{"title":"评估基于重氮的护理点胆红素检测法 careSTART S1 总胆红素试纸。","authors":"Woojae Kwoun,&nbsp;Jeong Joong Lee,&nbsp;Young-Ah Youn,&nbsp;Hyojin Chae,&nbsp;Jehoon Lee","doi":"10.1002/jcla.25093","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Neonatal jaundice (NNJ) affects a significant proportion of newborns globally, with an increased burden in low-resource settings. Effective health risk management of NNJ is hindered, particularly in resource-constrained environments, where early detection and treatment are challenging. The careSTART S1 Total Bilirubin Strip, a point-of-care testing (POCT) device based on a diazo-method, offers a potential solution by enabling onsite bilirubin measurement, thus, addressing the gap in early NNJ detection and management.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The current study evaluated the analytical performance of the careSTART S1 Total Bilirubin Strip for precision, linearity, method comparison, and lot-to-lot consistency following CLSI guidelines. For method comparison, 105 residual EDTA whole blood samples were analyzed with the careSTART S1 Total Bilirubin Strip and compared with reference measurements from the Roche Cobas c702 analyzer. Additionally, statistical analyses, including Passing-Bablok regression and Bland–Altman plots, were performed.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The careSTART S1 Total Bilirubin Strip showed allowable (&lt;10%) within-laboratory imprecision of 2.5%–3.6% across all levels and demonstrated linearity over the range of 4.16–439.3 μmol/L. Method comparison revealed a constant negative bias with a mean bias −4.19 μmol/L. However, the 95% confidence interval (−7.10 to −1.28 μmol/L) of the bias is covered by the prespecified allowable bias of 8.3%, at medical decision point. Lot-to-lot variation ranged from 0.14%–6.49%, and was within the acceptable critical difference of 8.3%.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The careSTART S1 Total Bilirubin Strip provided accurate and reliable bilirubin measurements that could contribute to neonatal care in settings lacking central laboratory facilities.</p>\n </section>\n </div>","PeriodicalId":15509,"journal":{"name":"Journal of Clinical Laboratory Analysis","volume":"38 15-16","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2024-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/jcla.25093","citationCount":"0","resultStr":"{\"title\":\"Evaluation of a Diazo-Based Point-Of-Care Bilirubin Assay careSTART S1 Total Bilirubin Strip\",\"authors\":\"Woojae Kwoun,&nbsp;Jeong Joong Lee,&nbsp;Young-Ah Youn,&nbsp;Hyojin Chae,&nbsp;Jehoon Lee\",\"doi\":\"10.1002/jcla.25093\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>Neonatal jaundice (NNJ) affects a significant proportion of newborns globally, with an increased burden in low-resource settings. Effective health risk management of NNJ is hindered, particularly in resource-constrained environments, where early detection and treatment are challenging. The careSTART S1 Total Bilirubin Strip, a point-of-care testing (POCT) device based on a diazo-method, offers a potential solution by enabling onsite bilirubin measurement, thus, addressing the gap in early NNJ detection and management.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>The current study evaluated the analytical performance of the careSTART S1 Total Bilirubin Strip for precision, linearity, method comparison, and lot-to-lot consistency following CLSI guidelines. For method comparison, 105 residual EDTA whole blood samples were analyzed with the careSTART S1 Total Bilirubin Strip and compared with reference measurements from the Roche Cobas c702 analyzer. Additionally, statistical analyses, including Passing-Bablok regression and Bland–Altman plots, were performed.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>The careSTART S1 Total Bilirubin Strip showed allowable (&lt;10%) within-laboratory imprecision of 2.5%–3.6% across all levels and demonstrated linearity over the range of 4.16–439.3 μmol/L. Method comparison revealed a constant negative bias with a mean bias −4.19 μmol/L. However, the 95% confidence interval (−7.10 to −1.28 μmol/L) of the bias is covered by the prespecified allowable bias of 8.3%, at medical decision point. Lot-to-lot variation ranged from 0.14%–6.49%, and was within the acceptable critical difference of 8.3%.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusion</h3>\\n \\n <p>The careSTART S1 Total Bilirubin Strip provided accurate and reliable bilirubin measurements that could contribute to neonatal care in settings lacking central laboratory facilities.</p>\\n </section>\\n </div>\",\"PeriodicalId\":15509,\"journal\":{\"name\":\"Journal of Clinical Laboratory Analysis\",\"volume\":\"38 15-16\",\"pages\":\"\"},\"PeriodicalIF\":2.6000,\"publicationDate\":\"2024-08-21\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1002/jcla.25093\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Clinical Laboratory Analysis\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/jcla.25093\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Laboratory Analysis","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/jcla.25093","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景:新生儿黄疸(NNJ)影响着全球很大一部分新生儿,在资源匮乏的环境中,新生儿黄疸的负担会加重。新生儿黄疸的有效健康风险管理受到阻碍,尤其是在资源有限的环境中,早期检测和治疗具有挑战性。CareSTART S1 总胆红素试纸条是一种基于重氮法的护理点检测(POCT)设备,它提供了一种潜在的解决方案,可实现现场胆红素测量,从而弥补早期 NNJ 检测和管理方面的不足:本研究按照 CLSI 指南评估了 CareSTART S1 总胆红素试纸条的分析性能,包括精确度、线性度、方法比较和批次间一致性。为了进行方法比较,使用 careSTART S1 总胆红素试纸条分析了 105 份残留 EDTA 全血样本,并与罗氏 Cobas c702 分析仪的参考测量值进行了比较。此外,还进行了统计分析,包括 Passing-Bablok 回归和 Bland-Altman 图:结果:careSTART S1 总胆红素试纸条显示出了可接受的范围:CareSTART S1 总胆红素试纸可提供准确可靠的胆红素测量结果,有助于在缺乏中心实验室设施的环境中开展新生儿护理工作。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Evaluation of a Diazo-Based Point-Of-Care Bilirubin Assay careSTART S1 Total Bilirubin Strip

Evaluation of a Diazo-Based Point-Of-Care Bilirubin Assay careSTART S1 Total Bilirubin Strip

Background

Neonatal jaundice (NNJ) affects a significant proportion of newborns globally, with an increased burden in low-resource settings. Effective health risk management of NNJ is hindered, particularly in resource-constrained environments, where early detection and treatment are challenging. The careSTART S1 Total Bilirubin Strip, a point-of-care testing (POCT) device based on a diazo-method, offers a potential solution by enabling onsite bilirubin measurement, thus, addressing the gap in early NNJ detection and management.

Methods

The current study evaluated the analytical performance of the careSTART S1 Total Bilirubin Strip for precision, linearity, method comparison, and lot-to-lot consistency following CLSI guidelines. For method comparison, 105 residual EDTA whole blood samples were analyzed with the careSTART S1 Total Bilirubin Strip and compared with reference measurements from the Roche Cobas c702 analyzer. Additionally, statistical analyses, including Passing-Bablok regression and Bland–Altman plots, were performed.

Results

The careSTART S1 Total Bilirubin Strip showed allowable (<10%) within-laboratory imprecision of 2.5%–3.6% across all levels and demonstrated linearity over the range of 4.16–439.3 μmol/L. Method comparison revealed a constant negative bias with a mean bias −4.19 μmol/L. However, the 95% confidence interval (−7.10 to −1.28 μmol/L) of the bias is covered by the prespecified allowable bias of 8.3%, at medical decision point. Lot-to-lot variation ranged from 0.14%–6.49%, and was within the acceptable critical difference of 8.3%.

Conclusion

The careSTART S1 Total Bilirubin Strip provided accurate and reliable bilirubin measurements that could contribute to neonatal care in settings lacking central laboratory facilities.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Clinical Laboratory Analysis
Journal of Clinical Laboratory Analysis 医学-医学实验技术
CiteScore
5.60
自引率
7.40%
发文量
584
审稿时长
6-12 weeks
期刊介绍: Journal of Clinical Laboratory Analysis publishes original articles on newly developing modes of technology and laboratory assays, with emphasis on their application in current and future clinical laboratory testing. This includes reports from the following fields: immunochemistry and toxicology, hematology and hematopathology, immunopathology, molecular diagnostics, microbiology, genetic testing, immunohematology, and clinical chemistry.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信