{"title":"新型独立式后腰椎面关节稳定装置的生物力学效应:体外尸体研究。","authors":"Temesgen G Assefa, Alexis C Ratliff, Anna G U Sawa, Brian P Kelly","doi":"10.1016/j.wneu.2024.08.168","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Although pedicle screw-rod (PSR) instrumentation remains the gold standard, it has complications, including pedicle breach and facet joint violation. There is current interest in facet joint stabilization with the potential to create a less invasive, natural arch of fixation that may avoid the complications of PSR instrumentation. This study examined the stabilizing potential of a novel facet joint fixation device for single-level (L4-L5) fixation in a human cadaveric model.</p><p><strong>Methods: </strong>Six L3-S1 specimens were tested multidirectionally under pure moment loading (7.5 Nm) in 3 conditions: 1) intact, 2) L4-L5 facet fixation without screws, and 3) L4-L5 facet fixation with screws. L4-L5 intervertebral disc angles were measured radiographically. Range of motion (ROM) and disc angles were compared using repeated-measures analysis of variance analysis, with statistical significance p<0.05.</p><p><strong>Results: </strong>Compared to the intact condition, L4-L5 bilateral facet fixation without or with screw fixation significantly reduced L4-L5 angular ROM in all directions (p≤0.003). No significant differences were observed in cranial and caudal adjacent-segment ROM (p≥0.08) except for L3-L4 fixation in extension, which exhibited small motion increases (0.12° without screws, 0.1° with screws) versus the intact condition (p≤0.003). No statistically significant differences were observed in disc angle values between the conditions (p=0.87).</p><p><strong>Conclusions: </strong>Bilateral lumbar facet fixation with and without supplemental transfacet screw fixation provided significant stability. Cranial and caudal adjacent-level ROM was not influenced by facet fixation except for a slight increase in cranial segment motion during extension. Facet fixation did not alter the lordotic intervertebral disc angle at the instrumented level.</p>","PeriodicalId":23906,"journal":{"name":"World neurosurgery","volume":null,"pages":null},"PeriodicalIF":1.9000,"publicationDate":"2024-09-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Biomechanical effects of a novel standalone posterior lumbar facet joint stabilization device: an in vitro cadaveric study.\",\"authors\":\"Temesgen G Assefa, Alexis C Ratliff, Anna G U Sawa, Brian P Kelly\",\"doi\":\"10.1016/j.wneu.2024.08.168\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>Although pedicle screw-rod (PSR) instrumentation remains the gold standard, it has complications, including pedicle breach and facet joint violation. There is current interest in facet joint stabilization with the potential to create a less invasive, natural arch of fixation that may avoid the complications of PSR instrumentation. This study examined the stabilizing potential of a novel facet joint fixation device for single-level (L4-L5) fixation in a human cadaveric model.</p><p><strong>Methods: </strong>Six L3-S1 specimens were tested multidirectionally under pure moment loading (7.5 Nm) in 3 conditions: 1) intact, 2) L4-L5 facet fixation without screws, and 3) L4-L5 facet fixation with screws. L4-L5 intervertebral disc angles were measured radiographically. Range of motion (ROM) and disc angles were compared using repeated-measures analysis of variance analysis, with statistical significance p<0.05.</p><p><strong>Results: </strong>Compared to the intact condition, L4-L5 bilateral facet fixation without or with screw fixation significantly reduced L4-L5 angular ROM in all directions (p≤0.003). No significant differences were observed in cranial and caudal adjacent-segment ROM (p≥0.08) except for L3-L4 fixation in extension, which exhibited small motion increases (0.12° without screws, 0.1° with screws) versus the intact condition (p≤0.003). No statistically significant differences were observed in disc angle values between the conditions (p=0.87).</p><p><strong>Conclusions: </strong>Bilateral lumbar facet fixation with and without supplemental transfacet screw fixation provided significant stability. Cranial and caudal adjacent-level ROM was not influenced by facet fixation except for a slight increase in cranial segment motion during extension. Facet fixation did not alter the lordotic intervertebral disc angle at the instrumented level.</p>\",\"PeriodicalId\":23906,\"journal\":{\"name\":\"World neurosurgery\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.9000,\"publicationDate\":\"2024-09-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"World neurosurgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1016/j.wneu.2024.08.168\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CLINICAL NEUROLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"World neurosurgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.wneu.2024.08.168","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:尽管椎弓根螺钉-螺杆(PSR)器械仍是黄金标准,但它也有并发症,包括椎弓根破损和面骨关节侵犯。目前,人们开始关注面关节的稳定问题,因为它有可能创造出一种创伤较小、自然拱形的固定方式,从而避免椎弓根螺钉器械的并发症。本研究在人体尸体模型中检验了一种新型面关节固定装置的稳定潜力,该装置可用于单水平(L4-L5)固定:方法:在 3 种条件下,对 6 个 L3-S1 标本进行了纯力矩加载(7.5 牛米)下的多方向测试:1)完好无损;2)L4-L5 椎面固定(无螺钉);3)L4-L5 椎面固定(有螺钉)。对 L4-L5 椎间盘角度进行放射学测量。采用重复测量方差分析比较了活动范围(ROM)和椎间盘角度,统计显著性为 p:与完好的情况相比,L4-L5 双侧面固定(无螺钉固定或有螺钉固定)显著降低了 L4-L5 在所有方向上的角度 ROM(p≤0.003)。除L3-L4伸展时的固定外,头侧和尾侧邻近节段的ROM没有观察到明显差异(p≥0.08),而L3-L4伸展时的固定与完好无损的情况相比,运动幅度略有增加(无螺钉时为0.12°,有螺钉时为0.1°)(p≤0.003)。两种情况下的椎间盘角度值差异无统计学意义(P=0.87):结论:使用或不使用经椎体外侧螺钉固定的双侧腰椎面固定具有明显的稳定性。除了伸展时头颅节段的运动略有增加外,头颅和尾部邻近水平的 ROM 均未受椎板固定的影响。面骨固定不会改变器械水平的椎间盘前凸角。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Biomechanical effects of a novel standalone posterior lumbar facet joint stabilization device: an in vitro cadaveric study.

Objective: Although pedicle screw-rod (PSR) instrumentation remains the gold standard, it has complications, including pedicle breach and facet joint violation. There is current interest in facet joint stabilization with the potential to create a less invasive, natural arch of fixation that may avoid the complications of PSR instrumentation. This study examined the stabilizing potential of a novel facet joint fixation device for single-level (L4-L5) fixation in a human cadaveric model.

Methods: Six L3-S1 specimens were tested multidirectionally under pure moment loading (7.5 Nm) in 3 conditions: 1) intact, 2) L4-L5 facet fixation without screws, and 3) L4-L5 facet fixation with screws. L4-L5 intervertebral disc angles were measured radiographically. Range of motion (ROM) and disc angles were compared using repeated-measures analysis of variance analysis, with statistical significance p<0.05.

Results: Compared to the intact condition, L4-L5 bilateral facet fixation without or with screw fixation significantly reduced L4-L5 angular ROM in all directions (p≤0.003). No significant differences were observed in cranial and caudal adjacent-segment ROM (p≥0.08) except for L3-L4 fixation in extension, which exhibited small motion increases (0.12° without screws, 0.1° with screws) versus the intact condition (p≤0.003). No statistically significant differences were observed in disc angle values between the conditions (p=0.87).

Conclusions: Bilateral lumbar facet fixation with and without supplemental transfacet screw fixation provided significant stability. Cranial and caudal adjacent-level ROM was not influenced by facet fixation except for a slight increase in cranial segment motion during extension. Facet fixation did not alter the lordotic intervertebral disc angle at the instrumented level.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
World neurosurgery
World neurosurgery CLINICAL NEUROLOGY-SURGERY
CiteScore
3.90
自引率
15.00%
发文量
1765
审稿时长
47 days
期刊介绍: World Neurosurgery has an open access mirror journal World Neurosurgery: X, sharing the same aims and scope, editorial team, submission system and rigorous peer review. The journal''s mission is to: -To provide a first-class international forum and a 2-way conduit for dialogue that is relevant to neurosurgeons and providers who care for neurosurgery patients. The categories of the exchanged information include clinical and basic science, as well as global information that provide social, political, educational, economic, cultural or societal insights and knowledge that are of significance and relevance to worldwide neurosurgery patient care. -To act as a primary intellectual catalyst for the stimulation of creativity, the creation of new knowledge, and the enhancement of quality neurosurgical care worldwide. -To provide a forum for communication that enriches the lives of all neurosurgeons and their colleagues; and, in so doing, enriches the lives of their patients. Topics to be addressed in World Neurosurgery include: EDUCATION, ECONOMICS, RESEARCH, POLITICS, HISTORY, CULTURE, CLINICAL SCIENCE, LABORATORY SCIENCE, TECHNOLOGY, OPERATIVE TECHNIQUES, CLINICAL IMAGES, VIDEOS
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信