使用戈德曼眼压计、iCare 眼压计和气囊式眼压计测量眼压的差异及其与中央角膜厚度相关性的横断面研究。

IF 1.2 Q3 OPHTHALMOLOGY
Journal of Current Ophthalmology Pub Date : 2024-08-10 eCollection Date: 2023-10-01 DOI:10.4103/joco.joco_180_23
Sameh Saad Mandour, Ahmed Elframawy, Mohammad Moataz Murad, Sara Abd Elmegeed Nage
{"title":"使用戈德曼眼压计、iCare 眼压计和气囊式眼压计测量眼压的差异及其与中央角膜厚度相关性的横断面研究。","authors":"Sameh Saad Mandour, Ahmed Elframawy, Mohammad Moataz Murad, Sara Abd Elmegeed Nage","doi":"10.4103/joco.joco_180_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To investigate the agreement between the Goldmann tonometer (GAT), the air-puff tonometer, and the iCare tonometer in intraocular pressure (IOP) evaluation as well as their association with central corneal thickness (CCT) in normal participants, glaucoma patients, and patients following refractive surgery.</p><p><strong>Methods: </strong>This is a cross-sectional study conducted on 204 eyes from 102 patients. The study consisted of three equal groups: group I (control group, <i>n</i> = 34), group II (glaucoma patients on medication, <i>n</i> = 34), and group III (refractive surgery patients, <i>n</i> = 34). All patients were subjected to examination (complete ocular examination, refraction, and IOP measurement).</p><p><strong>Results: </strong>A total of 102 participants were included in the study with both genders distributed equally. The mean ± standard deviation age was 44.12 ± 12.8 years in the control group while it was 46.29 ± 13.24 years in the glaucoma group and 40.68 ± 15.86 years in the refractive surgery group. Overall, there was a high correlation between the three methods. The mean IOP measured by GAT was 14.03 ± 3.43. The mean IOP measured by iCare was 15.16 ± 3.46. The mean IOP measured by air-puff was 16.66 ± 3.6. The iCare showed the most significant agreement with the GAT (intraclass correlation coefficient [ICC] 0.985, <i>P</i> > 0.05) and the mean difference in IOP between GAT and iCare was 1.1 (95% limits of agreement, -0.62-+2.85 mmHg). The mean difference in IOP between iCare and air-puff was 1.5 and it was 2.6 between GAT and air-puff. There were no significant differences in IOP measurements between GAT and iCare tonometer or between iCare tonometer and air-puff in all groups (<i>P</i> > 0.05). However, there were significant differences in IOP measurements between GAT and air-puff in all groups (<i>P</i> < 0.001). The ICC between all studied methods was strong (ICC > 0.92 for all). Regarding CCT, the mean CCT was 517.14 ± 29.82 μm. There were significant positive correlations between increasing CCT and increasing IOP by GAT, iCare, and air-puff tonometer among the three groups (<i>P</i> < 0.001).</p><p><strong>Conclusions: </strong>In conclusion, the iCare tonometer, specifically the iCare PRO RT model, is a reliable and efficient alternative instrument for assessing IOP. The IOP values obtained with the iCare PRO RT were found to be consistent with those obtained using the air-puff and GAT.</p>","PeriodicalId":15423,"journal":{"name":"Journal of Current Ophthalmology","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-08-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11392305/pdf/","citationCount":"0","resultStr":"{\"title\":\"Cross-Sectional Study of Differences between Intraocular Pressure Measurements using Goldmann, iCare, and Air-Puff Tonometers and their Correlation with Central Corneal Thickness.\",\"authors\":\"Sameh Saad Mandour, Ahmed Elframawy, Mohammad Moataz Murad, Sara Abd Elmegeed Nage\",\"doi\":\"10.4103/joco.joco_180_23\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To investigate the agreement between the Goldmann tonometer (GAT), the air-puff tonometer, and the iCare tonometer in intraocular pressure (IOP) evaluation as well as their association with central corneal thickness (CCT) in normal participants, glaucoma patients, and patients following refractive surgery.</p><p><strong>Methods: </strong>This is a cross-sectional study conducted on 204 eyes from 102 patients. The study consisted of three equal groups: group I (control group, <i>n</i> = 34), group II (glaucoma patients on medication, <i>n</i> = 34), and group III (refractive surgery patients, <i>n</i> = 34). All patients were subjected to examination (complete ocular examination, refraction, and IOP measurement).</p><p><strong>Results: </strong>A total of 102 participants were included in the study with both genders distributed equally. The mean ± standard deviation age was 44.12 ± 12.8 years in the control group while it was 46.29 ± 13.24 years in the glaucoma group and 40.68 ± 15.86 years in the refractive surgery group. Overall, there was a high correlation between the three methods. The mean IOP measured by GAT was 14.03 ± 3.43. The mean IOP measured by iCare was 15.16 ± 3.46. The mean IOP measured by air-puff was 16.66 ± 3.6. The iCare showed the most significant agreement with the GAT (intraclass correlation coefficient [ICC] 0.985, <i>P</i> > 0.05) and the mean difference in IOP between GAT and iCare was 1.1 (95% limits of agreement, -0.62-+2.85 mmHg). The mean difference in IOP between iCare and air-puff was 1.5 and it was 2.6 between GAT and air-puff. There were no significant differences in IOP measurements between GAT and iCare tonometer or between iCare tonometer and air-puff in all groups (<i>P</i> > 0.05). However, there were significant differences in IOP measurements between GAT and air-puff in all groups (<i>P</i> < 0.001). The ICC between all studied methods was strong (ICC > 0.92 for all). Regarding CCT, the mean CCT was 517.14 ± 29.82 μm. There were significant positive correlations between increasing CCT and increasing IOP by GAT, iCare, and air-puff tonometer among the three groups (<i>P</i> < 0.001).</p><p><strong>Conclusions: </strong>In conclusion, the iCare tonometer, specifically the iCare PRO RT model, is a reliable and efficient alternative instrument for assessing IOP. The IOP values obtained with the iCare PRO RT were found to be consistent with those obtained using the air-puff and GAT.</p>\",\"PeriodicalId\":15423,\"journal\":{\"name\":\"Journal of Current Ophthalmology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2024-08-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11392305/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Current Ophthalmology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/joco.joco_180_23\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/10/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q3\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Current Ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/joco.joco_180_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:研究在正常人、青光眼患者和屈光手术后患者中,戈德曼眼压计(GAT)、气囊式眼压计和 iCare 眼压计在眼压(IOP)评估中的一致性,以及它们与中央角膜厚度(CCT)的关联:这是一项横断面研究,研究对象是 102 名患者的 204 只眼睛。研究分为三组:第一组(对照组,34 人)、第二组(青光眼药物治疗患者,34 人)和第三组(屈光手术患者,34 人)。所有患者均接受了检查(眼部全面检查、屈光度和眼压测量):共有 102 人参与了研究,男女比例相当。对照组的平均年龄(标准差)为 44.12 ± 12.8 岁,青光眼组为 46.29 ± 13.24 岁,屈光手术组为 40.68 ± 15.86 岁。总体而言,三种方法之间存在高度相关性。GAT 测量的平均眼压为 14.03 ± 3.43。iCare 测量的平均眼压为 15.16 ± 3.46。气吹法测量的平均眼压为(16.66 ± 3.6)。iCare 与 GAT 的一致性最为显著(类内相关系数 [ICC] 0.985,P > 0.05),GAT 与 iCare 之间的平均眼压差为 1.1(95% 的一致性范围为 -0.62-+2.85 mmHg)。iCare 和空气吹气法的平均眼压差为 1.5,GAT 和空气吹气法的平均眼压差为 2.6。在所有组别中,GAT 与 iCare 眼压计之间或 iCare 眼压计与气囊式眼压计之间的眼压测量值均无明显差异(P > 0.05)。然而,在所有组别中,GAT 和气囊法测定的眼压有明显差异(P < 0.001)。所有研究方法之间的 ICC 都很高(均大于 0.92)。关于 CCT,平均 CCT 为 517.14 ± 29.82 μm。通过 GAT、iCare 和气囊式眼压计测量,三组 CCT 的增加与 IOP 的增加之间存在明显的正相关性(P < 0.001):总之,iCare 眼压计,特别是 iCare PRO RT 型号,是评估眼压的一种可靠、高效的替代仪器。使用 iCare PRO RT 得出的眼压值与使用气囊式眼压计和 GAT 得出的眼压值一致。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Cross-Sectional Study of Differences between Intraocular Pressure Measurements using Goldmann, iCare, and Air-Puff Tonometers and their Correlation with Central Corneal Thickness.

Purpose: To investigate the agreement between the Goldmann tonometer (GAT), the air-puff tonometer, and the iCare tonometer in intraocular pressure (IOP) evaluation as well as their association with central corneal thickness (CCT) in normal participants, glaucoma patients, and patients following refractive surgery.

Methods: This is a cross-sectional study conducted on 204 eyes from 102 patients. The study consisted of three equal groups: group I (control group, n = 34), group II (glaucoma patients on medication, n = 34), and group III (refractive surgery patients, n = 34). All patients were subjected to examination (complete ocular examination, refraction, and IOP measurement).

Results: A total of 102 participants were included in the study with both genders distributed equally. The mean ± standard deviation age was 44.12 ± 12.8 years in the control group while it was 46.29 ± 13.24 years in the glaucoma group and 40.68 ± 15.86 years in the refractive surgery group. Overall, there was a high correlation between the three methods. The mean IOP measured by GAT was 14.03 ± 3.43. The mean IOP measured by iCare was 15.16 ± 3.46. The mean IOP measured by air-puff was 16.66 ± 3.6. The iCare showed the most significant agreement with the GAT (intraclass correlation coefficient [ICC] 0.985, P > 0.05) and the mean difference in IOP between GAT and iCare was 1.1 (95% limits of agreement, -0.62-+2.85 mmHg). The mean difference in IOP between iCare and air-puff was 1.5 and it was 2.6 between GAT and air-puff. There were no significant differences in IOP measurements between GAT and iCare tonometer or between iCare tonometer and air-puff in all groups (P > 0.05). However, there were significant differences in IOP measurements between GAT and air-puff in all groups (P < 0.001). The ICC between all studied methods was strong (ICC > 0.92 for all). Regarding CCT, the mean CCT was 517.14 ± 29.82 μm. There were significant positive correlations between increasing CCT and increasing IOP by GAT, iCare, and air-puff tonometer among the three groups (P < 0.001).

Conclusions: In conclusion, the iCare tonometer, specifically the iCare PRO RT model, is a reliable and efficient alternative instrument for assessing IOP. The IOP values obtained with the iCare PRO RT were found to be consistent with those obtained using the air-puff and GAT.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
2.50
自引率
6.70%
发文量
45
审稿时长
8 weeks
期刊介绍: Peer Review under the responsibility of Iranian Society of Ophthalmology Journal of Current Ophthalmology, the official publication of the Iranian Society of Ophthalmology, is a peer-reviewed, open-access, scientific journal that welcomes high quality original articles related to vision science and all fields of ophthalmology. Journal of Current Ophthalmology is the continuum of Iranian Journal of Ophthalmology published since 1969.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信