地震系统中氡检测器的新校准方法

IF 1.2 4区 地球科学 Q3 Earth and Planetary Sciences
{"title":"地震系统中氡检测器的新校准方法","authors":"","doi":"10.1016/j.eqs.2024.06.001","DOIUrl":null,"url":null,"abstract":"<div><p>Radon observation is an important measurement item of seismic precursor network observation. The radon detector calibration is a key technical link for ensuring radon observation accuracy. At present, the radon detector calibration in seismic systems in China is faced with a series of bottleneck problems, such as aging and scrap, acquisition difficulties, high supervision costs, and transportation limitations of radon sources. As a result, a large number of radon detectors cannot be accurately calibrated regularly, seriously affecting the accuracy and reliability of radon observation data in China. To solve this problem, a new calibration method for radon detectors was established. The advantage of this method is that the dangerous radioactive substance, i.e., the radon source, can be avoided, but only “standard instruments” and water samples with certain dissolved radon concentrations can be used to realize radon detector calibration. This method avoids the risk of radioactive leakage and solves the current widespread difficulties and bottleneck of radon detector calibration in seismic systems in China. The comparison experiment with the traditional calibration method shows that the error of the calibration coefficient obtained by the new method is less than 5% compared with that by the traditional method, which meets the requirements of seismic observation systems, confirming the reliability of the new method. This new method can completely replace the traditional calibration method of using a radon source in seismic systems.</p></div>","PeriodicalId":46333,"journal":{"name":"Earthquake Science","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-08-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1674451924000612/pdfft?md5=57db9f4e2947e9ef8657cdc44b9359c2&pid=1-s2.0-S1674451924000612-main.pdf","citationCount":"0","resultStr":"{\"title\":\"A new calibration method for radon detector in seismic systems\",\"authors\":\"\",\"doi\":\"10.1016/j.eqs.2024.06.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>Radon observation is an important measurement item of seismic precursor network observation. The radon detector calibration is a key technical link for ensuring radon observation accuracy. At present, the radon detector calibration in seismic systems in China is faced with a series of bottleneck problems, such as aging and scrap, acquisition difficulties, high supervision costs, and transportation limitations of radon sources. As a result, a large number of radon detectors cannot be accurately calibrated regularly, seriously affecting the accuracy and reliability of radon observation data in China. To solve this problem, a new calibration method for radon detectors was established. The advantage of this method is that the dangerous radioactive substance, i.e., the radon source, can be avoided, but only “standard instruments” and water samples with certain dissolved radon concentrations can be used to realize radon detector calibration. This method avoids the risk of radioactive leakage and solves the current widespread difficulties and bottleneck of radon detector calibration in seismic systems in China. The comparison experiment with the traditional calibration method shows that the error of the calibration coefficient obtained by the new method is less than 5% compared with that by the traditional method, which meets the requirements of seismic observation systems, confirming the reliability of the new method. This new method can completely replace the traditional calibration method of using a radon source in seismic systems.</p></div>\",\"PeriodicalId\":46333,\"journal\":{\"name\":\"Earthquake Science\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2024-08-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S1674451924000612/pdfft?md5=57db9f4e2947e9ef8657cdc44b9359c2&pid=1-s2.0-S1674451924000612-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Earthquake Science\",\"FirstCategoryId\":\"89\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1674451924000612\",\"RegionNum\":4,\"RegionCategory\":\"地球科学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Earth and Planetary Sciences\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Earthquake Science","FirstCategoryId":"89","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1674451924000612","RegionNum":4,"RegionCategory":"地球科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Earth and Planetary Sciences","Score":null,"Total":0}
引用次数: 0

摘要

氡观测是地震前兆台网观测的重要测量项目。氡探测器校准是保证氡观测精度的关键技术环节。目前,我国地震系统氡检测仪校准面临着老化报废、采集困难、监管成本高、氡源运输受限等一系列瓶颈问题。因此,大量氡探测仪无法定期进行准确校准,严重影响了我国氡观测数据的准确性和可靠性。为了解决这一问题,一种新的氡检测仪校准方法应运而生。该方法的优点是避免了氡源这一危险的放射性物质,只能使用 "标准仪器 "和具有一定溶解氡浓度的水样来实现氡检测仪的校准。该方法避免了放射性物质泄漏的风险,解决了目前我国地震系统中普遍存在的氡探测器校准困难和瓶颈问题。与传统校准方法的对比实验表明,新方法得到的校准系数误差小于传统方法的5%,满足地震观测系统的要求,证实了新方法的可靠性。这种新方法可以完全取代地震系统中使用氡源的传统校准方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A new calibration method for radon detector in seismic systems

Radon observation is an important measurement item of seismic precursor network observation. The radon detector calibration is a key technical link for ensuring radon observation accuracy. At present, the radon detector calibration in seismic systems in China is faced with a series of bottleneck problems, such as aging and scrap, acquisition difficulties, high supervision costs, and transportation limitations of radon sources. As a result, a large number of radon detectors cannot be accurately calibrated regularly, seriously affecting the accuracy and reliability of radon observation data in China. To solve this problem, a new calibration method for radon detectors was established. The advantage of this method is that the dangerous radioactive substance, i.e., the radon source, can be avoided, but only “standard instruments” and water samples with certain dissolved radon concentrations can be used to realize radon detector calibration. This method avoids the risk of radioactive leakage and solves the current widespread difficulties and bottleneck of radon detector calibration in seismic systems in China. The comparison experiment with the traditional calibration method shows that the error of the calibration coefficient obtained by the new method is less than 5% compared with that by the traditional method, which meets the requirements of seismic observation systems, confirming the reliability of the new method. This new method can completely replace the traditional calibration method of using a radon source in seismic systems.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Earthquake Science
Earthquake Science GEOCHEMISTRY & GEOPHYSICS-
CiteScore
1.10
自引率
8.30%
发文量
42
审稿时长
3 months
期刊介绍: Earthquake Science (EQS) aims to publish high-quality, original, peer-reviewed articles on earthquake-related research subjects. It is an English international journal sponsored by the Seismological Society of China and the Institute of Geophysics, China Earthquake Administration. The topics include, but not limited to, the following ● Seismic sources of all kinds. ● Earth structure at all scales. ● Seismotectonics. ● New methods and theoretical seismology. ● Strong ground motion. ● Seismic phenomena of all kinds. ● Seismic hazards, earthquake forecasting and prediction. ● Seismic instrumentation. ● Significant recent or past seismic events. ● Documentation of recent seismic events or important observations. ● Descriptions of field deployments, new methods, and available software tools. The types of manuscripts include the following. There is no length requirement, except for the Short Notes. 【Articles】 Original contributions that have not been published elsewhere. 【Short Notes】 Short papers of recent events or topics that warrant rapid peer reviews and publications. Limited to 4 publication pages. 【Rapid Communications】 Significant contributions that warrant rapid peer reviews and publications. 【Review Articles】Review articles are by invitation only. Please contact the editorial office and editors for possible proposals. 【Toolboxes】 Descriptions of novel numerical methods and associated computer codes. 【Data Products】 Documentation of datasets of various kinds that are interested to the community and available for open access (field data, processed data, synthetic data, or models). 【Opinions】Views on important topics and future directions in earthquake science. 【Comments and Replies】Commentaries on a recently published EQS paper is welcome. The authors of the paper commented will be invited to reply. Both the Comment and the Reply are subject to peer review.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信