用无线超声波机测量肩峰下撞击综合征患者的肩峰距离:机器间可靠性研究。

IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Journal of Ultrasonography Pub Date : 2024-11-08 eCollection Date: 2024-11-01 DOI:10.15557/jou.2024.0018
Burak Tayyip Dede, Ebru Aytekin, Fatih Bağcier
{"title":"用无线超声波机测量肩峰下撞击综合征患者的肩峰距离:机器间可靠性研究。","authors":"Burak Tayyip Dede, Ebru Aytekin, Fatih Bağcier","doi":"10.15557/jou.2024.0018","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>Shoulder pain is the third most prevalent musculoskeletal condition that impairs function. Subacromial impingement syndrome is among the most typical causes of shoulder pain. The aim of this study was to evaluate inter-machine reliability of acromiohumeral distance measurements performed with standard ultrasound and wireless ultrasound devices in patients with subacromial impingement syndrome.</p><p><strong>Material and methods: </strong>A total of 61 participants diagnosed with subacromial impingement syndrome were included. Acromiohumeral distance was measured with wireless and standard ultrasound devices in a neutral position and at 60-degree abduction, respectively. The inter-machine intraclass correlation coefficient, standard error of measurements, and minimum detectable changes were calculated.</p><p><strong>Results: </strong>Inter-machine reliability measured in the neutral position was excellent (ICC = 0.97, 95% CI = 0.95-0.98); the standard error of measurement was 0.23 mm, and the minimum detectable change was 0.63 mm. Inter-machine reliability measured at 60 degrees abduction was excellent as well (ICC = 0.96; 95% CI; 0.93-0.97). The standard error of measurements was 0.20 mm, and the minimum detectable change was 0.55 mm. The mean difference between the two machines was 0.04 mm for the neutral position and 0.02 mm for the 60-degree abduction position.</p><p><strong>Conclusions: </strong>The study showed that wireless ultrasound devices were similar to standard ultrasound devices in measuring the acromiohumeral distance in patients with subacromial impingement syndrome. The findings could contribute to a significant improvement in the clinical use of wireless ultrasound devices.</p>","PeriodicalId":45612,"journal":{"name":"Journal of Ultrasonography","volume":"24 97","pages":"1-5"},"PeriodicalIF":1.3000,"publicationDate":"2024-11-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11546899/pdf/","citationCount":"0","resultStr":"{\"title\":\"Measures of acromiohumeral distance with wireless ultrasound machine in subacromial impingement syndrome: an inter-machine reliability study.\",\"authors\":\"Burak Tayyip Dede, Ebru Aytekin, Fatih Bağcier\",\"doi\":\"10.15557/jou.2024.0018\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Aim: </strong>Shoulder pain is the third most prevalent musculoskeletal condition that impairs function. Subacromial impingement syndrome is among the most typical causes of shoulder pain. The aim of this study was to evaluate inter-machine reliability of acromiohumeral distance measurements performed with standard ultrasound and wireless ultrasound devices in patients with subacromial impingement syndrome.</p><p><strong>Material and methods: </strong>A total of 61 participants diagnosed with subacromial impingement syndrome were included. Acromiohumeral distance was measured with wireless and standard ultrasound devices in a neutral position and at 60-degree abduction, respectively. The inter-machine intraclass correlation coefficient, standard error of measurements, and minimum detectable changes were calculated.</p><p><strong>Results: </strong>Inter-machine reliability measured in the neutral position was excellent (ICC = 0.97, 95% CI = 0.95-0.98); the standard error of measurement was 0.23 mm, and the minimum detectable change was 0.63 mm. Inter-machine reliability measured at 60 degrees abduction was excellent as well (ICC = 0.96; 95% CI; 0.93-0.97). The standard error of measurements was 0.20 mm, and the minimum detectable change was 0.55 mm. The mean difference between the two machines was 0.04 mm for the neutral position and 0.02 mm for the 60-degree abduction position.</p><p><strong>Conclusions: </strong>The study showed that wireless ultrasound devices were similar to standard ultrasound devices in measuring the acromiohumeral distance in patients with subacromial impingement syndrome. The findings could contribute to a significant improvement in the clinical use of wireless ultrasound devices.</p>\",\"PeriodicalId\":45612,\"journal\":{\"name\":\"Journal of Ultrasonography\",\"volume\":\"24 97\",\"pages\":\"1-5\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-11-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11546899/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Ultrasonography\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.15557/jou.2024.0018\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/11/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Ultrasonography","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15557/jou.2024.0018","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:肩部疼痛是影响功能的第三大肌肉骨骼疾病。肩峰下撞击综合征是导致肩痛的最典型原因之一。本研究旨在评估用标准超声波和无线超声波设备对肩峰下撞击综合征患者进行肩峰距离测量的机器间可靠性:共纳入61名诊断为肩峰下撞击综合征的患者。分别在中立位和外展60度时使用无线和标准超声设备测量肱骨桡侧距离。计算了机器间类内相关系数、测量标准误差和最小可检测变化:结果:在中立位时测量的机器间可靠性极佳(ICC = 0.97,95% CI = 0.95-0.98);测量的标准误差为 0.23 毫米,可检测到的最小变化为 0.63 毫米。外展 60 度时的机器间可靠性也非常好(ICC = 0.96;95% CI;0.93-0.97)。测量的标准误差为 0.20 毫米,可检测到的最小变化为 0.55 毫米。两台机器在中立位的平均差异为 0.04 毫米,在外展 60 度位的平均差异为 0.02 毫米:研究表明,在测量肩峰下撞击综合征患者的肩肱骨距离方面,无线超声设备与标准超声设备相似。该研究结果有助于大大提高无线超声设备的临床使用率。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Measures of acromiohumeral distance with wireless ultrasound machine in subacromial impingement syndrome: an inter-machine reliability study.

Aim: Shoulder pain is the third most prevalent musculoskeletal condition that impairs function. Subacromial impingement syndrome is among the most typical causes of shoulder pain. The aim of this study was to evaluate inter-machine reliability of acromiohumeral distance measurements performed with standard ultrasound and wireless ultrasound devices in patients with subacromial impingement syndrome.

Material and methods: A total of 61 participants diagnosed with subacromial impingement syndrome were included. Acromiohumeral distance was measured with wireless and standard ultrasound devices in a neutral position and at 60-degree abduction, respectively. The inter-machine intraclass correlation coefficient, standard error of measurements, and minimum detectable changes were calculated.

Results: Inter-machine reliability measured in the neutral position was excellent (ICC = 0.97, 95% CI = 0.95-0.98); the standard error of measurement was 0.23 mm, and the minimum detectable change was 0.63 mm. Inter-machine reliability measured at 60 degrees abduction was excellent as well (ICC = 0.96; 95% CI; 0.93-0.97). The standard error of measurements was 0.20 mm, and the minimum detectable change was 0.55 mm. The mean difference between the two machines was 0.04 mm for the neutral position and 0.02 mm for the 60-degree abduction position.

Conclusions: The study showed that wireless ultrasound devices were similar to standard ultrasound devices in measuring the acromiohumeral distance in patients with subacromial impingement syndrome. The findings could contribute to a significant improvement in the clinical use of wireless ultrasound devices.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Ultrasonography
Journal of Ultrasonography RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
2.30
自引率
0.00%
发文量
58
审稿时长
20 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信