纠正 PO2 测量中的系统误差,改进供氧能力 (α)的测量。

IF 4.6 Q2 MATERIALS SCIENCE, BIOMATERIALS
Alexander W. Timpe, Brad A. Seibel
{"title":"纠正 PO2 测量中的系统误差,改进供氧能力 (α)的测量。","authors":"Alexander W. Timpe,&nbsp;Brad A. Seibel","doi":"10.1016/j.cbpa.2024.111737","DOIUrl":null,"url":null,"abstract":"<div><p>An organism's oxygen supply capacity, measured as a ratio of a metabolic rate to its critical oxygen partial pressure, describes the efficacy of oxygen uptake and transport. This metric is sensitive to errors in oxygen measurement, especially near anoxia where the magnitude of instrument error as a proportion of total signal is magnified. Here, we present a conceptual and mathematical method that uses this sensitivity to identify, quantify, and therefore correct oxygen measurements collected using inaccurately calibrated sensors. When appropriate, adding a small correction value to each oxygen measurement counteracts the effects of this error and provides results that are comparable to data from accurately calibrated oxygen probes. We demonstrate, using simulated, laboratory, and literature datasets, how this method can be used post hoc to diagnose error in, correct the magnitude of, and reduce the variability in repeat measures of traits relevant to oxygen tolerance.</p></div>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-09-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Correcting systematic error in PO2 measurement to improve measures of oxygen supply capacity (α)\",\"authors\":\"Alexander W. Timpe,&nbsp;Brad A. Seibel\",\"doi\":\"10.1016/j.cbpa.2024.111737\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>An organism's oxygen supply capacity, measured as a ratio of a metabolic rate to its critical oxygen partial pressure, describes the efficacy of oxygen uptake and transport. This metric is sensitive to errors in oxygen measurement, especially near anoxia where the magnitude of instrument error as a proportion of total signal is magnified. Here, we present a conceptual and mathematical method that uses this sensitivity to identify, quantify, and therefore correct oxygen measurements collected using inaccurately calibrated sensors. When appropriate, adding a small correction value to each oxygen measurement counteracts the effects of this error and provides results that are comparable to data from accurately calibrated oxygen probes. We demonstrate, using simulated, laboratory, and literature datasets, how this method can be used post hoc to diagnose error in, correct the magnitude of, and reduce the variability in repeat measures of traits relevant to oxygen tolerance.</p></div>\",\"PeriodicalId\":2,\"journal\":{\"name\":\"ACS Applied Bio Materials\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":4.6000,\"publicationDate\":\"2024-09-06\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ACS Applied Bio Materials\",\"FirstCategoryId\":\"99\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1095643324001648\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MATERIALS SCIENCE, BIOMATERIALS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"99","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1095643324001648","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0

摘要

生物体的供氧能力是以新陈代谢率与其临界氧分压的比值来衡量的,它描述了氧气吸收和运输的效率。这一指标对氧气测量误差非常敏感,尤其是在缺氧状态下,仪器误差占总信号的比例会放大。在此,我们介绍一种概念和数学方法,利用这种敏感性来识别、量化并修正使用校准不准确的传感器收集的氧气测量值。在适当的时候,在每次氧气测量中加入一个小的修正值,就能抵消这种误差的影响,并提供与经过精确校准的氧气探头数据相媲美的结果。我们利用模拟、实验室和文献数据集演示了这种方法如何用于事后诊断误差、校正误差的大小以及减少重复测量与耐氧性相关性状的变异性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Correcting systematic error in PO2 measurement to improve measures of oxygen supply capacity (α)

Correcting systematic error in PO2 measurement to improve measures of oxygen supply capacity (α)

An organism's oxygen supply capacity, measured as a ratio of a metabolic rate to its critical oxygen partial pressure, describes the efficacy of oxygen uptake and transport. This metric is sensitive to errors in oxygen measurement, especially near anoxia where the magnitude of instrument error as a proportion of total signal is magnified. Here, we present a conceptual and mathematical method that uses this sensitivity to identify, quantify, and therefore correct oxygen measurements collected using inaccurately calibrated sensors. When appropriate, adding a small correction value to each oxygen measurement counteracts the effects of this error and provides results that are comparable to data from accurately calibrated oxygen probes. We demonstrate, using simulated, laboratory, and literature datasets, how this method can be used post hoc to diagnose error in, correct the magnitude of, and reduce the variability in repeat measures of traits relevant to oxygen tolerance.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
ACS Applied Bio Materials
ACS Applied Bio Materials Chemistry-Chemistry (all)
CiteScore
9.40
自引率
2.10%
发文量
464
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信