A-345 溶血对常规血气分析物的影响

IF 7.1 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
B Agana, B Overton, K Florendo, C Knezevic
{"title":"A-345 溶血对常规血气分析物的影响","authors":"B Agana, B Overton, K Florendo, C Knezevic","doi":"10.1093/clinchem/hvae106.339","DOIUrl":null,"url":null,"abstract":"Background Hemolysis is a major pre-analytical concern for most laboratory analytes. Detection of hemolysis and mitigation efforts are especially important for analytes, such as potassium, with high intracellular concentrations. Routine serum and plasma chemistry tests are performed on analyzers with the capacity to detect and measure the degree of hemolysis. However, for whole blood chemistries and blood gas measurements, the instruments utilized lack this capacity. The aim of this study was to evaluate the effect of hemolysis on whole blood and blood gas analytes and to compare visual assessments of hemolysis to measured hemolysis indices (H-index). Methods Remnant whole blood samples were split into two portions and each portion was aspirated and dispensed through a syringe one or more times. For the mock portion, only the syringe was used, while the hemolyzed portion had a needle affixed to the end of the syringe. The needle provided shear stress on the red blood cells to induce hemolysis, while the mock procedure was used to assess the impact of aspirating/dispensing on the sample in the absence of hemolysis. Each portion was then analyzed on a Radiometer ABL800 series instrument, spun down, and the H-index of the plasma portion was measured on a Roche cobas 8000 instrument. Medical technologists recorded their visual assessment of the specimens, with two technologists agreeing to the categorization of the specimen as either slightly, moderately, or severely hemolyzed. Degree of hemolysis was categorized by the delta (hemolyzed - mock) of the measured H-index: slight hemolysis was defined as an H-index delta of <100, moderate as 100-500 and severe as >500. Results The effect of hemolysis, with H-indices ranging from 2 to 3861, on 13 routine blood gas analytes was studied for 85 whole blood specimens. Hemolysis had little effect on metabolites, as percent bias was within ±3% at all levels of hemolysis for glucose, creatinine, and lactate. Similarly, most cooximetry components were minimally affected, with total hemoglobin, oxyhemoglobin, carboxyhemoglobin, and oxygen saturation within ±5% bias at all levels of hemolysis. Methemoglobin had a larger overall negative bias, with slight, moderate, and severe hemolysis levels yielding percent biases of -6.6, -12.3, and -13.3%, respectively. As expected, potassium displayed a significant positive bias with increasing hemolysis, with a generally linear trend. At moderate levels of hemolysis, the average potassium bias was 24.0% and at severe levels, over 100%. Sodium and ionized calcium also displayed overall linear trends but with a negative bias. At slight, moderate, and severe levels of hemolysis, sodium had a -0.56, -1.10, and -3.96% bias, and ionized calcium had a -2.99, -5.65, and -15.5% bias respectively. Conclusions Hemolysis can falsely increase or decrease a range of blood gas analytes and lead to misinterpretation of results and adversely affect clinical decision-making. Therefore, equipping current blood gas analyzers with hemolysis detection capabilities is crucial to enable laboratories to mitigate this effect and ensure accurate results for patient care.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"4 1","pages":""},"PeriodicalIF":7.1000,"publicationDate":"2024-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A-345 Effect of Hemolysis on Routine Blood Gas Analytes\",\"authors\":\"B Agana, B Overton, K Florendo, C Knezevic\",\"doi\":\"10.1093/clinchem/hvae106.339\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background Hemolysis is a major pre-analytical concern for most laboratory analytes. Detection of hemolysis and mitigation efforts are especially important for analytes, such as potassium, with high intracellular concentrations. Routine serum and plasma chemistry tests are performed on analyzers with the capacity to detect and measure the degree of hemolysis. However, for whole blood chemistries and blood gas measurements, the instruments utilized lack this capacity. The aim of this study was to evaluate the effect of hemolysis on whole blood and blood gas analytes and to compare visual assessments of hemolysis to measured hemolysis indices (H-index). Methods Remnant whole blood samples were split into two portions and each portion was aspirated and dispensed through a syringe one or more times. For the mock portion, only the syringe was used, while the hemolyzed portion had a needle affixed to the end of the syringe. The needle provided shear stress on the red blood cells to induce hemolysis, while the mock procedure was used to assess the impact of aspirating/dispensing on the sample in the absence of hemolysis. Each portion was then analyzed on a Radiometer ABL800 series instrument, spun down, and the H-index of the plasma portion was measured on a Roche cobas 8000 instrument. Medical technologists recorded their visual assessment of the specimens, with two technologists agreeing to the categorization of the specimen as either slightly, moderately, or severely hemolyzed. Degree of hemolysis was categorized by the delta (hemolyzed - mock) of the measured H-index: slight hemolysis was defined as an H-index delta of <100, moderate as 100-500 and severe as >500. Results The effect of hemolysis, with H-indices ranging from 2 to 3861, on 13 routine blood gas analytes was studied for 85 whole blood specimens. Hemolysis had little effect on metabolites, as percent bias was within ±3% at all levels of hemolysis for glucose, creatinine, and lactate. Similarly, most cooximetry components were minimally affected, with total hemoglobin, oxyhemoglobin, carboxyhemoglobin, and oxygen saturation within ±5% bias at all levels of hemolysis. Methemoglobin had a larger overall negative bias, with slight, moderate, and severe hemolysis levels yielding percent biases of -6.6, -12.3, and -13.3%, respectively. As expected, potassium displayed a significant positive bias with increasing hemolysis, with a generally linear trend. At moderate levels of hemolysis, the average potassium bias was 24.0% and at severe levels, over 100%. Sodium and ionized calcium also displayed overall linear trends but with a negative bias. At slight, moderate, and severe levels of hemolysis, sodium had a -0.56, -1.10, and -3.96% bias, and ionized calcium had a -2.99, -5.65, and -15.5% bias respectively. Conclusions Hemolysis can falsely increase or decrease a range of blood gas analytes and lead to misinterpretation of results and adversely affect clinical decision-making. Therefore, equipping current blood gas analyzers with hemolysis detection capabilities is crucial to enable laboratories to mitigate this effect and ensure accurate results for patient care.\",\"PeriodicalId\":10690,\"journal\":{\"name\":\"Clinical chemistry\",\"volume\":\"4 1\",\"pages\":\"\"},\"PeriodicalIF\":7.1000,\"publicationDate\":\"2024-10-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical chemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/clinchem/hvae106.339\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/clinchem/hvae106.339","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景溶血是大多数实验室分析物分析前的主要问题。对于钾等细胞内浓度较高的分析物而言,溶血的检测和缓解尤为重要。常规血清和血浆化学检测是在具有检测和测量溶血程度功能的分析仪上进行的。然而,在进行全血化学和血气测量时,所使用的仪器却缺乏这种能力。本研究的目的是评估溶血对全血和血气分析物的影响,并将溶血的肉眼评估与溶血指数(H-index)的测量结果进行比较。方法 将残留全血样本分成两份,每份样本都用注射器抽吸并分配一次或多次。模拟部分只使用注射器,而溶血部分则在注射器末端粘贴一根针。针头对红细胞产生剪切应力以诱导溶血,而模拟程序则用于评估在不溶血的情况下抽吸/分配对样本的影响。然后用 Radiometer ABL800 系列仪器对每份样本进行分析、离心,并用 Roche cobas 8000 仪器测量血浆部分的 H 指数。医疗技术人员记录他们对标本的目测评估,由两名技术人员一致同意将标本分为轻微溶血、中度溶血或严重溶血。溶血程度根据测得的 H 指数δ(溶血-模拟)进行分类:H 指数δ为<100 时为轻微溶血,100-500 时为中度溶血,500 时为重度溶血。结果 研究了 85 份全血标本中溶血(H 指数从 2 到 3861 不等)对 13 种常规血气分析物的影响。溶血对代谢物的影响很小,因为在所有溶血水平下,葡萄糖、肌酐和乳酸盐的偏差百分比都在±3%以内。同样,大多数共氧血红蛋白成分受到的影响也很小,总血红蛋白、氧合血红蛋白、碳氧血红蛋白和血氧饱和度在所有溶血程度下的偏差都在±5%以内。高铁血红蛋白的总体负偏差较大,轻度、中度和重度溶血水平的偏差百分比分别为-6.6%、-12.3%和-13.3%。不出所料,随着溶血量的增加,钾的偏倚率呈显著的正值,且总体呈线性趋势。中度溶血时,钾的平均偏差为 24.0%,严重溶血时超过 100%。钠和离子钙也呈总体线性趋势,但偏差为负。在轻度、中度和重度溶血时,钠的偏差分别为-0.56%、-1.10%和-3.96%,离子钙的偏差分别为-2.99%、-5.65%和-15.5%。结论 溶血会使一系列血气分析物发生错误的增减,导致对结果的误读,并对临床决策产生不利影响。因此,在当前的血气分析仪上配备溶血检测功能至关重要,可使实验室减轻这种影响,确保为患者提供准确的结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A-345 Effect of Hemolysis on Routine Blood Gas Analytes
Background Hemolysis is a major pre-analytical concern for most laboratory analytes. Detection of hemolysis and mitigation efforts are especially important for analytes, such as potassium, with high intracellular concentrations. Routine serum and plasma chemistry tests are performed on analyzers with the capacity to detect and measure the degree of hemolysis. However, for whole blood chemistries and blood gas measurements, the instruments utilized lack this capacity. The aim of this study was to evaluate the effect of hemolysis on whole blood and blood gas analytes and to compare visual assessments of hemolysis to measured hemolysis indices (H-index). Methods Remnant whole blood samples were split into two portions and each portion was aspirated and dispensed through a syringe one or more times. For the mock portion, only the syringe was used, while the hemolyzed portion had a needle affixed to the end of the syringe. The needle provided shear stress on the red blood cells to induce hemolysis, while the mock procedure was used to assess the impact of aspirating/dispensing on the sample in the absence of hemolysis. Each portion was then analyzed on a Radiometer ABL800 series instrument, spun down, and the H-index of the plasma portion was measured on a Roche cobas 8000 instrument. Medical technologists recorded their visual assessment of the specimens, with two technologists agreeing to the categorization of the specimen as either slightly, moderately, or severely hemolyzed. Degree of hemolysis was categorized by the delta (hemolyzed - mock) of the measured H-index: slight hemolysis was defined as an H-index delta of <100, moderate as 100-500 and severe as >500. Results The effect of hemolysis, with H-indices ranging from 2 to 3861, on 13 routine blood gas analytes was studied for 85 whole blood specimens. Hemolysis had little effect on metabolites, as percent bias was within ±3% at all levels of hemolysis for glucose, creatinine, and lactate. Similarly, most cooximetry components were minimally affected, with total hemoglobin, oxyhemoglobin, carboxyhemoglobin, and oxygen saturation within ±5% bias at all levels of hemolysis. Methemoglobin had a larger overall negative bias, with slight, moderate, and severe hemolysis levels yielding percent biases of -6.6, -12.3, and -13.3%, respectively. As expected, potassium displayed a significant positive bias with increasing hemolysis, with a generally linear trend. At moderate levels of hemolysis, the average potassium bias was 24.0% and at severe levels, over 100%. Sodium and ionized calcium also displayed overall linear trends but with a negative bias. At slight, moderate, and severe levels of hemolysis, sodium had a -0.56, -1.10, and -3.96% bias, and ionized calcium had a -2.99, -5.65, and -15.5% bias respectively. Conclusions Hemolysis can falsely increase or decrease a range of blood gas analytes and lead to misinterpretation of results and adversely affect clinical decision-making. Therefore, equipping current blood gas analyzers with hemolysis detection capabilities is crucial to enable laboratories to mitigate this effect and ensure accurate results for patient care.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical chemistry
Clinical chemistry 医学-医学实验技术
CiteScore
11.30
自引率
4.30%
发文量
212
审稿时长
1.7 months
期刊介绍: Clinical Chemistry is a peer-reviewed scientific journal that is the premier publication for the science and practice of clinical laboratory medicine. It was established in 1955 and is associated with the Association for Diagnostics & Laboratory Medicine (ADLM). The journal focuses on laboratory diagnosis and management of patients, and has expanded to include other clinical laboratory disciplines such as genomics, hematology, microbiology, and toxicology. It also publishes articles relevant to clinical specialties including cardiology, endocrinology, gastroenterology, genetics, immunology, infectious diseases, maternal-fetal medicine, neurology, nutrition, oncology, and pediatrics. In addition to original research, editorials, and reviews, Clinical Chemistry features recurring sections such as clinical case studies, perspectives, podcasts, and Q&A articles. It has the highest impact factor among journals of clinical chemistry, laboratory medicine, pathology, analytical chemistry, transfusion medicine, and clinical microbiology. The journal is indexed in databases such as MEDLINE and Web of Science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信