Borja Rivero-Santana, Carlos Galán, Claudia Pérez-Martínez, Borja Ibañez, Armando Pérez de Prado, María Fernández-Velasco, Raúl Moreno, Alfonso Jurado-Roman
{"title":"ELLIS 研究:通过扫描电子显微镜评估准分子激光冠状动脉血管成形术和血管内碎石术对药物洗脱支架的作用的比较分析。","authors":"Borja Rivero-Santana, Carlos Galán, Claudia Pérez-Martínez, Borja Ibañez, Armando Pérez de Prado, María Fernández-Velasco, Raúl Moreno, Alfonso Jurado-Roman","doi":"10.1161/CIRCINTERVENTIONS.124.014505","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Stent underexpansion is a significant challenge in percutaneous coronary intervention, critically impacting patient outcomes. While excimer laser coronary angioplasty (ELCA) and intravascular lithotripsy (IVL) are increasingly used to address this issue, their full impact on the integrity of drug-eluting stents remains unclear, raising concerns about their safety and efficacy.</p><p><strong>Methods: </strong>This in vitro study assessed the effects of ELCA and IVL on the structural integrity of drug-eluting stents using scanning electron microscopy. Nine stents, 5 Onyx Frontier (with durable circumferential polymer coating) and 4 Cre8 (polymer-free), were implanted in a 3-dimensional coronary artery simulator following standardized protocols. After implantation, treatments with saline-ELCA, contrast-ELCA, IVL, and high-pressure balloon dilatation were applied. A comprehensive evaluation of the stent surface was performed at 60-fold magnification.</p><p><strong>Results: </strong>Scanning electron microscopy analysis revealed significant differences in polymer damage between the techniques. High-pressure balloon dilatation and contrast-ELCA exhibited substantial polymer fragmentation and detachment compared with IVL, saline-ELCA, and conventional dilatation. High-pressure balloon dilatation demonstrated the highest incidence of polymer shaving and overcoating. No significant alterations were observed in polymer-free stents, regardless of the technique used.</p><p><strong>Conclusions: </strong>IVL and saline-ELCA applied immediately after stent implantation produce minimal polymer damage, whereas high-pressure balloon dilatation and contrast-ELCA cause significant damage to the polymer coating. The integrity of polymer-free drug-eluting stent appears stable regardless of the technique used. Further research is needed to validate these findings and explore their clinical implications.</p>","PeriodicalId":10330,"journal":{"name":"Circulation: Cardiovascular Interventions","volume":null,"pages":null},"PeriodicalIF":6.1000,"publicationDate":"2024-10-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"ELLIS Study: Comparative Analysis of Excimer Laser Coronary Angioplasty and Intravascular Lithotripsy on Drug-Eluting Stent as Assessed by Scanning Electron Microscopy.\",\"authors\":\"Borja Rivero-Santana, Carlos Galán, Claudia Pérez-Martínez, Borja Ibañez, Armando Pérez de Prado, María Fernández-Velasco, Raúl Moreno, Alfonso Jurado-Roman\",\"doi\":\"10.1161/CIRCINTERVENTIONS.124.014505\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Stent underexpansion is a significant challenge in percutaneous coronary intervention, critically impacting patient outcomes. While excimer laser coronary angioplasty (ELCA) and intravascular lithotripsy (IVL) are increasingly used to address this issue, their full impact on the integrity of drug-eluting stents remains unclear, raising concerns about their safety and efficacy.</p><p><strong>Methods: </strong>This in vitro study assessed the effects of ELCA and IVL on the structural integrity of drug-eluting stents using scanning electron microscopy. Nine stents, 5 Onyx Frontier (with durable circumferential polymer coating) and 4 Cre8 (polymer-free), were implanted in a 3-dimensional coronary artery simulator following standardized protocols. After implantation, treatments with saline-ELCA, contrast-ELCA, IVL, and high-pressure balloon dilatation were applied. A comprehensive evaluation of the stent surface was performed at 60-fold magnification.</p><p><strong>Results: </strong>Scanning electron microscopy analysis revealed significant differences in polymer damage between the techniques. High-pressure balloon dilatation and contrast-ELCA exhibited substantial polymer fragmentation and detachment compared with IVL, saline-ELCA, and conventional dilatation. High-pressure balloon dilatation demonstrated the highest incidence of polymer shaving and overcoating. No significant alterations were observed in polymer-free stents, regardless of the technique used.</p><p><strong>Conclusions: </strong>IVL and saline-ELCA applied immediately after stent implantation produce minimal polymer damage, whereas high-pressure balloon dilatation and contrast-ELCA cause significant damage to the polymer coating. The integrity of polymer-free drug-eluting stent appears stable regardless of the technique used. Further research is needed to validate these findings and explore their clinical implications.</p>\",\"PeriodicalId\":10330,\"journal\":{\"name\":\"Circulation: Cardiovascular Interventions\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":6.1000,\"publicationDate\":\"2024-10-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Circulation: Cardiovascular Interventions\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1161/CIRCINTERVENTIONS.124.014505\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Circulation: Cardiovascular Interventions","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1161/CIRCINTERVENTIONS.124.014505","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
ELLIS Study: Comparative Analysis of Excimer Laser Coronary Angioplasty and Intravascular Lithotripsy on Drug-Eluting Stent as Assessed by Scanning Electron Microscopy.
Background: Stent underexpansion is a significant challenge in percutaneous coronary intervention, critically impacting patient outcomes. While excimer laser coronary angioplasty (ELCA) and intravascular lithotripsy (IVL) are increasingly used to address this issue, their full impact on the integrity of drug-eluting stents remains unclear, raising concerns about their safety and efficacy.
Methods: This in vitro study assessed the effects of ELCA and IVL on the structural integrity of drug-eluting stents using scanning electron microscopy. Nine stents, 5 Onyx Frontier (with durable circumferential polymer coating) and 4 Cre8 (polymer-free), were implanted in a 3-dimensional coronary artery simulator following standardized protocols. After implantation, treatments with saline-ELCA, contrast-ELCA, IVL, and high-pressure balloon dilatation were applied. A comprehensive evaluation of the stent surface was performed at 60-fold magnification.
Results: Scanning electron microscopy analysis revealed significant differences in polymer damage between the techniques. High-pressure balloon dilatation and contrast-ELCA exhibited substantial polymer fragmentation and detachment compared with IVL, saline-ELCA, and conventional dilatation. High-pressure balloon dilatation demonstrated the highest incidence of polymer shaving and overcoating. No significant alterations were observed in polymer-free stents, regardless of the technique used.
Conclusions: IVL and saline-ELCA applied immediately after stent implantation produce minimal polymer damage, whereas high-pressure balloon dilatation and contrast-ELCA cause significant damage to the polymer coating. The integrity of polymer-free drug-eluting stent appears stable regardless of the technique used. Further research is needed to validate these findings and explore their clinical implications.
期刊介绍:
Circulation: Cardiovascular Interventions, an American Heart Association journal, focuses on interventional techniques pertaining to coronary artery disease, structural heart disease, and vascular disease, with priority placed on original research and on randomized trials and large registry studies. In addition, pharmacological, diagnostic, and pathophysiological aspects of interventional cardiology are given special attention in this online-only journal.