump provides a sustained reduction in IOP while the device is worn with negative pressure applied with an IOP reduction exceeding 25% across 8 hours of continuous wear.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
4.10
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信