评估用于脉宽调制人工耳蜗听觉映射的电诱发镫骨反射阈值(eSRT):多电极刺激的影响。

IF 1.9 3区 医学 Q3 CLINICAL NEUROLOGY
Pedro Goiana Martins, Leonel Luís, Manuel Segovia-Martinez, Behnam Molaee-Ardekani
{"title":"评估用于脉宽调制人工耳蜗听觉映射的电诱发镫骨反射阈值(eSRT):多电极刺激的影响。","authors":"Pedro Goiana Martins, Leonel Luís, Manuel Segovia-Martinez, Behnam Molaee-Ardekani","doi":"10.1097/MAO.0000000000004354","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study assesses the electrically evoked stapedius reflex threshold (eSRT) as an objective method for generating auditory maps in individuals with pulse-width-modulation cochlear implants (CIs). It investigates the impact of both single- and multiple-electrode stimulations on eSRT detection rates and their association with patients' comfort loudness levels (C-levels). Despite eSRT exploration by several CI brands incorporating pulse-amplitude-modulation stimulation technology, there is a lack of information on systems using pulse-width modulation.</p><p><strong>Approach: </strong>The eSRT was measured in 19 ears ipsilaterally and contralaterally in response to stimulations in groups of 1, 3, and 5 electrodes across five distinct cochlear regions, spanning from apical to basal. For each group, the eSRT detection rate and its correlation with the C-levels were analyzed both regionally and overall.</p><p><strong>Main results: </strong>In both contralateral and ipsilateral sides, higher electrically evoked stapedius reflex (eSR) detection rates were obtained for multielectrode (67%, 47%) than for single-electrode (41%, 23%) stimulation. Invoking eSR in the apical and basal regions was easier than in the middle region. The multiple stimulation just increased slightly the correlation coefficients between eSRT and C-levels from 0.73 to 0.77 and from 0.77 to 0.85 (p < 0.05) for the contralateral and ipsilateral ears, respectively.</p><p><strong>Significance: </strong>Our research indicates that using a multielectrode stimulation approach enhances the detection rate of eSRT and slightly improves the correlation coefficient between eSRT and C-levels. It is found preferable to obtain measurements from the contralateral side due to its higher detection rate, although the ipsilateral side yields a marginally higher correlation coefficient. The findings suggest that eSRTs obtained from pulse width modulation implants could aid audiologists in programming the device.</p>","PeriodicalId":19732,"journal":{"name":"Otology & Neurotology","volume":"45 10","pages":"1128-1134"},"PeriodicalIF":1.9000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Assessment of Electrically Evoked Stapedius Reflex Threshold (eSRT) for Auditory Mapping in Pulse Width Modulation Cochlear Implants: The Effect of Multielectrode Stimulation.\",\"authors\":\"Pedro Goiana Martins, Leonel Luís, Manuel Segovia-Martinez, Behnam Molaee-Ardekani\",\"doi\":\"10.1097/MAO.0000000000004354\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>This study assesses the electrically evoked stapedius reflex threshold (eSRT) as an objective method for generating auditory maps in individuals with pulse-width-modulation cochlear implants (CIs). It investigates the impact of both single- and multiple-electrode stimulations on eSRT detection rates and their association with patients' comfort loudness levels (C-levels). Despite eSRT exploration by several CI brands incorporating pulse-amplitude-modulation stimulation technology, there is a lack of information on systems using pulse-width modulation.</p><p><strong>Approach: </strong>The eSRT was measured in 19 ears ipsilaterally and contralaterally in response to stimulations in groups of 1, 3, and 5 electrodes across five distinct cochlear regions, spanning from apical to basal. For each group, the eSRT detection rate and its correlation with the C-levels were analyzed both regionally and overall.</p><p><strong>Main results: </strong>In both contralateral and ipsilateral sides, higher electrically evoked stapedius reflex (eSR) detection rates were obtained for multielectrode (67%, 47%) than for single-electrode (41%, 23%) stimulation. Invoking eSR in the apical and basal regions was easier than in the middle region. The multiple stimulation just increased slightly the correlation coefficients between eSRT and C-levels from 0.73 to 0.77 and from 0.77 to 0.85 (p < 0.05) for the contralateral and ipsilateral ears, respectively.</p><p><strong>Significance: </strong>Our research indicates that using a multielectrode stimulation approach enhances the detection rate of eSRT and slightly improves the correlation coefficient between eSRT and C-levels. It is found preferable to obtain measurements from the contralateral side due to its higher detection rate, although the ipsilateral side yields a marginally higher correlation coefficient. The findings suggest that eSRTs obtained from pulse width modulation implants could aid audiologists in programming the device.</p>\",\"PeriodicalId\":19732,\"journal\":{\"name\":\"Otology & Neurotology\",\"volume\":\"45 10\",\"pages\":\"1128-1134\"},\"PeriodicalIF\":1.9000,\"publicationDate\":\"2024-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Otology & Neurotology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MAO.0000000000004354\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CLINICAL NEUROLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Otology & Neurotology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MAO.0000000000004354","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

摘要

研究目的本研究将电诱发镫骨反射阈值(eSRT)作为一种客观的方法,对使用脉宽调制人工耳蜗(CI)的患者生成听觉地图进行评估。它研究了单电极和多电极刺激对 eSRT 检测率的影响及其与患者舒适响度水平(C 级)的关联。尽管多个采用脉冲幅度调制刺激技术的人工耳蜗品牌都对 eSRT 进行了研究,但有关采用脉宽调制技术的系统的信息还很缺乏:在同侧和对侧各测量了 19 只耳朵的 eSRT,这些耳朵对 1、3 和 5 个电极组的刺激做出了反应,这些电极组横跨从耳尖到基底的五个不同耳蜗区域。对每组的 eSRT 检测率及其与 C 级的相关性进行了区域和整体分析:主要结果:在对侧和同侧,多电极(67%、47%)刺激的电诱发镫骨反射(eSR)检测率高于单电极(41%、23%)刺激。在顶端和基底区域诱发 eSR 比在中间区域更容易。多重刺激只是略微提高了对侧耳和同侧耳的 eSRT 与 C 级之间的相关系数,分别从 0.73 升至 0.77 和从 0.77 升至 0.85(p < 0.05):我们的研究表明,使用多电极刺激方法可提高 eSRT 的检出率,并略微改善 eSRT 与 C 级之间的相关系数。尽管同侧的相关系数略高,但由于对侧的检测率更高,因此从对侧获得测量结果更为可取。研究结果表明,从脉宽调制植入体获得的 eSRT 可以帮助听力学家对设备进行编程。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Assessment of Electrically Evoked Stapedius Reflex Threshold (eSRT) for Auditory Mapping in Pulse Width Modulation Cochlear Implants: The Effect of Multielectrode Stimulation.

Objective: This study assesses the electrically evoked stapedius reflex threshold (eSRT) as an objective method for generating auditory maps in individuals with pulse-width-modulation cochlear implants (CIs). It investigates the impact of both single- and multiple-electrode stimulations on eSRT detection rates and their association with patients' comfort loudness levels (C-levels). Despite eSRT exploration by several CI brands incorporating pulse-amplitude-modulation stimulation technology, there is a lack of information on systems using pulse-width modulation.

Approach: The eSRT was measured in 19 ears ipsilaterally and contralaterally in response to stimulations in groups of 1, 3, and 5 electrodes across five distinct cochlear regions, spanning from apical to basal. For each group, the eSRT detection rate and its correlation with the C-levels were analyzed both regionally and overall.

Main results: In both contralateral and ipsilateral sides, higher electrically evoked stapedius reflex (eSR) detection rates were obtained for multielectrode (67%, 47%) than for single-electrode (41%, 23%) stimulation. Invoking eSR in the apical and basal regions was easier than in the middle region. The multiple stimulation just increased slightly the correlation coefficients between eSRT and C-levels from 0.73 to 0.77 and from 0.77 to 0.85 (p < 0.05) for the contralateral and ipsilateral ears, respectively.

Significance: Our research indicates that using a multielectrode stimulation approach enhances the detection rate of eSRT and slightly improves the correlation coefficient between eSRT and C-levels. It is found preferable to obtain measurements from the contralateral side due to its higher detection rate, although the ipsilateral side yields a marginally higher correlation coefficient. The findings suggest that eSRTs obtained from pulse width modulation implants could aid audiologists in programming the device.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Otology & Neurotology
Otology & Neurotology 医学-耳鼻喉科学
CiteScore
3.80
自引率
14.30%
发文量
509
审稿时长
3-6 weeks
期刊介绍: ​​​​​Otology & Neurotology publishes original articles relating to both clinical and basic science aspects of otology, neurotology, and cranial base surgery. As the foremost journal in its field, it has become the favored place for publishing the best of new science relating to the human ear and its diseases. The broadly international character of its contributing authors, editorial board, and readership provides the Journal its decidedly global perspective.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信