四种手持式血液乳酸分析仪在低强度到高强度运动中的精确度和准确性。

IF 4.6 Q2 MATERIALS SCIENCE, BIOMATERIALS
ACS Applied Bio Materials Pub Date : 2024-12-01 Epub Date: 2024-08-19 DOI:10.1007/s00421-024-05572-6
Fredrik Mentzoni, Martin Skaugen, Ingrid Eythorsdottir, Stian Roterud, Espen Spro Johansen, Thomas Losnegard
{"title":"四种手持式血液乳酸分析仪在低强度到高强度运动中的精确度和准确性。","authors":"Fredrik Mentzoni, Martin Skaugen, Ingrid Eythorsdottir, Stian Roterud, Espen Spro Johansen, Thomas Losnegard","doi":"10.1007/s00421-024-05572-6","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the precision and accuracy in measured blood lactate concentrations among four commonly used handheld lactate analyzers compared to two stationary analyzers.</p><p><strong>Methods: </strong>Venous blood samples were taken at exercise intensities ranging from low to high. The blood lactate concentration was measured simultaneously with four pairs of handheld lactate analyzers (two new units of each brand: Lactate Plus, Lactate Pro2, Lactate Scout 4, and TaiDoc TD-4289), and compared with two stationary analyzers (Biosen C-Line and YSI Sport 1500). Measurements were repeated for a range of blood lactate concentrations (measured with Biosen) from 0.88 to 4.89 mM with a median difference between measurements of 0.10 mM.</p><p><strong>Results: </strong>The mean relative differences to the Biosen analyzer were <math><mo>-</mo></math> 7% (Plus), 7% (Pro), <math><mo>-</mo></math> 10% (Scout), 42% (Tai), and <math><mo>-</mo></math> 32% (Ysi). The residual standard errors after linear regression against Biosen were 0.18 mM (Plus), 0.20 mM (Pro), 0.22 mM (Scout), 0.15 mM (Tai), and 0.06 mM (Ysi). Accordingly, a blood lactate concentration of 3 mM measured with Biosen yielded 95% prediction intervals that were 0.72 mM (Plus), 0.80 mM (Pro), 0.87 mM (Scout), 0.60 mM (Tai), and 0.23 mM (Ysi) wide.</p><p><strong>Conclusion: </strong>Compared to our two stationary analyzers, the precision of the four handheld lactate analyzers evaluated in this study was poor. Among the four, Tai was the most precise; however, this analyzer had low accuracy with a substantial mean difference to the reference analyzer.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Precision and accuracy of four handheld blood lactate analyzers across low to high exercise intensities.\",\"authors\":\"Fredrik Mentzoni, Martin Skaugen, Ingrid Eythorsdottir, Stian Roterud, Espen Spro Johansen, Thomas Losnegard\",\"doi\":\"10.1007/s00421-024-05572-6\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To evaluate the precision and accuracy in measured blood lactate concentrations among four commonly used handheld lactate analyzers compared to two stationary analyzers.</p><p><strong>Methods: </strong>Venous blood samples were taken at exercise intensities ranging from low to high. The blood lactate concentration was measured simultaneously with four pairs of handheld lactate analyzers (two new units of each brand: Lactate Plus, Lactate Pro2, Lactate Scout 4, and TaiDoc TD-4289), and compared with two stationary analyzers (Biosen C-Line and YSI Sport 1500). Measurements were repeated for a range of blood lactate concentrations (measured with Biosen) from 0.88 to 4.89 mM with a median difference between measurements of 0.10 mM.</p><p><strong>Results: </strong>The mean relative differences to the Biosen analyzer were <math><mo>-</mo></math> 7% (Plus), 7% (Pro), <math><mo>-</mo></math> 10% (Scout), 42% (Tai), and <math><mo>-</mo></math> 32% (Ysi). The residual standard errors after linear regression against Biosen were 0.18 mM (Plus), 0.20 mM (Pro), 0.22 mM (Scout), 0.15 mM (Tai), and 0.06 mM (Ysi). Accordingly, a blood lactate concentration of 3 mM measured with Biosen yielded 95% prediction intervals that were 0.72 mM (Plus), 0.80 mM (Pro), 0.87 mM (Scout), 0.60 mM (Tai), and 0.23 mM (Ysi) wide.</p><p><strong>Conclusion: </strong>Compared to our two stationary analyzers, the precision of the four handheld lactate analyzers evaluated in this study was poor. Among the four, Tai was the most precise; however, this analyzer had low accuracy with a substantial mean difference to the reference analyzer.</p>\",\"PeriodicalId\":2,\"journal\":{\"name\":\"ACS Applied Bio Materials\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":4.6000,\"publicationDate\":\"2024-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ACS Applied Bio Materials\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s00421-024-05572-6\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/8/19 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"MATERIALS SCIENCE, BIOMATERIALS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s00421-024-05572-6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/19 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0

摘要

目的:评估四种常用手持式乳酸分析仪与两种固定式乳酸分析仪测量血乳酸浓度的精确度和准确性:方法:在运动强度从低到高的情况下采集静脉血样本。方法:在低强度、高强度的运动中采集静脉血样本,用四对手持式乳酸分析仪(每种品牌各两台新仪器,分别为 Lactate Plus、Lactate Plus 和 Lactate Plus)同时测量血液中的乳酸浓度:Lactate Plus、Lactate Pro2、Lactate Scout 4 和 TaiDoc TD-4289),并与两台固定式分析仪(Biosen C-Line 和 YSI Sport 1500)进行比较。在 0.88 至 4.89 毫摩尔的血乳酸浓度范围内重复测量(用 Biosen 测量),测量结果之间的中位差为 0.10 毫摩尔:与 Biosen 分析仪的平均相对差异为 - 7%(Plus)、7%(Pro)、- 10%(Scout)、42%(Tai)和 - 32%(Ysi)。与 Biosen 进行线性回归后的残差标准误差分别为 0.18 毫摩尔(Plus)、0.20 毫摩尔(Pro)、0.22 毫摩尔(Scout)、0.15 毫摩尔(Tai)和 0.06 毫摩尔(Ysi)。相应地,使用 Biosen 测量 3 mM 的血乳酸浓度时,95% 的预测区间分别为 0.72 mM (Plus)、0.80 mM (Pro)、0.87 mM (Scout)、0.60 mM (Tai) 和 0.23 mM (Ysi):结论:与我们的两台固定式分析仪相比,本研究中评估的四台手持式乳酸分析仪的精确度较差。结论:与我们的两台固定式分析仪相比,本研究评估的四台手持式乳酸分析仪的精确度较差。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Precision and accuracy of four handheld blood lactate analyzers across low to high exercise intensities.

Precision and accuracy of four handheld blood lactate analyzers across low to high exercise intensities.

Purpose: To evaluate the precision and accuracy in measured blood lactate concentrations among four commonly used handheld lactate analyzers compared to two stationary analyzers.

Methods: Venous blood samples were taken at exercise intensities ranging from low to high. The blood lactate concentration was measured simultaneously with four pairs of handheld lactate analyzers (two new units of each brand: Lactate Plus, Lactate Pro2, Lactate Scout 4, and TaiDoc TD-4289), and compared with two stationary analyzers (Biosen C-Line and YSI Sport 1500). Measurements were repeated for a range of blood lactate concentrations (measured with Biosen) from 0.88 to 4.89 mM with a median difference between measurements of 0.10 mM.

Results: The mean relative differences to the Biosen analyzer were - 7% (Plus), 7% (Pro), - 10% (Scout), 42% (Tai), and - 32% (Ysi). The residual standard errors after linear regression against Biosen were 0.18 mM (Plus), 0.20 mM (Pro), 0.22 mM (Scout), 0.15 mM (Tai), and 0.06 mM (Ysi). Accordingly, a blood lactate concentration of 3 mM measured with Biosen yielded 95% prediction intervals that were 0.72 mM (Plus), 0.80 mM (Pro), 0.87 mM (Scout), 0.60 mM (Tai), and 0.23 mM (Ysi) wide.

Conclusion: Compared to our two stationary analyzers, the precision of the four handheld lactate analyzers evaluated in this study was poor. Among the four, Tai was the most precise; however, this analyzer had low accuracy with a substantial mean difference to the reference analyzer.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
ACS Applied Bio Materials
ACS Applied Bio Materials Chemistry-Chemistry (all)
CiteScore
9.40
自引率
2.10%
发文量
464
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信