{"title":"使用 3 毫米磁共振兼容导管进行间隙双模超声,用于图像引导 HIFU 和定向体外组织消融。","authors":"Thomas Biscaldi, Romain L'Huillier, Laurent Milot, W Apoutou N'Djin","doi":"10.1109/TUFFC.2024.3458067","DOIUrl":null,"url":null,"abstract":"<p><p>Current interstitial techniques of tumor ablation face challenges that ultrasound technologies could meet. The ablation radius and directionality of the ultrasound beam could improve the efficiency and precision. Here, a 9-gauge MR-compatible dual-mode ultrasound catheter prototype was experimentally evaluated for Ultrasound Image-guided High Intensity Focused Ultrasound (USgHIFU) conformal ablations. The prototype consisted of 64 piezocomposite linear array elements and was driven by an open research programmable dual-mode ultrasound platform. After verifying the US-image guidance capabilities of the prototype, the HIFU output performances (dynamic focusing and HIFU intensities) were quantitatively characterized, together with the associated 3D HIFU-induced thermal heating in tissue phantoms (using MR thermometry). Finally, the ability to produce robustly HIFU-induced thermal ablations in in-vitro liver was studied experimentally and compared to numerical modeling. Investigations of several HIFU dynamic focusing allowed overcoming the challenges of miniaturizing the device: mono-focal focusing maximized deep energy deposition, while multi-focal strategies eliminated grating lobes. The linear-array design of the prototype made it possible to produce interstitial ultrasound images of tissue and tumor mimics in situ. Multi-focal pressure fields were generated without grating lobes and transducer surface intensities reached up to Isapa = 14 W·cm<sup>-2</sup>. Seventeen elementary thermal ablations were performed in vitro. Rotation of the catheter proved the directionality of ablation, sparing non-targeted tissue. This experimental proof of concept demonstrates the feasibility of treating volumes comparable to those of primary solid tumors with a miniaturized USgHIFU catheter whose dimensions are close to those of tools traditionally used in interventional radiology, while offering new functionalities.</p>","PeriodicalId":13322,"journal":{"name":"IEEE transactions on ultrasonics, ferroelectrics, and frequency control","volume":"PP ","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2024-09-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Interstitial dual-mode ultrasound with a 3-mm MR-compatible catheter for image-guided HIFU and directional in-vitro tissue ablations.\",\"authors\":\"Thomas Biscaldi, Romain L'Huillier, Laurent Milot, W Apoutou N'Djin\",\"doi\":\"10.1109/TUFFC.2024.3458067\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Current interstitial techniques of tumor ablation face challenges that ultrasound technologies could meet. The ablation radius and directionality of the ultrasound beam could improve the efficiency and precision. Here, a 9-gauge MR-compatible dual-mode ultrasound catheter prototype was experimentally evaluated for Ultrasound Image-guided High Intensity Focused Ultrasound (USgHIFU) conformal ablations. The prototype consisted of 64 piezocomposite linear array elements and was driven by an open research programmable dual-mode ultrasound platform. After verifying the US-image guidance capabilities of the prototype, the HIFU output performances (dynamic focusing and HIFU intensities) were quantitatively characterized, together with the associated 3D HIFU-induced thermal heating in tissue phantoms (using MR thermometry). Finally, the ability to produce robustly HIFU-induced thermal ablations in in-vitro liver was studied experimentally and compared to numerical modeling. Investigations of several HIFU dynamic focusing allowed overcoming the challenges of miniaturizing the device: mono-focal focusing maximized deep energy deposition, while multi-focal strategies eliminated grating lobes. The linear-array design of the prototype made it possible to produce interstitial ultrasound images of tissue and tumor mimics in situ. Multi-focal pressure fields were generated without grating lobes and transducer surface intensities reached up to Isapa = 14 W·cm<sup>-2</sup>. Seventeen elementary thermal ablations were performed in vitro. Rotation of the catheter proved the directionality of ablation, sparing non-targeted tissue. This experimental proof of concept demonstrates the feasibility of treating volumes comparable to those of primary solid tumors with a miniaturized USgHIFU catheter whose dimensions are close to those of tools traditionally used in interventional radiology, while offering new functionalities.</p>\",\"PeriodicalId\":13322,\"journal\":{\"name\":\"IEEE transactions on ultrasonics, ferroelectrics, and frequency control\",\"volume\":\"PP \",\"pages\":\"\"},\"PeriodicalIF\":3.0000,\"publicationDate\":\"2024-09-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IEEE transactions on ultrasonics, ferroelectrics, and frequency control\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1109/TUFFC.2024.3458067\",\"RegionNum\":2,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"ACOUSTICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE transactions on ultrasonics, ferroelectrics, and frequency control","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1109/TUFFC.2024.3458067","RegionNum":2,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ACOUSTICS","Score":null,"Total":0}
引用次数: 0

摘要

目前的肿瘤间质消融技术面临着超声技术可以应对的挑战。超声束的消融半径和方向性可以提高效率和精确度。在此,我们对超声图像引导下高强度聚焦超声(USgHIFU)保形消融的 9 号磁共振兼容双模超声导管原型进行了实验评估。该原型由 64 个压电复合材料线性阵列元件组成,由开放式可编程双模超声平台驱动。在验证了原型的 US 图像引导能力后,对 HIFU 输出性能(动态聚焦和 HIFU 强度)以及相关的三维 HIFU 引起的组织模型热加热(使用 MR 测温)进行了定量表征。最后,通过实验研究了在体外肝脏中产生强力 HIFU 诱导热消融的能力,并与数值建模进行了比较。对几种 HIFU 动态聚焦的研究克服了设备小型化的挑战:单焦点聚焦最大限度地提高了深层能量沉积,而多焦点策略消除了光栅裂片。原型的线性阵列设计使其能够在原位生成组织和肿瘤模拟的间隙超声图像。产生的多焦点压力场没有光栅裂片,换能器表面强度高达 Isapa = 14 W-cm-2。体外进行了 17 次基本热消融。导管的旋转证明了消融的方向性,使非目标组织不受影响。这一实验性概念证明了使用微型 USgHIFU 导管治疗与原发性实体瘤体积相当的肿瘤的可行性,该导管的尺寸与传统上用于介入放射学的工具接近,同时具有新的功能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Interstitial dual-mode ultrasound with a 3-mm MR-compatible catheter for image-guided HIFU and directional in-vitro tissue ablations.

Current interstitial techniques of tumor ablation face challenges that ultrasound technologies could meet. The ablation radius and directionality of the ultrasound beam could improve the efficiency and precision. Here, a 9-gauge MR-compatible dual-mode ultrasound catheter prototype was experimentally evaluated for Ultrasound Image-guided High Intensity Focused Ultrasound (USgHIFU) conformal ablations. The prototype consisted of 64 piezocomposite linear array elements and was driven by an open research programmable dual-mode ultrasound platform. After verifying the US-image guidance capabilities of the prototype, the HIFU output performances (dynamic focusing and HIFU intensities) were quantitatively characterized, together with the associated 3D HIFU-induced thermal heating in tissue phantoms (using MR thermometry). Finally, the ability to produce robustly HIFU-induced thermal ablations in in-vitro liver was studied experimentally and compared to numerical modeling. Investigations of several HIFU dynamic focusing allowed overcoming the challenges of miniaturizing the device: mono-focal focusing maximized deep energy deposition, while multi-focal strategies eliminated grating lobes. The linear-array design of the prototype made it possible to produce interstitial ultrasound images of tissue and tumor mimics in situ. Multi-focal pressure fields were generated without grating lobes and transducer surface intensities reached up to Isapa = 14 W·cm-2. Seventeen elementary thermal ablations were performed in vitro. Rotation of the catheter proved the directionality of ablation, sparing non-targeted tissue. This experimental proof of concept demonstrates the feasibility of treating volumes comparable to those of primary solid tumors with a miniaturized USgHIFU catheter whose dimensions are close to those of tools traditionally used in interventional radiology, while offering new functionalities.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
7.70
自引率
16.70%
发文量
583
审稿时长
4.5 months
期刊介绍: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control includes the theory, technology, materials, and applications relating to: (1) the generation, transmission, and detection of ultrasonic waves and related phenomena; (2) medical ultrasound, including hyperthermia, bioeffects, tissue characterization and imaging; (3) ferroelectric, piezoelectric, and piezomagnetic materials, including crystals, polycrystalline solids, films, polymers, and composites; (4) frequency control, timing and time distribution, including crystal oscillators and other means of classical frequency control, and atomic, molecular and laser frequency control standards. Areas of interest range from fundamental studies to the design and/or applications of devices and systems.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信