评估旋转器械润滑剂对牙科种植体骨结合的影响:肋骨胫骨实验研究。

Heon-Young Kim, Ju-Dong Song, Il-Seok Jang, Sun-Jong Kim, Jin-Woo Kim
{"title":"评估旋转器械润滑剂对牙科种植体骨结合的影响:肋骨胫骨实验研究。","authors":"Heon-Young Kim, Ju-Dong Song, Il-Seok Jang, Sun-Jong Kim, Jin-Woo Kim","doi":"10.11607/jomi.11036","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To investigate the impact of mineral oil lubricants used in rotary instruments on osseointegration within rabbit tibias with a specific focus on potential contamination from dental handpieces.</p><p><strong>Materials and methods: </strong>A total of 12 New Zealand rabbits were included in this study, each receiving 2 implants in each tibia, resulting in a total of 48 implants across the study. Groups were organized based on the time until euthanasia and the degree of implant contamination. Three contamination levels were defined: The first group received implants without any lubricant in the handpiece (control group); the second group received implants with handpieces managed as recommended; and the third group received presoaked implants that were placed via a handpiece treated as per the second group's protocol. These groups were further subdivided based on euthanization periods of 2 and 4 weeks. Both the removal torque and the bone-to-implant contact (BIC) were measured and analyzed.</p><p><strong>Results: </strong>No significant correlation was observed between the degree of implant contamination and removal torque. However, there was a significant reduction in BIC associated with higher contamination levels, particularly after 4 weeks.</p><p><strong>Conclusions: </strong>Even brief exposure to lubricants from handpieces can jeopardize the osseointegration of implants in bone. Therefore, it is imperative to implement thorough procedures for lubricant removal after application and to employ precise cleaning and suction during implant drilling and placement to minimize residual oil on the implant surface.</p>","PeriodicalId":94230,"journal":{"name":"The International journal of oral & maxillofacial implants","volume":" ","pages":"181-187"},"PeriodicalIF":0.0000,"publicationDate":"2025-04-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluating the Impact of Rotary Instrument Lubricants on Dental Implant Osseointegration: An Experimental Rabbit Tibia Study.\",\"authors\":\"Heon-Young Kim, Ju-Dong Song, Il-Seok Jang, Sun-Jong Kim, Jin-Woo Kim\",\"doi\":\"10.11607/jomi.11036\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To investigate the impact of mineral oil lubricants used in rotary instruments on osseointegration within rabbit tibias with a specific focus on potential contamination from dental handpieces.</p><p><strong>Materials and methods: </strong>A total of 12 New Zealand rabbits were included in this study, each receiving 2 implants in each tibia, resulting in a total of 48 implants across the study. Groups were organized based on the time until euthanasia and the degree of implant contamination. Three contamination levels were defined: The first group received implants without any lubricant in the handpiece (control group); the second group received implants with handpieces managed as recommended; and the third group received presoaked implants that were placed via a handpiece treated as per the second group's protocol. These groups were further subdivided based on euthanization periods of 2 and 4 weeks. Both the removal torque and the bone-to-implant contact (BIC) were measured and analyzed.</p><p><strong>Results: </strong>No significant correlation was observed between the degree of implant contamination and removal torque. However, there was a significant reduction in BIC associated with higher contamination levels, particularly after 4 weeks.</p><p><strong>Conclusions: </strong>Even brief exposure to lubricants from handpieces can jeopardize the osseointegration of implants in bone. Therefore, it is imperative to implement thorough procedures for lubricant removal after application and to employ precise cleaning and suction during implant drilling and placement to minimize residual oil on the implant surface.</p>\",\"PeriodicalId\":94230,\"journal\":{\"name\":\"The International journal of oral & maxillofacial implants\",\"volume\":\" \",\"pages\":\"181-187\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-04-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The International journal of oral & maxillofacial implants\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.11607/jomi.11036\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The International journal of oral & maxillofacial implants","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11607/jomi.11036","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:牙科植入物的骨结合受到各种表面特性的极大影响,如化学性质、纹理和整体清洁度。本研究旨在调查旋转器械中使用的矿物油润滑剂对兔胫骨骨结合的影响,特别关注牙科手钳可能造成的污染。材料和方法:本研究共纳入 12 只新西兰兔,每只兔的每根胫骨都接受了两个植入物,因此整个研究中共有 48 个植入物。根据安乐死前的时间和种植体污染程度分组。污染程度分为三个等级:第一组植入的植入体手机中未添加任何润滑剂(对照组);第二组植入的植入体手机按建议进行了管理;第三组植入的植入体使用了预先浸泡在润滑剂中的夹具。根据两周和四周的安乐死期,这些组别又进一步细分。我们测量并分析了移除扭矩和骨-种植体接触情况。结果:我们观察到夹具污染的严重程度与拔除扭矩之间存在非显著的反相关性。然而,污染程度越高,骨与种植体的接触越少,尤其是在四周之后。结论:即使是短暂接触手机的润滑剂,也会危及种植体在骨中的骨结合。因此,必须在使用后彻底清除润滑剂,并在种植体钻孔和植入过程中进行精确的清洁和抽吸,以尽量减少种植体表面的残留油。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluating the Impact of Rotary Instrument Lubricants on Dental Implant Osseointegration: An Experimental Rabbit Tibia Study.

Purpose: To investigate the impact of mineral oil lubricants used in rotary instruments on osseointegration within rabbit tibias with a specific focus on potential contamination from dental handpieces.

Materials and methods: A total of 12 New Zealand rabbits were included in this study, each receiving 2 implants in each tibia, resulting in a total of 48 implants across the study. Groups were organized based on the time until euthanasia and the degree of implant contamination. Three contamination levels were defined: The first group received implants without any lubricant in the handpiece (control group); the second group received implants with handpieces managed as recommended; and the third group received presoaked implants that were placed via a handpiece treated as per the second group's protocol. These groups were further subdivided based on euthanization periods of 2 and 4 weeks. Both the removal torque and the bone-to-implant contact (BIC) were measured and analyzed.

Results: No significant correlation was observed between the degree of implant contamination and removal torque. However, there was a significant reduction in BIC associated with higher contamination levels, particularly after 4 weeks.

Conclusions: Even brief exposure to lubricants from handpieces can jeopardize the osseointegration of implants in bone. Therefore, it is imperative to implement thorough procedures for lubricant removal after application and to employ precise cleaning and suction during implant drilling and placement to minimize residual oil on the implant surface.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信