{"title":"护理点 CBC 血液分析仪的分析性能,包括 5 部分鉴别:一项前瞻性研究,评估基于微流体流式细胞仪的分析仪在免检环境中的应用。","authors":"Jane F Emerson, Hao Wang, Imran N Siddiqi","doi":"10.1093/ajcp/aqae149","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>A microfluidic flow cytometer-based point-of-care (POC) analyzer was validated against an in-laboratory hematology analyzer (Sysmex XN Automated Hematology System). Concordance on a full complete blood cell count (CBC) with 5-part differential, as performed by operators with no prior clinical laboratory experience, was evaluated.</p><p><strong>Methods: </strong>We prospectively collected 376 venous blood specimens (376) from individuals with self-reported medical conditions and from apparently healthy individuals. Forty-six additional remnant specimens were acquired to ensure coverage of analytic measuring ranges. Parallel testing was performed, with up to 7 hours between testing on the POC and Sysmex XN analyzers.</p><p><strong>Results: </strong>Regression analysis resulted in r values of 0.998 to 0.932 for all parameters of a 5-part differential CBC other than basophils (0.709). The mean percentage bias from the reference method, inclusive of the upper and lower reporting limits, was less than 2% for parameters other than lymphocytes (-6.4%), monocytes (25.9%), eosinophils (12.2%), and basophils (-15%). Overall agreement on abnormal flagging was 93.3%.</p><p><strong>Conclusions: </strong>The Cito CBC microflow cytometer (CytoChip Inc) provides a CBC with a 5-part differential with accuracy, precision, and abnormal flagging equivalent to a moderate-complexity hematology analyzer. It has the key features required of a POC device that can be operated in a waived setting: minimum space requirements, rapid results, single-action measurement (no sample processing or dilution), ease of use, and minimal blood volume.</p>","PeriodicalId":7506,"journal":{"name":"American journal of clinical pathology","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2024-11-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Analytical performance of a point-of-care CBC hematology analyzer, including a 5-part differential: A prospective study to evaluate a microfluidic flow cytometry-based analyzer in waived settings.\",\"authors\":\"Jane F Emerson, Hao Wang, Imran N Siddiqi\",\"doi\":\"10.1093/ajcp/aqae149\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>A microfluidic flow cytometer-based point-of-care (POC) analyzer was validated against an in-laboratory hematology analyzer (Sysmex XN Automated Hematology System). Concordance on a full complete blood cell count (CBC) with 5-part differential, as performed by operators with no prior clinical laboratory experience, was evaluated.</p><p><strong>Methods: </strong>We prospectively collected 376 venous blood specimens (376) from individuals with self-reported medical conditions and from apparently healthy individuals. Forty-six additional remnant specimens were acquired to ensure coverage of analytic measuring ranges. Parallel testing was performed, with up to 7 hours between testing on the POC and Sysmex XN analyzers.</p><p><strong>Results: </strong>Regression analysis resulted in r values of 0.998 to 0.932 for all parameters of a 5-part differential CBC other than basophils (0.709). The mean percentage bias from the reference method, inclusive of the upper and lower reporting limits, was less than 2% for parameters other than lymphocytes (-6.4%), monocytes (25.9%), eosinophils (12.2%), and basophils (-15%). Overall agreement on abnormal flagging was 93.3%.</p><p><strong>Conclusions: </strong>The Cito CBC microflow cytometer (CytoChip Inc) provides a CBC with a 5-part differential with accuracy, precision, and abnormal flagging equivalent to a moderate-complexity hematology analyzer. It has the key features required of a POC device that can be operated in a waived setting: minimum space requirements, rapid results, single-action measurement (no sample processing or dilution), ease of use, and minimal blood volume.</p>\",\"PeriodicalId\":7506,\"journal\":{\"name\":\"American journal of clinical pathology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2024-11-09\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"American journal of clinical pathology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/ajcp/aqae149\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"PATHOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of clinical pathology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/ajcp/aqae149","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"PATHOLOGY","Score":null,"Total":0}
Analytical performance of a point-of-care CBC hematology analyzer, including a 5-part differential: A prospective study to evaluate a microfluidic flow cytometry-based analyzer in waived settings.
Objectives: A microfluidic flow cytometer-based point-of-care (POC) analyzer was validated against an in-laboratory hematology analyzer (Sysmex XN Automated Hematology System). Concordance on a full complete blood cell count (CBC) with 5-part differential, as performed by operators with no prior clinical laboratory experience, was evaluated.
Methods: We prospectively collected 376 venous blood specimens (376) from individuals with self-reported medical conditions and from apparently healthy individuals. Forty-six additional remnant specimens were acquired to ensure coverage of analytic measuring ranges. Parallel testing was performed, with up to 7 hours between testing on the POC and Sysmex XN analyzers.
Results: Regression analysis resulted in r values of 0.998 to 0.932 for all parameters of a 5-part differential CBC other than basophils (0.709). The mean percentage bias from the reference method, inclusive of the upper and lower reporting limits, was less than 2% for parameters other than lymphocytes (-6.4%), monocytes (25.9%), eosinophils (12.2%), and basophils (-15%). Overall agreement on abnormal flagging was 93.3%.
Conclusions: The Cito CBC microflow cytometer (CytoChip Inc) provides a CBC with a 5-part differential with accuracy, precision, and abnormal flagging equivalent to a moderate-complexity hematology analyzer. It has the key features required of a POC device that can be operated in a waived setting: minimum space requirements, rapid results, single-action measurement (no sample processing or dilution), ease of use, and minimal blood volume.
期刊介绍:
The American Journal of Clinical Pathology (AJCP) is the official journal of the American Society for Clinical Pathology and the Academy of Clinical Laboratory Physicians and Scientists. It is a leading international journal for publication of articles concerning novel anatomic pathology and laboratory medicine observations on human disease. AJCP emphasizes articles that focus on the application of evolving technologies for the diagnosis and characterization of diseases and conditions, as well as those that have a direct link toward improving patient care.