利用探针萃取与液相色谱-高分辨质谱联用技术检测未修饰的单克隆血清游离光链以确定克隆性

IF 7.1 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Priscilla S W Yeung, Yajing Liu, Samuel Yang, Ashley Ruan, Christina R Kerr, Carolyn V Wong, Run-Zhang Shi, David J Iberri, Ruben Y Luo
{"title":"利用探针萃取与液相色谱-高分辨质谱联用技术检测未修饰的单克隆血清游离光链以确定克隆性","authors":"Priscilla S W Yeung, Yajing Liu, Samuel Yang, Ashley Ruan, Christina R Kerr, Carolyn V Wong, Run-Zhang Shi, David J Iberri, Ruben Y Luo","doi":"10.1093/clinchem/hvae130","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Serum free light chains (FLCs) are an essential clinical biomarker for the diagnosis and monitoring of patients with plasma cell neoplasms. The current widely used immunoassay methods quantify total serum FLCs, which include monoclonal FLCs as well as FLCs in the polyclonal background. Patients with chronic diseases, inflammatory disorders, or renal dysfunction can have elevated total FLCs that lead to ambiguous results. These patients may benefit from a direct measurement of monoclonal FLCs. The purpose of this study was to develop a method that couples on-probe extraction (OPEX) with liquid chromatography-high-resolution mass spectrometry (LC-HR-MS), abbreviated to OPEX-MS, to directly determine the clonality of FLCs.</p><p><strong>Methods: </strong>OPEX immunocapture was performed using microprobes loaded with anti-kappa or anti-lambda light chain antibodies. Captured proteins were separated by reversed-phase LC and analyzed by HR-MS.</p><p><strong>Results: </strong>Four cohorts of samples from unique patients were tested based on immunoassay FLC results. The LC-HR-MS analysis in the OPEX-MS method provides both a unique retention time along with deconvoluted masses of FLC monomers and dimers for each clone. The study found that 16 out of 49 (33%) kappa FLC elevated samples as well as 83 out of 100 (83%) dual kappa and lambda FLC elevated samples did not have monoclonal FLCs, which is consistent with the knowledge that there is often no clonal population in samples with mildly elevated FLC immunoassay results.</p><p><strong>Conclusions: </strong>The OPEX-MS method can serve as a complementary approach to directly determine clonality in patients with difficult-to-interpret FLC immunoassay results.</p>","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":" ","pages":""},"PeriodicalIF":7.1000,"publicationDate":"2024-10-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Clonality Determination by Detecting Unmodified Monoclonal Serum Free Light Chains Using On-Probe Extraction Coupled with Liquid Chromatography-High-Resolution Mass Spectrometry.\",\"authors\":\"Priscilla S W Yeung, Yajing Liu, Samuel Yang, Ashley Ruan, Christina R Kerr, Carolyn V Wong, Run-Zhang Shi, David J Iberri, Ruben Y Luo\",\"doi\":\"10.1093/clinchem/hvae130\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Serum free light chains (FLCs) are an essential clinical biomarker for the diagnosis and monitoring of patients with plasma cell neoplasms. The current widely used immunoassay methods quantify total serum FLCs, which include monoclonal FLCs as well as FLCs in the polyclonal background. Patients with chronic diseases, inflammatory disorders, or renal dysfunction can have elevated total FLCs that lead to ambiguous results. These patients may benefit from a direct measurement of monoclonal FLCs. The purpose of this study was to develop a method that couples on-probe extraction (OPEX) with liquid chromatography-high-resolution mass spectrometry (LC-HR-MS), abbreviated to OPEX-MS, to directly determine the clonality of FLCs.</p><p><strong>Methods: </strong>OPEX immunocapture was performed using microprobes loaded with anti-kappa or anti-lambda light chain antibodies. Captured proteins were separated by reversed-phase LC and analyzed by HR-MS.</p><p><strong>Results: </strong>Four cohorts of samples from unique patients were tested based on immunoassay FLC results. The LC-HR-MS analysis in the OPEX-MS method provides both a unique retention time along with deconvoluted masses of FLC monomers and dimers for each clone. The study found that 16 out of 49 (33%) kappa FLC elevated samples as well as 83 out of 100 (83%) dual kappa and lambda FLC elevated samples did not have monoclonal FLCs, which is consistent with the knowledge that there is often no clonal population in samples with mildly elevated FLC immunoassay results.</p><p><strong>Conclusions: </strong>The OPEX-MS method can serve as a complementary approach to directly determine clonality in patients with difficult-to-interpret FLC immunoassay results.</p>\",\"PeriodicalId\":10690,\"journal\":{\"name\":\"Clinical chemistry\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":7.1000,\"publicationDate\":\"2024-10-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical chemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/clinchem/hvae130\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/clinchem/hvae130","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景:血清游离轻链(FLCs)是诊断和监测浆细胞肿瘤患者的重要临床生物标志物。目前广泛使用的免疫测定方法可定量检测血清总游离轻链,其中包括单克隆游离轻链和多克隆游离轻链。慢性疾病、炎症性疾病或肾功能不全患者的总 FLCs 可能会升高,从而导致结果不明确。直接测量单克隆 FLCs 可能会使这些患者受益。本研究旨在开发一种将探针提取(OPEX)与液相色谱-高分辨质谱(LC-HR-MS)(简称 OPEX-MS)相结合的方法,以直接测定 FLCs 的克隆性:方法:使用装有抗卡帕或抗λ轻链抗体的微探针进行 OPEX 免疫捕获。捕获的蛋白质通过反相液相色谱进行分离,并通过 HR-MS 进行分析:根据免疫测定 FLC 的结果,对来自不同患者的四组样本进行了检测。OPEX-MS 方法中的 LC-HR-MS 分析为每个克隆提供了独特的保留时间以及 FLC 单体和二聚体的去卷积质量。研究发现,49 份 kappa FLC 升高的样本中有 16 份(33%)以及 100 份 kappa 和 lambda FLC 双升高的样本中有 83 份(83%)没有单克隆 FLC,这与 FLC 免疫测定结果轻度升高的样本中通常没有克隆群的知识是一致的:OPEX-MS方法可作为一种补充方法,直接确定FLC免疫测定结果难以解释的患者的克隆性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Clonality Determination by Detecting Unmodified Monoclonal Serum Free Light Chains Using On-Probe Extraction Coupled with Liquid Chromatography-High-Resolution Mass Spectrometry.

Background: Serum free light chains (FLCs) are an essential clinical biomarker for the diagnosis and monitoring of patients with plasma cell neoplasms. The current widely used immunoassay methods quantify total serum FLCs, which include monoclonal FLCs as well as FLCs in the polyclonal background. Patients with chronic diseases, inflammatory disorders, or renal dysfunction can have elevated total FLCs that lead to ambiguous results. These patients may benefit from a direct measurement of monoclonal FLCs. The purpose of this study was to develop a method that couples on-probe extraction (OPEX) with liquid chromatography-high-resolution mass spectrometry (LC-HR-MS), abbreviated to OPEX-MS, to directly determine the clonality of FLCs.

Methods: OPEX immunocapture was performed using microprobes loaded with anti-kappa or anti-lambda light chain antibodies. Captured proteins were separated by reversed-phase LC and analyzed by HR-MS.

Results: Four cohorts of samples from unique patients were tested based on immunoassay FLC results. The LC-HR-MS analysis in the OPEX-MS method provides both a unique retention time along with deconvoluted masses of FLC monomers and dimers for each clone. The study found that 16 out of 49 (33%) kappa FLC elevated samples as well as 83 out of 100 (83%) dual kappa and lambda FLC elevated samples did not have monoclonal FLCs, which is consistent with the knowledge that there is often no clonal population in samples with mildly elevated FLC immunoassay results.

Conclusions: The OPEX-MS method can serve as a complementary approach to directly determine clonality in patients with difficult-to-interpret FLC immunoassay results.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical chemistry
Clinical chemistry 医学-医学实验技术
CiteScore
11.30
自引率
4.30%
发文量
212
审稿时长
1.7 months
期刊介绍: Clinical Chemistry is a peer-reviewed scientific journal that is the premier publication for the science and practice of clinical laboratory medicine. It was established in 1955 and is associated with the Association for Diagnostics & Laboratory Medicine (ADLM). The journal focuses on laboratory diagnosis and management of patients, and has expanded to include other clinical laboratory disciplines such as genomics, hematology, microbiology, and toxicology. It also publishes articles relevant to clinical specialties including cardiology, endocrinology, gastroenterology, genetics, immunology, infectious diseases, maternal-fetal medicine, neurology, nutrition, oncology, and pediatrics. In addition to original research, editorials, and reviews, Clinical Chemistry features recurring sections such as clinical case studies, perspectives, podcasts, and Q&A articles. It has the highest impact factor among journals of clinical chemistry, laboratory medicine, pathology, analytical chemistry, transfusion medicine, and clinical microbiology. The journal is indexed in databases such as MEDLINE and Web of Science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信