{"title":"评估基于重氮的护理点胆红素检测法 careSTART S1 总胆红素试纸。","authors":"Woojae Kwoun, Jeong Joong Lee, Young-Ah Youn, Hyojin Chae, Jehoon Lee","doi":"10.1002/jcla.25093","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Neonatal jaundice (NNJ) affects a significant proportion of newborns globally, with an increased burden in low-resource settings. Effective health risk management of NNJ is hindered, particularly in resource-constrained environments, where early detection and treatment are challenging. The careSTART S1 Total Bilirubin Strip, a point-of-care testing (POCT) device based on a diazo-method, offers a potential solution by enabling onsite bilirubin measurement, thus, addressing the gap in early NNJ detection and management.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The current study evaluated the analytical performance of the careSTART S1 Total Bilirubin Strip for precision, linearity, method comparison, and lot-to-lot consistency following CLSI guidelines. For method comparison, 105 residual EDTA whole blood samples were analyzed with the careSTART S1 Total Bilirubin Strip and compared with reference measurements from the Roche Cobas c702 analyzer. Additionally, statistical analyses, including Passing-Bablok regression and Bland–Altman plots, were performed.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The careSTART S1 Total Bilirubin Strip showed allowable (<10%) within-laboratory imprecision of 2.5%–3.6% across all levels and demonstrated linearity over the range of 4.16–439.3 μmol/L. Method comparison revealed a constant negative bias with a mean bias −4.19 μmol/L. However, the 95% confidence interval (−7.10 to −1.28 μmol/L) of the bias is covered by the prespecified allowable bias of 8.3%, at medical decision point. Lot-to-lot variation ranged from 0.14%–6.49%, and was within the acceptable critical difference of 8.3%.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The careSTART S1 Total Bilirubin Strip provided accurate and reliable bilirubin measurements that could contribute to neonatal care in settings lacking central laboratory facilities.</p>\n </section>\n </div>","PeriodicalId":15509,"journal":{"name":"Journal of Clinical Laboratory Analysis","volume":"38 15-16","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2024-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/jcla.25093","citationCount":"0","resultStr":"{\"title\":\"Evaluation of a Diazo-Based Point-Of-Care Bilirubin Assay careSTART S1 Total Bilirubin Strip\",\"authors\":\"Woojae Kwoun, Jeong Joong Lee, Young-Ah Youn, Hyojin Chae, Jehoon Lee\",\"doi\":\"10.1002/jcla.25093\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>Neonatal jaundice (NNJ) affects a significant proportion of newborns globally, with an increased burden in low-resource settings. Effective health risk management of NNJ is hindered, particularly in resource-constrained environments, where early detection and treatment are challenging. The careSTART S1 Total Bilirubin Strip, a point-of-care testing (POCT) device based on a diazo-method, offers a potential solution by enabling onsite bilirubin measurement, thus, addressing the gap in early NNJ detection and management.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>The current study evaluated the analytical performance of the careSTART S1 Total Bilirubin Strip for precision, linearity, method comparison, and lot-to-lot consistency following CLSI guidelines. For method comparison, 105 residual EDTA whole blood samples were analyzed with the careSTART S1 Total Bilirubin Strip and compared with reference measurements from the Roche Cobas c702 analyzer. Additionally, statistical analyses, including Passing-Bablok regression and Bland–Altman plots, were performed.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>The careSTART S1 Total Bilirubin Strip showed allowable (<10%) within-laboratory imprecision of 2.5%–3.6% across all levels and demonstrated linearity over the range of 4.16–439.3 μmol/L. Method comparison revealed a constant negative bias with a mean bias −4.19 μmol/L. However, the 95% confidence interval (−7.10 to −1.28 μmol/L) of the bias is covered by the prespecified allowable bias of 8.3%, at medical decision point. Lot-to-lot variation ranged from 0.14%–6.49%, and was within the acceptable critical difference of 8.3%.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusion</h3>\\n \\n <p>The careSTART S1 Total Bilirubin Strip provided accurate and reliable bilirubin measurements that could contribute to neonatal care in settings lacking central laboratory facilities.</p>\\n </section>\\n </div>\",\"PeriodicalId\":15509,\"journal\":{\"name\":\"Journal of Clinical Laboratory Analysis\",\"volume\":\"38 15-16\",\"pages\":\"\"},\"PeriodicalIF\":2.6000,\"publicationDate\":\"2024-08-21\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1002/jcla.25093\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Clinical Laboratory Analysis\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/jcla.25093\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Laboratory Analysis","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/jcla.25093","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
Evaluation of a Diazo-Based Point-Of-Care Bilirubin Assay careSTART S1 Total Bilirubin Strip
Background
Neonatal jaundice (NNJ) affects a significant proportion of newborns globally, with an increased burden in low-resource settings. Effective health risk management of NNJ is hindered, particularly in resource-constrained environments, where early detection and treatment are challenging. The careSTART S1 Total Bilirubin Strip, a point-of-care testing (POCT) device based on a diazo-method, offers a potential solution by enabling onsite bilirubin measurement, thus, addressing the gap in early NNJ detection and management.
Methods
The current study evaluated the analytical performance of the careSTART S1 Total Bilirubin Strip for precision, linearity, method comparison, and lot-to-lot consistency following CLSI guidelines. For method comparison, 105 residual EDTA whole blood samples were analyzed with the careSTART S1 Total Bilirubin Strip and compared with reference measurements from the Roche Cobas c702 analyzer. Additionally, statistical analyses, including Passing-Bablok regression and Bland–Altman plots, were performed.
Results
The careSTART S1 Total Bilirubin Strip showed allowable (<10%) within-laboratory imprecision of 2.5%–3.6% across all levels and demonstrated linearity over the range of 4.16–439.3 μmol/L. Method comparison revealed a constant negative bias with a mean bias −4.19 μmol/L. However, the 95% confidence interval (−7.10 to −1.28 μmol/L) of the bias is covered by the prespecified allowable bias of 8.3%, at medical decision point. Lot-to-lot variation ranged from 0.14%–6.49%, and was within the acceptable critical difference of 8.3%.
Conclusion
The careSTART S1 Total Bilirubin Strip provided accurate and reliable bilirubin measurements that could contribute to neonatal care in settings lacking central laboratory facilities.
期刊介绍:
Journal of Clinical Laboratory Analysis publishes original articles on newly developing modes of technology and laboratory assays, with emphasis on their application in current and future clinical laboratory testing. This includes reports from the following fields: immunochemistry and toxicology, hematology and hematopathology, immunopathology, molecular diagnostics, microbiology, genetic testing, immunohematology, and clinical chemistry.